Chen-Chen Zhao1,2, Guo-Xin Qu1,2, Shi-Gui Yan1,2, Xun-Zi Cai3,4.
Abstract
BACKGROUND: Postoperative squeaking in patients who applied the fourth-generation ceramic bearing in primary hip replacement has not been reported systematically; we aim to study the squeaking incidence in the fourth-generation ceramic bearing and related risk factors for squeaking, and we also attempt to explore the relationship between squeaking and prosthetic brands.Entities:
Keywords: Fourth-generation ceramics; Prosthesis brands; Squeaking; Total hip replacement
Mesh:
Year: 2018 PMID: 29859126 PMCID: PMC5984797 DOI: 10.1186/s13018-018-0841-y
Source DB: PubMed Journal: J Orthop Surg Res ISSN: 1749-799X Impact factor: 2.359
Studies included in the meta-analysis
| Author | Year | Country | Total hips | Patients with squeak | Acetabular cup | Femoral stem | Mean follow-up time (months) |
|---|---|---|---|---|---|---|---|
| Cai et al. | 2012 | China | 43 | 0 | NA | NA | 39.7 |
| McDonnell et al. | 2013 | Australia | 208 | 26 | Delta Motion (DePuy) | Finsbury Type C (Finsbury Orthopaedics)/SL Plus MIA stem (Smith & Nephew)/Tri-Lock and Corail (DePuy) | 21 |
| Wang et al. | 2014 | China | 177 | 2 | Pinnacle (DePuy) | Corail and Summit (DePuy) | 12 |
| Hamilton et al. | 2015 | Canada | 345 | 26 | Pinnacle (DePuy) | AML, Prodigy, Summit, Srom, and Corail (DePuy) | 64 |
| Aoude et al. | 2015 | Canada | 133 | 1 | Pinnacle (DePuy) | Corail, SROM, Summit, Trilock and Prodigy (DePuy) | 72 |
| Baek et al. | 2015 | Korea | 94 | 0 | Bencox (Corentec) | Bencox (Corentec) | 60 |
| Kim et al. | 2016 | Korea | 334 | 2 | Pinnacle (DePuy) | Proxima (DePuy) | 157.2 |
| Lee et al. | 2016 | Korea | 269 | 0 | Exceed eABT™ (Biomet) | Taperloc (Biomet) | 24 |
| Lim et al. | 2016 | Korea | 53 | 1 | Bencox (Corentec) | Bencox (Corentec) | 64 |
| Buttaro et al. | 2016 | Argentina | 939 | 1 | Pinnacle (DePuy)/Combi (Wlink)/Trinity (Corin) | Corail, SROM and Cstem (DePuy)/CFP (Wlink)/Minihip (Corin) | 64 |
| Boden et al. | 2017 | UK | 266 | 2 | Pinnacle (DePuy) | Corail (DePuy) | 18 |
| Salo et al. | 2017 | Finland | 336 | 37 | Continuum (Zimmer)/Exceed (Biomet)/Pinnacle (DePuy) | ML Taper (Zimmer)/Bimetric (Biomet)/Corail, Srom, and Summit (DePuy) | 25.6 |
| Goldhofer et al. | 2017 | Australia | 206 | 15 | Delta Motion (DePuy) | Secur-Fit stem (Stryker) | 60 |
| Lee et al. | 2017 | Korea | 286 | 7 | Pinnacle (DePuy) | Proxima (DePuy) | 66.5 |
Fig. 1Forest plots for squeaking incidence of all included studies. I2, τ2, and P value are tested for heterogeneity
Meta-regression analysis for the factors affecting the squeaking incidence
| Variables | Total hip | Study | 95% CI | |
|---|---|---|---|---|
| Age | 3689 | 14 | 0.498 | − 0.430 to 0.223 |
| Gender | 3689 | 14 | 0.727 | − 15.57 to 21.62 |
| BMI | 2145 | 10 | 0.464 | − 1.935 to 0.967 |
| Abduction | 1537 | 7 | 0.606 | − 0.889 to 1.415 |
| Anteversion | 1494 | 6 | 0.817 | − 1.309 to 1.584 |
BMI body mass index, CI confidence interval
Fig. 2Forest plots for squeaking incidence in less than 5-year and more than or equal to 5-year follow up groups. I2, τ2, and P value are tested for heterogeneity
Fig. 3Forest plots for squeaking incidence in each subgroup based on acetabular cup brands. I2, τ2, and P value are tested for heterogeneity
Fig. 4Forest plots for squeaking incidence in each subgroup based on femoral stem brands. I2, τ2, and P value are tested for heterogeneity