BACKGROUND: The current study compared optical quality before and after implantation of a posterior chamber phakic intraocular lens with a central hole [V4c implantable collamer lens (ICL)] under bright and dark lighting conditions by means of the Optical Quality Analysis System™ (OQAS; Visiometrics, Terrassa, Spain).Entities:
Keywords: Implantable collamer lens with a central hole; Lighting condition; OQAS; Optical quality
Mesh:
Year: 2020 PMID: 32131800 PMCID: PMC7055093 DOI: 10.1186/s12886-020-01340-0
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Demographic data of the study population undergoing V4c ICL implantation
| Demographic data | |
| Number of patients (eyes) | 46 (91) |
| Sex (male:female) | 15:31 |
| Age (years) | 28.46 ± 0.53 (range: 18–44) |
| Preoperative | |
| Manifest spherical equivalent (D) | −10.50 ± 0.33 (range: − 21.75 to −5.75) |
| LogMAR UDVA | 1.66 ± 0.04 (range: 1–3) |
| LogMAR BDVA | 0.08 ± 0.01 (range: 0–0.7) |
| IOP (mmHg) | 14.37 ± 0.29 (range: 9–21) |
| Postoperative (3 months) | |
| Manifest spherical equivalent (D) | 0.03 ± 0.29 (range: −3.23–1) |
| LogMAR UDVA | 0.03 ± 0.04 (range: 0–0.7) |
| IOP (mmHg) | 13.96 ± 0.27 (range: 8–21) |
D diopters, logMAR logarithm of the minimal angle of resolution, UDVA uncorrected distance visual acuity, BDVA best-corrected distance visual acuity, IOP intraocular pressure
Fig. 1Optical quality preoperatively and 1 week, 1 month, 3 months postoperatively after implantation of posterior chamber phakic implantable collamer lens with a central hole. MTF, modulation transfer function; OSI, objective scattering index; PVA, predicted visual acuity
Changes in optical quality parameters in eyes undergoing V4c ICL implantation under different lighting conditions
| Time | Parameter | Bright | Dark | |
|---|---|---|---|---|
| 1 week | ΔOSI | −0.606 ± 0.131 | − 0.920 ± 0.207 | 0.02 |
| ΔMTF cutoff frequency (cpd) | 2.789 ± 1.133 | 4.187 ± 1.274 | 0.22 | |
| ΔStrehl ratio | 0.013 ± 0.007 | 0.019 ± 0.006 | 0.37 | |
| ΔPVA100% | 0.099 ± 0.038 | 0.131 ± 0.042 | 0.40 | |
| ΔPVA20% | 0.060 ± 0.029 | 0.086 ± 0.033 | 0.38 | |
| ΔPVA9% | 0.032 ± 0.021 | 0.051 ± 0.019 | 0.37 | |
| 1 month | ΔOSI | −0.686 ± 0.120 | −1.034 ± 0.217 | 0.004 |
| ΔMTF cutoff frequency (cpd) | 4.589 ± 1.109 | 4.462 ± 1.136 | 0.91 | |
| ΔStrehl ratio | 0.018 ± 0.007 | 0.019 ± 0.007 | 0.95 | |
| ΔPVA100% | 0.164 ± 0.036 | 0.143 ± 0.046 | 0.56 | |
| ΔPVA20% | 0.111 ± 0.029 | 0.113 ± 0.035 | 0.94 | |
| ΔPVA9% | 0.049 ± 0.021 | 0.052 ± 0.021 | 0.88 | |
| 3 months | ΔOSI | −0.689 ± 0.121 | −1.056 ± 0.217 | 0.002 |
| ΔMTF cutoff frequency (cpd) | 6.155 ± 1.207 | 6.627 ± 1.214 | 0.70 | |
| ΔStrehl ratio | 0.025 ± 0.008 | 0.031 ± 0.007 | 0.44 | |
| ΔPVA100% | 0.212 ± 0.039 | 0.209 ± 0.041 | 0.94 | |
| ΔPVA20% | 0.136 ± 0.033 | 0.162 ± 0.032 | 0.44 | |
