Literature DB >> 23584925

Effective dose span of ten different cone beam CT devices.

D Rottke1, S Patzelt, P Poxleitner, D Schulze.   

Abstract

OBJECTIVES: Evaluation and reduction of dose are important issues. Since cone beam CT (CBCT) has been established now not just in dentistry, the number of acquired examinations continues to rise. Unfortunately, it is very difficult to compare the doses of available devices on the market owing to different exposition parameters, volumes and geometries. The aim of this study was to evaluate the spans of effective doses (EDs) of ten different CBCT devices.
METHODS: 48 thermoluminescent dosemeters were placed in 24 sites in a RANDO(®) head phantom. Protocols with lowest exposition parameters and protocols with highest exposition parameters were performed for each of the ten devices. The ED was calculated from the measured energy doses according to the International Commission on Radiological Protection 2007 recommendations for each protocol and device, and the statistical values were evaluated afterwards.
RESULTS: The calculation of the ED resulted in values between 17.2 µSv and 396 µSv for the ten devices. The mean values for protocols with lowest and highest exposition parameters were 31.6 µSv and 209 µSv, respectively.
CONCLUSIONS: It was not the aim of this study to evaluate the image quality depending on different exposition parameters but to define the spans of EDs in which different CBCT devices work. There is a wide span of ED for different CBCT devices depending on the selected exposition parameters, required spatial resolution and many other factors.

Keywords:  cone beam computed tomography; dose range; effective doses; thermoluminescent dosimetry

Mesh:

Year:  2013        PMID: 23584925      PMCID: PMC3699973          DOI: 10.1259/dmfr.20120417

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  8 in total

1.  The comparison of radiation dose between C-arm flat-detector CT (DynaCT) and multi-slice CT (MSCT): a phantom study.

Authors:  Mei Bai; Bin Liu; Hongqiu Mu; Xianghua Liu; Yuan Jiang
Journal:  Eur J Radiol       Date:  2011-10-02       Impact factor: 3.528

2.  Effective dose range for dental cone beam computed tomography scanners.

Authors:  Ruben Pauwels; Jilke Beinsberger; Bruno Collaert; Chrysoula Theodorakou; Jessica Rogers; Anne Walker; Lesley Cockmartin; Hilde Bosmans; Reinhilde Jacobs; Ria Bogaerts; Keith Horner
Journal:  Eur J Radiol       Date:  2010-12-31       Impact factor: 3.528

3.  Dosimetry of 3 CBCT devices for oral and maxillofacial radiology: CB Mercuray, NewTom 3G and i-CAT.

Authors:  J B Ludlow; L E Davies-Ludlow; S L Brooks; W B Howerton
Journal:  Dentomaxillofac Radiol       Date:  2006-07       Impact factor: 2.419

4.  Effective dose from cone beam CT examinations in dentistry.

Authors:  J A Roberts; N A Drage; J Davies; D W Thomas
Journal:  Br J Radiol       Date:  2008-10-13       Impact factor: 3.039

5.  A manufacturer's role in reducing the dose of cone beam computed tomography examinations: effect of beam filtration.

Authors:  J B Ludlow
Journal:  Dentomaxillofac Radiol       Date:  2011-02       Impact factor: 2.419

6.  Dosimetry of a cone beam CT device for oral and maxillofacial radiology using Monte Carlo techniques and ICRP adult reference computational phantoms.

Authors:  J J Morant; M Salvadó; I Hernández-Girón; R Casanovas; R Ortega; A Calzado
Journal:  Dentomaxillofac Radiol       Date:  2012-08-29       Impact factor: 2.419

7.  Assessment of radiation exposure in dental cone-beam computerized tomography with the use of metal-oxide semiconductor field-effect transistor (MOSFET) dosimeters and Monte Carlo simulations.

