Giancarlo Cordasco1, Marco Portelli, Angela Militi, Riccardo Nucera, Antonino Lo Giudice, Elda Gatto, Alessandra Lucchese.
Abstract
BACKGROUND: The aim of this study was to evaluate the amount of radiation doses absorbed by soft tissues (entrance skin dose) with a low-dose spiral computed tomography (CT) protocol compared to conventional X-ray techniques commonly used in orthodontics.Entities:
Mesh:
Year: 2013 PMID: 24325970 PMCID: PMC4384968 DOI: 10.1186/2196-1042-14-24
Source DB: PubMed Journal: Prog Orthod ISSN: 1723-7785 Impact factor: 2.750
Figure 1CT scan (Somaton Sensation 16 Siemens®).
Figure 2CT images.
Absorbed doses (mGy)
| X-ray technique | Right eye lens | Left eye lens | Right parotid gland | Left parotid gland | Right thyroid gland | Left thyroid gland | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Mean | SD | Mean | SD | Mean | SD | Mean | SD | Mean | SD | Mean | SD | |
| LC | 0.06 | 0.012 | 0.09 | 0.021 | 0.04 | 0.0098 | 0.3 | 0.052 | 0.06 | 0.013 | 0.07 | 0.017 |
| PAC | 0.28 | 0.04 | 0.31 | 0.056 | 0.03 | 0.0087 | 0.05 | 0.011 | 0.07 | 0.018 | 0.04 | 0.009 |
| PR | 0.07 | 0.016 | 0.08 | 0.019 | 0.52 | 0.14 | 0.49 | 0.12 | 0.08 | 0.039 | 0.07 | 0.0195 |
| LC + PAC + PR | 0.41 | 0.097 | 0.48 | 0.101 | 0.59 | 0.158 | 0.84 | 0.32 | 0.21 | 0.097 | 0.18 | 0.022 |
| Low-dose CT | 0.57 | 0.11 | 0.58 | 0.12 | 0.78 | 0.289 | 0.79 | 0.30 | 0.18 | 0.081 | 0.17 | 0.067 |
For each X-ray technique and organ, the doses have been registered five times. The mean value and the standard deviation (SD) are reported.
Image quality scores attributed and mean absorbed doses registered
| Image quality | Mean absorbed dose (mGy) | |
|---|---|---|
| LC | Satisfactory (3) | 0.10 |
| PAC | Moderate (2) | 0.78 |
| PR | Good (4) | 0.21 |
| LC + PAC + PR | Satisfactory (3) | 0.45 |
| Low-dose CT | Excellent (5) | 0.51 |
Wilcoxon's signed-rank test to identify significant difference in terms of image quality and absorbed dose
| LC + PAC + PR | Low-dose CT |
| |
|---|---|---|---|
| Image quality | Satisfactory (3) | Excellent (5) | <0.05 |
| Mean absorbed dose (mGy) | 0.45 | 0.51 | NS |
The level of significance was set at P < 0.05. NS, not statistically significant.