Design and evaluation of the MAMMI dedicated breast PET.

L Moliner1, A J Gonzalez, A Soriano, F Sanchez, C Correcher, A Orero, M Carles, L F Vidal, J Barbera, L Caballero, M Seimetz, C Vazquez, J M Benlloch.   

Abstract

PURPOSE: A breast dedicated positron emission tomography (PET) scanner has been developed based on monolithic LYSO crystals coupled to position sensitive photomultiplier tubes (PSPMTs). In this study, we describe the design of the PET system and report on its performance evaluation.
METHODS: MAMMI is a breast PET scanner based on monolithic LYSO crystals. It consists of 12 compact modules with a transaxial field of view (FOV) of 170 mm in diameter and 40 mm axial FOV that translates to cover up to 170 mm. The patient lies down in a prone position that facilitates maximum breast elongation. Quantitative performance analysis of the calculated method for the attenuation correction specifically developed for MAMMI, and based on PET image segmentation, has also been conducted in this evaluation. In order to fully determine the MAMMI prototype's performance, we have adapted the measurements suggested for National Electrical Manufacturers Association (NEMA) NU 2-2007 and NU 4-2008 protocol tests, as they are defined for whole-body and small animal PET scanners, respectively.
RESULTS: Spatial resolutions of 1.6, 1.8, and 1.9 mm were measured in the axial, radial, and tangential directions, respectively. A scatter fraction of 20.8% was obtained and the maximum NEC was determined to be 25 kcps at 44 MBq. The average sensitivity of the system was observed to be 1% for an energy window of (250 keV-750 keV) and a maximum absolute sensitivity of 1.8% was measured at the FOV center.
CONCLUSIONS: The overall performance of the MAMMI reported on this evaluation quantifies its ability to produce high quality PET images. Spatial resolution values below 3 mm were measured in most of the FOV. Only the radial component of spatial resolution exceeds the 3 mm at radial positions larger than 60 mm. This study emphasizes the need for standardized testing methodologies for dedicated breast PET systems similar to NEMA standards for whole-body and small animal PET scanners.

Entities:  

Mesh:

Year:  2012        PMID: 22957607     DOI: 10.1118/1.4742850

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  22 in total

Review 1.  Innovations in Instrumentation for Positron Emission Tomography.

Authors:  Eric Berg; Simon R Cherry
Journal:  Semin Nucl Med       Date:  2018-03-12       Impact factor: 4.446

Review 2.  Nuclear imaging of the breast: translating achievements in instrumentation into clinical use.

Authors:  Carrie B Hruska; Michael K O'Connor
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

3.  Design study of a dedicated head and neck cancer PET system.

Authors:  Mohan Li; Brett Yockey; Shiva Abbaszadeh
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2020-01-08

4.  Molecular breast imaging: an emerging modality for breast cancer screening.

Authors:  Michael K O'Connor
Journal:  Breast Cancer Manag       Date:  2015-01-01

Review 5.  Update on latest advances in time-of-flight PET.

Authors:  Suleman Surti; Joel S Karp
Journal:  Phys Med       Date:  2020-11-16       Impact factor: 2.685

6.  Simulation study of quantitative precision of the PET/X dedicated breast PET scanner.

Authors:  Chengeng Zeng; Paul E Kinahan; Hua Qian; Robert L Harrison; Kyle M Champley; Lawrence R MacDonald
Journal:  J Med Imaging (Bellingham)       Date:  2017-10-30

7.  Design Optimization of a TOF, Breast PET Scanner.

Authors:  Eunsin Lee; Matthew E Werner; Joel S Karp; Suleman Surti
Journal:  IEEE Trans Nucl Sci       Date:  2013-06       Impact factor: 1.679

8.  Texture analysis of high-resolution dedicated breast 18 F-FDG PET images correlates with immunohistochemical factors and subtype of breast cancer.

Authors:  Alexis Moscoso; Álvaro Ruibal; Inés Domínguez-Prado; Anxo Fernández-Ferreiro; Míchel Herranz; Luis Albaina; Sonia Argibay; Jesús Silva-Rodríguez; Juan Pardo-Montero; Pablo Aguiar
Journal:  Eur J Nucl Med Mol Imaging       Date:  2017-09-21       Impact factor: 9.236

9.  A dedicated breast-PET/CT scanner: Evaluation of basic performance characteristics.

Authors:  Raymond R Raylman; Will Van Kampen; Alexander V Stolin; Wenbo Gong; Gangadhar Jaliparthi; Peter F Martone; Mark F Smith; David Sarment; Neal H Clinthorne; Mark Perna
Journal:  Med Phys       Date:  2018-02-23       Impact factor: 4.071

Review 10.  Radionuclide methods and instrumentation for breast cancer detection and diagnosis.

Authors:  Suleman Surti
Journal:  Semin Nucl Med       Date:  2013-07       Impact factor: 4.446

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.