Authors:  J Koivisto; T Kiljunen; M Tapiovaara; J Wolff; M Kortesniemi
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol       Date:  2012-09

8.  Comparison of effective dose for imaging of mandible between multi-detector CT and cone-beam CT.

Authors:  Dae-Kyo Jeong; Sang-Chul Lee; Kyung-Hoe Huh; Won-Jin Yi; Min-Suk Heo; Sam-Sun Lee; Soon-Chul Choi
Journal:  Imaging Sci Dent       Date:  2012-06-25
  8 in total
  20 in total

Review 1.  Effective dose of cone beam CT (CBCT) of the facial skeleton: a systematic review.

Authors:  A Al-Okshi; C Lindh; H Salé; M Gunnarsson; M Rohlin
Journal:  Br J Radiol       Date:  2015-01       Impact factor: 3.039

2.  The effective dose of different scanning protocols using the Sirona GALILEOS(®) comfort CBCT scanner.

Authors:  D Chambers; R Bohay; L Kaci; R Barnett; J Battista
Journal:  Dentomaxillofac Radiol       Date:  2014-10-31       Impact factor: 2.419

3.  TLD measurements and Monte Carlo calculations of head and neck organ and effective doses for cone beam computed tomography using 3D Accuitomo 170.

Authors:  Marina Ernst; Peter Manser; Karl Dula; Werner Volken; Marco Fm Stampanoni; Michael K Fix
Journal:  Dentomaxillofac Radiol       Date:  2017-07-27       Impact factor: 2.419

4.  Investigation of an optimized scanning protocol for the dentomaxillofacial region using 320-slice multidetector computed tomography.

Authors:  Wei-Chan Lin; Hsueh-Han Wang; Wen-Lin Hsu; Bi-Hui Cao; De-Ji Chen; Chia-Jung Tsai
Journal:  Dentomaxillofac Radiol       Date:  2017-03-07       Impact factor: 2.419

5.  A new method to evaluate image quality of CBCT images quantitatively without observers.

Authors:  Yohei Takeshita; Mayumi Shimizu; Kazutoshi Okamura; Shoko Yoshida; Warangkana Weerawanich; Kenji Tokumori; Gainer R Jasa; Kazunori Yoshiura
Journal:  Dentomaxillofac Radiol       Date:  2017-02-17       Impact factor: 2.419

6.  Effective dose assessment in the maxillofacial region using thermoluminescent (TLD) and metal oxide semiconductor field-effect transistor (MOSFET) dosemeters: a comparative study.

Authors:  J Koivisto; D Schulze; J Wolff; D Rottke
Journal:  Dentomaxillofac Radiol       Date:  2014-08-21       Impact factor: 2.419

7.  Effective radiation dose and eye lens dose in dental cone beam CT: effect of field of view and angle of rotation.

Authors:  R Pauwels; G Zhang; C Theodorakou; A Walker; H Bosmans; R Jacobs; R Bogaerts; K Horner
Journal:  Br J Radiol       Date:  2014-10       Impact factor: 3.039

Review 8.  Guidelines for clinical use of CBCT: a review.

Authors:  K Horner; L O'Malley; K Taylor; A-M Glenny
Journal:  Dentomaxillofac Radiol       Date:  2015       Impact factor: 2.419

9.  Evidence of genotoxicity and cytotoxicity of X-rays in the oral mucosa epithelium of adults subjected to cone beam CT.

Authors:  Juliana Bm da Fonte; Taís M de Andrade; Ricardo Lc Albuquerque; Maria de Fátima B de Melo; Wilton M Takeshita
Journal:  Dentomaxillofac Radiol       Date:  2017-11-13       Impact factor: 2.419

Review 10.  Clinical guidelines for dental cone-beam computed tomography.

Authors:  Takafumi Hayashi; Yoshinori Arai; Toru Chikui; Sachiko Hayashi-Sakai; Kazuya Honda; Hiroko Indo; Taisuke Kawai; Kaoru Kobayashi; Shumei Murakami; Masako Nagasawa; Munetaka Naitoh; Eiji Nakayama; Yutaka Nikkuni; Hideyoshi Nishiyama; Noriaki Shoji; Shigeaki Suenaga; Ray Tanaka
Journal:  Oral Radiol       Date:  2018-01-11       Impact factor: 1.852

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.