| ΔPVA9% | 0.078 ± 0.023 | 0.094 ± 0.023 | 0.48 |
ICL implantable collamer lens, MTF modulation transfer function, OSI objective scattering index, PVA predicted visual acuity
Comparison of changes in optical quality between Group H and S under bright conditions
| Parameter | Time | Group H | Group S | P-value |
|---|---|---|---|---|
| ΔOSI | 1 week | − 0.063 ± 0.098 | −1.127 ± 0.136 | 0.005 |
| 1 month | −0.108 ± 0.088 | −1.267 ± 0.151 | 0.003 | |
| 3 months | −0.158 ± 0.098 | −1.227 ± 0.127 | 0.004 | |
| ΔMTF cutoff frequency | 1 week | 0.018 ± 1.362 | 6.042 ± 2.334 | 0.010 |
| 1 month | 1.458 ± 1.450 | 8.311 ± 2.508 | 0.006 | |
| 3 months | 4.465 ± 1.472 | 8.808 ± 2.393 | 0.070 | |
| ΔStrehl ratio | 1 week | 0.000 ± 0.009 | 0.029 ± 0.013 | 0.028 |
| 1 month | 0.001 ± 0.010 | 0.038 ± 0.015 | 0.012 | |
| 3 months | 0.014 ± 0.011 | 0.038 ± 0.015 | 0.098 | |
| ΔOV100% | 1 week | 0.011 ± 0.050 | 0.203 ± 0.080 | 0.017 |
| 1 month | 0.063 ± 0.049 | 0.281 ± 0.084 | 0.009 | |
| 3 months | 0.162 ± 0.049 | 0.288 ± 0.080 | 0.117 | |
| ΔOV20% | 1 week | −0.002 ± 0.045 | 0.143 ± 0.066 | 0.029 |
| 1 month | 0.038 ± 0.047 | 0.194 ± 0.071 | 0.028 | |
| 3 months | 0.094 ± 0.048 | 0.199 ± 0.066 | 0.115 | |
| ΔOV9% | 1 week | −0.012 ± 0.031 | 0.087 ± 0.042 | 0.018 |
| 1 month | −0.006 ± 0.032 | 0.113 ± 0.046 | 0.011 | |
| 3 months | 0.050 ± 0.034 | 0.111 ± 0.044 | 0.165 |
ICL implantable collamer lens, OSI objective scattering index, MTF modulation transfer function, OV optical quality analysis system value
Comparison of optical quality changes between Group H and S under dark conditions
| Parameter | Time | Group H | Group S | P-value |
|---|---|---|---|---|
| ΔOSI | 1 week | −0.325 ± 0.121 | − 1.606 ± 0.198 | 0.001 |
| 1 month | −0.379 ± 0.116 | −1.746 ± 0.212 | 0.001 | |
| 3 months | −0.429 ± 0.125 | − 1.727 ± 0.217 | 0.002 | |
| ΔMTF cutoff frequency | 1 week | 1.623 ± 1.587 | 7.443 ± 2.308 | 0.012 |
| 1 month | 1.347 ± 1.593 | 8.399 ± 2.476 | 0.004 | |
| 3 months | 2.994 ± 1.533 | 10.503 ± 2.274 | 0.001 | |
| ΔStrehl ratio | 1 week | 0.006 ± 0.008 | 0.036 ± 0.012 | 0.011 |
| 1 month | 0.006 ± 0.009 | 0.038 ± 0.013 | 0.011 | |
| 3 months | 0.016 ± 0.010 | 0.048 ± 0.013 | 0.016 | |
| ΔOV100% | 1 week | 0.058 ± 0.053 | 0.237 ± 0.077 | 0.020 |
| 1 month | 0.042 ± 0.053 | 0.275 ± 0.082 | 0.005 | |
| 3 months | 0.099 ± 0.051 | 0.335 ± 0.076 | 0.002 | |
| ΔOV20% | 1 week | 0.025 ± 0.043 | 0.169 ± 0.061 | 0.018 |
| 1 month | 0.052 ± 0.042 | 0.203 ± 0.063 | 0.016 | |
| 3 months | 0.089 ± 0.043 | 0.254 ± 0.060 | 0.006 | |
| ΔOV9% | 1 week | 0.015 ± 0.026 | 0.106 ± 0.037 | 0.013 |
| 1 month | 0.019 ± 0.027 | 0.110 ± 0.038 | 0.017 | |
| 3 months | 0.058 ± 0.034 | 0.137 ± 0.043 | 0.048 |
ICL implantable collamer lens, OSI objective scattering index, MTF modulation transfer function, OV optical quality analysis system value