Visual quality comparison of conventional and Hole-Visian implantable collamer lens at different degrees of decentering.

Cari Pérez-Vives1, Teresa Ferrer-Blasco, David Madrid-Costa, Santiago García-Lázaro, Robert Montés-Micó.   

Abstract

PURPOSE: To compare the visual quality of implantable collamer lens (ICL) with and without central hole (Hole ICL and conventional ICL) at different degrees of decentering.
METHODS: An adaptive optics visual simulator (crx1, Imagine Eyes, Orsay, France) was used to simulate the -3, -6 and -12 dioptres (D) conventional and Hole ICLs in three conditions: centred and decentred 0.3 and 0.6 mm. Visual acuity (VA) at high-contrast, medium-contrast and low-contrast and contrast sensitivity (CS) were measured in 15 observers for 3 and 4.5 mm pupils.
RESULTS: No statistically significant differences in VA and CS were found between conventional and Hole ICLs for any ICL powers and pupil sizes evaluated (p>0.05). Regarding the effect of the ICL decentration on visual performance, we did not find statistically significant differences in VA and CS between centred, 0.3 and 0.6 mm decentred (p>0.05). Moreover, the ICL decentration affected the same manner on the conventional and Hole ICLs.
CONCLUSIONS: The outcomes showed that conventional and Hole ICLs provided good and comparable visual performance for all powers and pupil sizes evaluated. Besides, ICL decentering affects the same manner both ICL models evaluated. The ICL decentering did not have any effect on the visual performance.

Keywords:  Optics and Refraction

Mesh:

Year:  2013        PMID: 24246374     DOI: 10.1136/bjophthalmol-2013-303787

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  9 in total

1.  Comparison of postoperative optical quality according to the degree of decentering of V4c implantable collamer lens.

Authors:  Min Ji Park; Hye Min Jeon; Kyoung Heon Lee; Sang Youp Han
Journal:  Int J Ophthalmol       Date:  2017-04-18       Impact factor: 1.779

2.  Effects of tilt and decentration of Visian Implantable Collamer Lens (ICL V4c) on visual quality: an observational study.

Authors:  Lingling Niu; Zhe Zhang; Huamao Miao; Jing Zhao; Meiyan Li; Ji C He; Peijun Yao; Xingtao Zhou
Journal:  BMC Ophthalmol       Date:  2022-07-05       Impact factor: 2.086

Review 3.  Meta-analysis and review: effectiveness, safety, and central port design of the intraocular collamer lens.

Authors:  Mark Packer
Journal:  Clin Ophthalmol       Date:  2016-06-09

4.  Posterior chamber collagen copolymer phakic intraocular lens with a central hole for moderate-to-high myopia: First experience in China.

Authors:  Xinfang Cao; Weiliang Wu; Yang Wang; Chen Xie; Jianping Tong; Ye Shen
Journal:  Medicine (Baltimore)       Date:  2016-09       Impact factor: 1.889

Review 5.  In vivo optical quality of posterior-chamber phakic implantable collamer lenses with a central port.

Authors:  Robert Montés-Micó; Francisco Pastor-Pascual; Enrique Artiaga-Elordi; Ramón Ruiz-Mesa; Pedro Tañá-Rivero
Journal:  Eye Vis (Lond)       Date:  2021-08-16

6.  Pure ICL Implantation: A Novel Ophthalmic Viscosurgical Device-Free Method.

Authors:  Qin Qin; Lianyun Bao; Zifang He; Feifei Chen; Dandan Zhu; Si Zhang; Wenwen Zhang; Yajun Liu; Ruiying Gao; Zhenggao Xie
Journal:  J Ophthalmol       Date:  2021-10-06       Impact factor: 1.909

7.  Effects of V4c-ICL Implantation on Myopic Patients' Vision-Related Daily Activities.

Authors:  Taixiang Liu; Shaorong Linghu; Le Pan; Rong Shi
Journal:  J Ophthalmol       Date:  2016-11-14       Impact factor: 1.909

8.  Implantable collamer lens with central hole: 3-year follow-up.

Authors:  Luis Fernández-Vega-Cueto; Carlos Lisa; José J Esteve-Taboada; Robert Montés-Micó; José F Alfonso
Journal:  Clin Ophthalmol       Date:  2018-10-11

9.  Refractive outcomes and optical quality after implantation of posterior chamber phakic implantable collamer lens with a central hole (ICL V4c).

Authors:  Huamao Miao; Xun Chen; Mi Tian; Yingjun Chen; Xiaoying Wang; Xingtao Zhou
Journal:  BMC Ophthalmol       Date:  2018-06-14       Impact factor: 2.209

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.