Effective radiation dose of ProMax 3D cone-beam computerized tomography scanner with different dental protocols.

Xing-min Qu1, Gang Li, John B Ludlow, Zu-yan Zhang, Xu-chen Ma.   

Abstract

OBJECTIVES: The aim of this study was to compare effective doses resulting from different scan protocols for cone-beam computerized tomography (CBCT) using International Commission on Radiological Protection (ICRP) 1990 and 2007 calculations of dose. STUDY
DESIGN: Average tissue-absorbed dose, equivalent dose, and effective dose for a ProMax 3D CBCT with different dental protocols were calculated using thermoluminescent dosimeter chips in a human equivalent phantom. Effective doses were derived using ICRP 1990 and the superseding 2007 recommendations.
RESULTS: Effective doses (ICRP 2007) for default patient sizes from small to large ranged from 102 to 298 μSv. The coefficient of determination (R(2)) between tube current and effective dose (ICRP 2007) was 0.90. When scanning with lower resolution settings, the effective doses were reduced significantly (P < .05).
CONCLUSIONS: ProMax 3D can provide a wide range of radiation dose levels. Reduction in radiation dose can be achieved when using lower settings of exposure parameters.
Copyright © 2010 Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20952220     DOI: 10.1016/j.tripleo.2010.06.013

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  32 in total

1.  Dose reduction of cone beam CT scanning for the entire oral and maxillofacial regions with thyroid collars.

Authors:  X M Qu; G Li; G C H Sanderink; Z Y Zhang; X C Ma
Journal:  Dentomaxillofac Radiol       Date:  2012-07       Impact factor: 2.419

2.  Shielding effect of thyroid collar for digital panoramic radiography.

Authors:  G-S Han; J-G Cheng; G Li; X-C Ma
Journal:  Dentomaxillofac Radiol       Date:  2013-09-04       Impact factor: 2.419

3.  Detection accuracy of proximal caries by phosphor plate and cone-beam computerized tomography images scanned with different resolutions.

Authors:  Jun-Ge Cheng; Zhi-Ling Zhang; Xiao-Yan Wang; Zu-Yan Zhang; Xu-Chen Ma; Gang Li
Journal:  Clin Oral Investig       Date:  2011-07-30       Impact factor: 3.573

Review 4.  Effective dose of cone beam CT (CBCT) of the facial skeleton: a systematic review.

Authors:  A Al-Okshi; C Lindh; H Salé; M Gunnarsson; M Rohlin
Journal:  Br J Radiol       Date:  2015-01       Impact factor: 3.039

5.  [Diagnostic accuracy of cone beam computed tomography with different resolution settings for external root resorption].

Authors:  X Y Xie; S M Jia; Z H Sun; Z Y Zhang
Journal:  Beijing Da Xue Xue Bao Yi Xue Ban       Date:  2019-02-18

6.  Dose optimization for assessment of periodontal structures in cone beam CT examinations.

Authors:  Ayman Al-Okshi; Chrysoula Theodorakou; Christina Lindh
Journal:  Dentomaxillofac Radiol       Date:  2017-02-17       Impact factor: 2.419

7.  Dosimetry of a cone beam CT device for oral and maxillofacial radiology using Monte Carlo techniques and ICRP adult reference computational phantoms.

Authors:  J J Morant; M Salvadó; I Hernández-Girón; R Casanovas; R Ortega; A Calzado
Journal:  Dentomaxillofac Radiol       Date:  2012-08-29       Impact factor: 2.419

8.  Using GafChromic film to estimate the effective dose from dental cone beam CT and panoramic radiography.

Authors:  A Al-Okshi; M Nilsson; A Petersson; M Wiese; C Lindh
Journal:  Dentomaxillofac Radiol       Date:  2013-04-22       Impact factor: 2.419

9.  Patient radiation dose and protection from cone-beam computed tomography.

Authors:  Gang Li
Journal:  Imaging Sci Dent       Date:  2013-06-14

10.  Detection accuracy of condylar bony defects in Promax 3D cone beam CT images scanned with different protocols.

Authors:  Z-L Zhang; J-G Cheng; G Li; X-Q Shi; J-Z Zhang; Z-Y Zhang; X-C Ma
Journal:  Dentomaxillofac Radiol       Date:  2013-02-18       Impact factor: 2.419

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.