| Literature DB >> 30547030 |
Antonio J Gonzalez1, Stuart S Berr2, Gabriel Cañizares1, Andrea Gonzalez-Montoro1, Abel Orero3, Carlos Correcher3, Ahmadreza Rezaei4, Johan Nuyts4,5, Filomeno Sanchez1, Stan Majewski2, Jose M Benlloch1.
Abstract
There are drawbacks with using a Positron Emission Tomography (PET) scanner design employing the traditional arrangement of multiple detectors in an array format. Typically PET systems are constructed with many regular gaps between the detector modules in a ring or box configuration, with additional axial gaps between the rings. Although this has been significantly reduced with the use of the compact high granularity SiPM photodetector technology, such a scanner design leads to a decrease in the number of annihilation photons that are detected causing lower scanner sensitivity. Moreover, the ability to precisely determine the line of response (LOR) along which the positron annihilated is diminished closer to the detector edges because the spatial resolution there is degraded due to edge effects. This happens for both monolithic based designs, caused by the truncation of the scintillation light distribution, but also for detector blocks that use crystal arrays with a number of elements that are larger than the number of photosensors and, therefore, make use of the light sharing principle. In this report we present a design for a small-animal PET scanner based on a single monolithic annulus-like scintillator that can be used as a PET insert in high-field Magnetic Resonance systems. We provide real data showing the performance improvement when edge-less modules are used. We also describe the specific proposed design for a rodent scanner that employs facetted outside faces in a single LYSO tube. In a further step, in order to support and prove the proposed edgeless geometry, simulations of that scanner have been performed and lately reconstructed showing the advantages of the design.Entities:
Keywords: Positron emission tomography; Preclinical imaging; SiPM; hybrid PET-MR; monolithic crystal
Year: 2018 PMID: 30547030 PMCID: PMC6279866 DOI: 10.3389/fmed.2018.00328
Source DB: PubMed Journal: Front Med (Lausanne) ISSN: 2296-858X
Figure 1Sketches of a PET scanner based on individual modules (A) and built as a single monolithic annulus-like crystal (B).
Figure 2(A) Sketch of the crystal with two scintillation light distributions, the dashed lines represent the 60% of the crystal volume. The flood map in (B) shows the 11 × 11 collimated positron-emitter sources.
Figure 3(A) Sketch of the design with 10-face scintillation tube. (B) Photograph of the manufactured monolithic LYSO tube.
Figure 4(A) Measured spatial resolution in a 10mm thick monolithic block as a function of the impact position. (B) Energy resolution dependence.
Figure 5(A) Reconstructed Derenzo-like phantom images using all crystal impacts. (B) The same reconstruction considering data only in the 30 × 30mm central region. (C) Profiles for both cases (yellow line is for the filtered case).
Figure 6(A) Image reconstruction of two point sources, 15mm separated. (B) Lines profiles across the two sources. (C) 3D reconstruction of the normalization annulus. (D) Reconstruction of the Derenzo-like phantom (transverse and coronal views). Notice that one of the hot rods was filled with 10% of the concentration in the other rods. (E) Line projections across the smallest rods.
Figure 7(A) Light distributions using 3mm bins for different impact positions (in angles). The red lines are Gaussian fits. (B) Measured centroid as a function of known position (in angle) showing a good linearity without edge effects.
Figure 8Light distributions for different DOI impact positions in the crystal, projected onto the axial (A) and transaxial planes (B), respectively.