Literature DB >> 35169874

New Radioprotective Device that can be Used for Fluoroscopic Exam: Possibility to Contribute to Staff Exposure Protection During VFSS.

Yoshiaki Morishima1,2, Koichi Chida3, Osamu Ito4.   

Abstract

The videofluoroscopic swallowing study (VFSS) is a recognized standard diagnostic imaging technique that is used to investigate swallowing disorders and dysphagia. Patients were assessed in a seated posture on a chair or wheelchair. Using X-ray fluoroscopy, the state of patients' swallowing was checked by eating and drinking according to the physician's instructions. VFSS procedures are prolonged, and VFSS staff members are exposed to radiation. Therefore, we evaluated original lead shielding device (OLSD) that can be attached to the handrail of a table and placed vertically. The OLSD has a lead-equivalent thickness of 0.3 mmPb, weighs about 6 kg, and has the dimensions 50 cm × 50 cm × 8.0 mm. We used a human phantom and a radiation survey meter with and without protection from scattered radiation at the positions of the physician and medical staff at the height of 150 cm above the floor (i.e., the height of the eye's crystalline lens). After measuring the scattered radiation, we created radiation maps with and without the OLSD. The dose rate at the physician's position without and with the OLSD was 190 µSv/h and 92 µSv/h, respectively, and a dose reduction of 51.6% with the plate. Moreover, the radiation maps added clarity to the distribution of the scattered radiation. Such information should lead to greater awareness about exposures to physicians and other medical staff. Thus, the OLSD effectively provided protection from scattered radiation at the physician's position during fluoroscopy. It may contribute to the reduction of staff exposure for VFSS.
© 2022. The Author(s), under exclusive licence to Springer Science+Business Media, LLC, part of Springer Nature.

Entities:  

Keywords:  Fluoroscopy; Original lead shielding device; Scattered radiation protection; Swallowing disorders

Year:  2022        PMID: 35169874     DOI: 10.1007/s00455-022-10411-x

Source DB:  PubMed          Journal:  Dysphagia        ISSN: 0179-051X            Impact factor:   3.438


  20 in total

1.  A double-blind, randomized, sham-controlled trial of the effect of a radiation-attenuating drape on radiation exposure to endoscopy staff during ERCP.

Authors:  Thiruvengadam Muniraj; Harry R Aslanian; Loren Laine; James Farrell; Maria M Ciarleglio; Yanhong Deng; Henry Ho; Priya A Jamidar
Journal:  Am J Gastroenterol       Date:  2015-03-31       Impact factor: 10.864

2.  Radiation protection in digestive endoscopy: European Society of Digestive Endoscopy (ESGE) guideline.

Authors:  J-M Dumonceau; F J Garcia-Fernandez; F R Verdun; E Carinou; L Donadille; J Damilakis; I Mouzas; K Paraskeva; N Ruiz-Lopez; L Struelens; V Tsapaki; F Vanhavere; V Valatas; M Sans-Merce
Journal:  Endoscopy       Date:  2012-03-21       Impact factor: 10.093

3.  Patients' radiation dose during videofluoroscopic swallowing studies according to underlying characteristics.

Authors:  Hong Min Kim; Kyoung Hyo Choi; Tae Woo Kim
Journal:  Dysphagia       Date:  2012-09-09       Impact factor: 3.438

4.  Radiation exposure from videofluoroscopic swallow studies in children with a type 1 laryngeal cleft and pharyngeal dysphagia: A retrospective review.

Authors:  Cheryl Hersh; Carissa Wentland; Sarah Sally; Marie de Stadler; Steven Hardy; M Shannon Fracchia; Bob Liu; Christopher Hartnick
Journal:  Int J Pediatr Otorhinolaryngol       Date:  2016-07-26       Impact factor: 1.675

Review 5.  Radiation Exposure to Staff and Patient During Videofluoroscopic Swallowing Studies and Recommended Protection Strategies.

Authors:  Victoria Jean Earl; Mohamed Khaldoun Badawy
Journal:  Dysphagia       Date:  2018-09-22       Impact factor: 3.438

6.  Estimation of the Dose of Radiation Received by Patient and Physician During a Videofluoroscopic Swallowing Study.

Authors:  Yoshiaki Morishima; Koichi Chida; Hiroshi Watanabe
Journal:  Dysphagia       Date:  2016-06-18       Impact factor: 3.438

7.  Role of the modified barium swallow in management of patients with dysphagia.

Authors:  J A Logemann
Journal:  Otolaryngol Head Neck Surg       Date:  1997-03       Impact factor: 5.591

8.  Occupational Radiation Exposure during Endoscopic Retrograde Cholangiopancreatography and Usefulness of Radiation Protective Curtains.

Authors:  Tomoyuki Minami; Tamito Sasaki; Masahiro Serikawa; Michihiro Kamigaki; Masanobu Yukutake; Takashi Ishigaki; Yasutaka Ishii; Teruo Mouri; Satoshi Yoshimi; Akinori Shimizu; Tomofumi Tsuboi; Keisuke Kurihara; Yumiko Tatsukawa; Eisuke Miyaki; Kazuaki Chayama
Journal:  Gastroenterol Res Pract       Date:  2014-11-13       Impact factor: 2.260

9.  Effectiveness of additional lead shielding to protect staff from scattering radiation during endoscopic retrograde cholangiopancreatography procedures.

Authors:  Yoshiaki Morishima; Koichi Chida; Takayoshi Meguro
Journal:  J Radiat Res       Date:  2018-03-01       Impact factor: 2.724

10.  Radiation Dose Reduction during Radial Cardiac Catheterization: Evaluation of a Dedicated Radial Angiography Absorption Shielding Drape.

Authors:  Andrew Ertel; Jeffrey Nadelson; Adhir R Shroff; Ranya Sweis; Dean Ferrera; Mladen I Vidovich
Journal:  ISRN Cardiol       Date:  2012-09-04
View more