Effect of leaded glasses and thyroid shielding on cone beam CT radiation dose in an adult female phantom.

A D Goren1, R D Prins, L T Dauer, B Quinn, A Al-Najjar, R D Faber, G Patchell, I Branets, D C Colosi.   

Abstract

OBJECTIVES: This study aims to demonstrate the effectiveness of leaded glasses in reducing the lens of eye dose and of lead thyroid collars in reducing the dose to the thyroid gland of an adult female from dental cone beam CT (CBCT). The effect of collimation on the radiation dose in head organs is also examined.
METHODS: Dose measurements were conducted by placing optically stimulated luminescent dosemeters in an anthropomorphic female phantom. Eye lens dose was measured by placing a dosemeter on the anterior surface of the phantom eye location. All exposures were performed on one commercially available dental CBCT machine, using selected collimation and exposure techniques. Each scan technique was performed without any lead shielding and then repeated with lead shielding in place. To calculate the percent reduction from lead shielding, the dose measured with lead shielding was divided by the dose measured without lead shielding. The percent reduction from collimation was calculated by comparing the dose measured with collimation to the dose measured without collimation.
RESULTS: The dose to the internal eye for one of the scans without leaded glasses or thyroid shield was 0.450 cGy and with glasses and thyroid shield was 0.116 cGy (a 74% reduction). The reduction to the lens of the eye was from 0.396 cGy to 0.153 cGy (a 61% reduction). Without glasses or thyroid shield, the thyroid dose was 0.158 cGy; and when both glasses and shield were used, the thyroid dose was reduced to 0.091 cGy (a 42% reduction).
CONCLUSIONS: Collimation alone reduced the dose to the brain by up to 91%, with a similar reduction in other organs. Based on these data, leaded glasses, thyroid collars and collimation minimize the dose to organs outside the field of view.

Mesh:

Substances:

Year:  2013        PMID: 23412460      PMCID: PMC3667524          DOI: 10.1259/dmfr.20120260

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  25 in total

1.  Radiological protection of the worker in medicine and dentistry. A report of a Task Group of Committee 3 of the International Commission on Radiological Protection.

Authors: 
Journal:  Ann ICRP       Date:  1989

2.  Dose reduction in maxillofacial imaging using low dose Cone Beam CT.

Authors:  Kostas Tsiklakis; Catherine Donta; Sophia Gavala; Kety Karayianni; Vasiliki Kamenopoulou; Costas J Hourdakis
Journal:  Eur J Radiol       Date:  2005-06-22       Impact factor: 3.528

3.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103.

Authors: 
Journal:  Ann ICRP       Date:  2007

4.  Diagnostic X-ray exposure and female papillary thyroid cancer: a pooled analysis of two Swedish studies.

Authors:  G Wingren; A Hallquist; L Hardell
Journal:  Eur J Cancer Prev       Date:  1997-12       Impact factor: 2.497

5.  Significant reduction in dental cone beam computed tomography (CBCT) eye dose through the use of leaded glasses.

Authors:  R Prins; L T Dauer; D C Colosi; B Quinn; N J Kleiman; G C Bohle; B Holohan; A Al-Najjar; T Fernandez; M Bonvento; R D Faber; H Ching; A D Goren
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2011-07-28

6.  A leaded apron for use in panoramic dental radiography.

Authors:  B L Whitcher; B M Gratt; E A Sickles
Journal:  Oral Surg Oral Med Oral Pathol       Date:  1980-05

Review 7.  Medical exposure to radiation and thyroid cancer.

Authors:  S J Schonfeld; C Lee; A Berrington de González
Journal:  Clin Oncol (R Coll Radiol)       Date:  2011-02-05       Impact factor: 4.126

8.  Radiation dose to the thyroid gland and breast from multidetector computed tomography of the cervical spine: does bismuth shielding with and without a cervical collar reduce dose?

