Huamao Miao1, Xun Chen1, Mi Tian1, Yingjun Chen1, Xiaoying Wang2, Xingtao Zhou3.
Abstract
BACKGROUND: To investigate refractive outcomes and optical quality after implantation of posterior chamber phakic implantable collamer lens with a central hole (ICL V4c) to correct high myopia.Entities:
Keywords: Intraocular pressure; Intraocular scattering; Myopia; Optical quality; Phakic intraocular Lens; Refractive outcomes; V4c
Mesh:
Year: 2018 PMID: 29898694 PMCID: PMC6001218 DOI: 10.1186/s12886-018-0805-3
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Fig. 1Refractive outcomes at 3 months postoperatively in 67 eyes with high myopia after implantation of posterior chamber phakic implantable collamer lens (ICL V4c) with a central hole, including (a) postoperative uncorrected distance visual acuity, (b) change in corrected distance visual acuity, (c) attempted versus achieved spherical equivalent, (d) spherical equivalent refractive accuracy, (e) refractive astigmatism, and (f) stability of spherical equivalent refraction. UDVA = uncorrected distance visual acuity; CDVA = corrected distance visual acuity; D = diopters; Postop = postoperative; Preop = preoperative; SD = standard deviation; mo = months
Refractive Outcomes at 1 and 3 Months after ICL V4c Implantation for Correcting High Myopia. (n = 35)
| Parameters | 1 Month | 3 Months |
| ||
|---|---|---|---|---|---|
| Mean | SD | Mean | SD | ||
| Sphere (D) | 0.09 | 0.56 | 0.10 | 0.54 | 0.906 |
| Cylinder (D) | −0.73 | 0.54 | −0.71 | 0.52 | 0.786 |
| SE (D) | −0.27 | 0.51 | −0.26 | 0.47 | 0.793 |
| logMAR UDVA | −0.01 | 0.11 | −0.01 | 0.09 | 0.580 |
| logMAR CDVA | −0.06 | 0.09 | −0.08 | 0.08 | 0.071 |
| Efficacy index | 1.11 | 0.22 | 1.13 | 0.24 | 0.797 |
| Safety index | 1.25 | 0.20 | 1.31 | 0.22 | 0.060 |
ICL V4c Posterior Chamber Phakic Implantable collamer lens With a Central Hole, n number of eyes, SD standard deviation, SE spherical equivalent refraction, D diopters, logMAR the logarithm of the minimal angle of resolution, UDVA uncorrected distance visual acuity, CDVA corrected distance visual acuity
Optical Quality at 1 and 3 Months after ICL V4c Implantation for Correcting High Myopia (n = 35)
| Parameters | 1 Month | 3 Months |
| ||
|---|---|---|---|---|---|
| Mean | SD | Mean | SD | ||
| MTFcutoff (cpd) | 38.49 | 10.35 | 37.64 | 9.74 | 0.475 |
| Strehl2D Ratio | 0.22 | 0.07 | 0.21 | 0.06 | 0.587 |
| OV100% | 1.28 | 0.35 | 1.25 | 0.32 | 0.475 |
| OV20% | 1.30 | 0.43 | 1.25 | 0.39 | 0.408 |
| OV9% | 1.30 | 0.47 | 1.24 | 0.44 | 0.401 |
| OSI | 1.00 | 0.83 | 0.98 | 0.71 | 0.954 |
MTF modulation transfer function cutoff frequency, cpd cycles per degree, Strehl2D Ratio Strehl in two dimensions ratio, OV optical quality analysis system value, OSI objective scatter index, SD standard deviation
Fig. 2Intraocular pressure after ICL V4c implantation