Comparison of anterior chamber depth measurements by 3-dimensional optical coherence tomography, partial coherence interferometry biometry, Scheimpflug rotating camera imaging, and ultrasound biomicroscopy.

Shunsuke Nakakura1, Etsuko Mori, Nozomi Nagatomi, Hitoshi Tabuchi, Yoshiaki Kiuchi.   

Abstract

PURPOSE: To evaluate the congruity of anterior chamber depth (ACD) measurements using 4 devices.
SETTING: Saneikai Tsukazaki Hospital, Himeji City, Japan.
DESIGN: Comparative case series.
METHODS: In 1 eye of 42 healthy participants, the ACD was measured by 3-dimensional corneal and anterior segment optical coherence tomography (CAS-OCT), partial coherence interferometry (PCI), Scheimpflug imaging, and ultrasound biomicroscopy (UBM). The differences between the measurements were evaluated by 2-way analysis of variance and post hoc analysis. Agreement between the measurements was evaluated using Bland-Altman analysis. To evaluate the true ACD using PCI, the automatically calculated ACD minus the central corneal thickness measured by CAS-OCT was defined as PCI true. Two ACD measurements were also taken with CAS-OCT.
RESULTS: The mean ACD was 3.72 mm ± 0.23 (SD) (PCI), 3.18 ± 0.23 mm (PCI true), 3.24 ± 0.25 mm (Scheimpflug), 3.03 ± 0.25 mm (UBM), 3.14 ± 0.24 mm (CAS-OCT auto), and 3.12 ± 0.24 mm (CAS-OCT manual). A significant difference was observed between PCI biometry, Scheimpflug imaging, and UBM measurements and the other methods. Post hoc analysis showed no significant differences between PCI true and CAS-OCT auto or between CAS-OCT auto and CAS-OCT manual. Strong correlations were observed between all measurements; however, Bland-Altman analysis showed good agreement only between PCI true and Scheimpflug imaging and between CAS-OCT auto and CAS OCT manual.
CONCLUSION: The ACD measurements obtained from PCI biometry, Scheimpflug imaging, CAS-OCT, and UBM were significantly different and not interchangeable except for PCI true and CAS-OCT auto and CAS-OCT auto and CAS-OCT manual. FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.
Copyright © 2012 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22613688     DOI: 10.1016/j.jcrs.2012.02.036

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  6 in total

1.  Comparison of anterior section parameters using anterior segment optical coherence tomography and ultrasound biomicroscopy in myopic patients after ICL implantation.

Authors:  Jing Zhang; Hui-Hui Luo; Jing Zhuang; Ke-Ming Yu
Journal:  Int J Ophthalmol       Date:  2016-01-18       Impact factor: 1.779

2.  Comparison of Scheimpflug and swept-source anterior segment optical coherence tomography in normal and keratoconus eyes.

Authors:  Seyed Mohammad Ghoreishi; Seyed Ali Akbar Mortazavi; Zahra-Alsadat Abtahi; Mohammad-Ali Abtahi; Seyed Ali Sonbolestan; Seyed-Hossein Abtahi; Mohadeseh Mohammadinia; Kimia Nekouie Isfahani
Journal:  Int Ophthalmol       Date:  2016-09-26       Impact factor: 2.031

3.  Diagnosis of lacrimal canalicular diseases using ultrasound biomicroscopy: a preliminary study.

Authors:  Hai Tao; Li-Ping Xu; Cui Han; Peng Wang; Fang Bai
Journal:  Int J Ophthalmol       Date:  2014-08-18       Impact factor: 1.779

4.  Comparison of anterior segment measurements obtained by three different devices in healthy eyes.

Authors:  Carmen Lopez de la Fuente; Ana Sanchez-Cano; Francisco Segura; Isabel Pinilla
Journal:  Biomed Res Int       Date:  2014-06-02       Impact factor: 3.411

5.  Comparative study of anterior segment measurements using 3 different instruments in myopic patients after ICL implantation.

Authors:  Ting Wan; Houfa Yin; Yi Yang; Fang Wu; Zhiyi Wu; Yabo Yang
Journal:  BMC Ophthalmol       Date:  2019-08-13       Impact factor: 2.209

6.  OCT on Anterior Segment Anomalies of the Eye in a Polish Paediatric Cohort: Diagnostic and Therapeutic Challenges.

Authors:  Katarzyna Lepska; Dariusz Dobrowolski; Katarzyna Krysik; Anita Lyssek-Boroń; Edward Wylęgała
Journal:  Biomed Res Int       Date:  2021-02-18       Impact factor: 3.411

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.