Authors:  Martin L Gunn; Kalpana M Kanal; Orpheus Kolokythas; Yoshimi Anzai
Journal:  J Comput Assist Tomogr       Date:  2009 Nov-Dec       Impact factor: 1.826

9.  Medical diagnostic and therapeutic ionizing radiation and the risk for thyroid cancer: a case-control study.

Authors:  A Hallquist; L Hardell; A Degerman; G Wingren; L Boquist
Journal:  Eur J Cancer Prev       Date:  1994-05       Impact factor: 2.497

10.  Thyroid shields and neck exposures in cephalometric radiography.

Authors:  Philippe Hujoel; Lars Hollender; Anne-Marie Bollen; John D Young; Joana Cunha-Cruz; Molly McGee; Alex Grosso
Journal:  BMC Med Imaging       Date:  2006-06-13       Impact factor: 1.930

View more
  8 in total

1.  Prevalence of Protective Shielding Utilization for Radiation Dose Reduction in Adult Patients Undergoing Body Scanning Using Computed Tomography.

Authors:  Shoaib Safiullah; Roshan Patel; Brittany Uribe; Kyle Spradling; Chandana Lall; Lishi Zhang; Zhamshid Okhunov; Ralph V Clayman; Jaime Landman
Journal:  J Endourol       Date:  2017-08-14       Impact factor: 2.942

2.  Thyroid shielding in cone beam computed tomography: recommendations towards appropriate use.

Authors:  Ruben Pauwels; Keith Horner; Jenia Vassileva; Madan M Rehani
Journal:  Dentomaxillofac Radiol       Date:  2019-07-09       Impact factor: 2.419

3.  The relationship between 3D dentofacial photogrammetry measurements and traditional cephalometric measurements.

Authors:  Jose C Castillo; Grace Gianneschi; Demyana Azer; Amornrut Manosudprasit; Arshan Haghi; Neetu Bansal; Veerasathpurush Allareddy; Mohamed I Masoud
Journal:  Angle Orthod       Date:  2019-03       Impact factor: 2.079

4.  Can modifying shielding, field of view, and exposure settings make the effective dose of a cone-beam computed tomography comparable to traditional radiographs used for orthodontic diagnosis?

Authors:  Stephanie Ting; Diana Attaia; K Brandon Johnson; Samer Shoukry Kossa; Bernard Friedland; Veerasathpurush Allareddy; Mohamed I Masoud
Journal:  Angle Orthod       Date:  2020-09-01       Impact factor: 2.079

5.  Effective radiation dose and eye lens dose in dental cone beam CT: effect of field of view and angle of rotation.

Authors:  R Pauwels; G Zhang; C Theodorakou; A Walker; H Bosmans; R Jacobs; R Bogaerts; K Horner
Journal:  Br J Radiol       Date:  2014-10       Impact factor: 3.039

6.  Effectiveness of thyroid gland shielding in dental CBCT using a paediatric anthropomorphic phantom.

Authors:  A Hidalgo; J Davies; K Horner; C Theodorakou
Journal:  Dentomaxillofac Radiol       Date:  2014-11-20       Impact factor: 2.419

Review 7.  Clinical guidelines for dental cone-beam computed tomography.

Authors:  Takafumi Hayashi; Yoshinori Arai; Toru Chikui; Sachiko Hayashi-Sakai; Kazuya Honda; Hiroko Indo; Taisuke Kawai; Kaoru Kobayashi; Shumei Murakami; Masako Nagasawa; Munetaka Naitoh; Eiji Nakayama; Yutaka Nikkuni; Hideyoshi Nishiyama; Noriaki Shoji; Shigeaki Suenaga; Ray Tanaka
Journal:  Oral Radiol       Date:  2018-01-11       Impact factor: 1.852

8.  Dosimetry of Critical Organs in Maxillofacial Imaging with Cone-beam Computed Tomography.

Authors:  R Ghanbarnezhad Farshi; A Mesbahi; M Johari; Ü Kara; N Gharehaghaji
Journal:  J Biomed Phys Eng       Date:  2019-02-01
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.