Literature DB >> 19768515

THA with Delta ceramic on ceramic: results of a multicenter investigational device exemption trial.

William G Hamilton1, James P McAuley, Douglas A Dennis, Jeffrey A Murphy, Thomas J Blumenfeld, Joel Politi.   

Abstract

UNLABELLED: Although the published studies on the outcomes of total hip arthroplasty (THA) performed with currently available ceramic components show high survivorship and low bearing wear at midterm followup, concern over ceramic fracture and squeaking persist. For these reasons, the use of ceramic is limited. Recently, a new alumina matrix composite material (Delta ceramic) with improved material properties was developed to address these concerns. We report the early outcomes and complications of a prospective, randomized, multicenter trial of 263 patients (264 hips) at eight centers, comparing a Delta ceramic-on-ceramic (COC) articulation with a Delta ceramic head-crosslinked polyethylene bearing combination (COP). There were 177 COC hips and 87 COP hips. Complications were reported for all patients, whereas clinical and radiographic results were provided for the 233 patients with minimum 2-year followup (average, 31.2 months; range, 21-49 months). The Harris hip scores and clinical, radiographic, and survivorship outcomes were similar in both groups. There were four (2%) revisions in the COC group and two (2%) in the COP group. We encountered three intraoperative ceramic liner-related events. In addition, one patient receiving the COC underwent revision for chipping of the ceramic liner, and a second had ceramic fragmentation on followup radiographs but has not undergone revision. These liner related complications remain a concern. No patient reported squeaking in either group; this leaves us hopeful the new material will lessen the frequency of squeaking. In the short term, the Delta COC articulation provided similar functional scores and survivorship and complication rates with the ceramic head mated with crosslinked polyethylene. LEVEL OF EVIDENCE: Level I, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19768515      PMCID: PMC2807004          DOI: 10.1007/s11999-009-1091-4

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  45 in total

1.  Incomplete seating of the liner with the Trident acetabular system: a cause for concern?

Authors:  A J Langdown; R J Pickard; C M Hobbs; H J Clarke; D J N Dalton; M L Grover
Journal:  J Bone Joint Surg Br       Date:  2007-03

2.  Mid-term results of cementless total hip replacement using a ceramic-on-ceramic bearing with and without computer navigation.

Authors:  N Sugano; T Nishii; H Miki; H Yoshikawa; Y Sato; S Tamura
Journal:  J Bone Joint Surg Br       Date:  2007-04

3.  Early failure of a cross-linked polyethylene acetabular liner. A case report.

Authors:  K David Moore; Preston R Beck; Donald W Petersen; John M Cuckler; Jack E Lemons; Alan W Eberhardt
Journal:  J Bone Joint Surg Am       Date:  2008-11       Impact factor: 5.284

4.  Failure analysis of a ceramic bearing acetabular component.

Authors:  Robert A Poggie; Thomas R Turgeon; Richard D Coutts
Journal:  J Bone Joint Surg Am       Date:  2007-02       Impact factor: 5.284

5.  Ceramic liner fracture after cementless alumina-on-alumina total hip arthroplasty.

Authors:  Yong-Chan Ha; Shin-Yoon Kim; Hee Joong Kim; Jeong Joon Yoo; Kyung-Hoi Koo
Journal:  Clin Orthop Relat Res       Date:  2007-05       Impact factor: 4.176

6.  Alumina femoral head fracture in uncemented total hip arthroplasty with a ceramic sandwich cup.

Authors:  Deuk-Soo Hwang; Young-Mo Kim; Chang-Hwan Lee
Journal:  J Arthroplasty       Date:  2007-04       Impact factor: 4.757

7.  Two- to 9-year clinical results of alumina ceramic-on-ceramic THA.

Authors:  Stephen B Murphy; Timo M Ecker; Moritz Tannast
Journal:  Clin Orthop Relat Res       Date:  2006-12       Impact factor: 4.176

8.  Delayed catastrophic failure of a ceramic head in hybrid total hip arthroplasty.

Authors:  Imran Anwar; Gauraang Bhatnagar; Salah Atrah
Journal:  J Arthroplasty       Date:  2008-06-05       Impact factor: 4.757

9.  Squeaking in ceramic-on-ceramic hips: the importance of acetabular component orientation.

Authors:  William L Walter; Gary C O'toole; William K Walter; Andrew Ellis; Bernard A Zicat
Journal:  J Arthroplasty       Date:  2007-03-28       Impact factor: 4.757

10.  Ceramic-on-ceramic total hip arthroplasty: update.

Authors:  William N Capello; James A D'Antonio; Judy R Feinberg; Michael T Manley; Marybeth Naughton
Journal:  J Arthroplasty       Date:  2008-10       Impact factor: 4.757

View more
  42 in total

1.  Causes of failure of ceramic-on-ceramic and metal-on-metal hip arthroplasties.

Authors:  Manny Porat; Javad Parvizi; Peter F Sharkey; Keith R Berend; Adolph V Lombardi; Robert L Barrack
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

Review 2.  Survival of hard-on-hard bearings in total hip arthroplasty: a systematic review.

Authors:  Michael G Zywiel; Siraj A Sayeed; Aaron J Johnson; Thomas P Schmalzried; Michael A Mont
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

3.  Preventing ceramic liner fracture after Delta ceramic-on-ceramic total hip arthroplasty.

Authors:  Young-Kyun Lee; Jae-Young Lim; Yong-Chan Ha; Tae-Young Kim; Woon-Hwa Jung; Kyung-Hoi Koo
Journal:  Arch Orthop Trauma Surg       Date:  2020-06-12       Impact factor: 3.067

4.  Ideal femoral head size in total hip arthroplasty balances stability and volumetric wear.

Authors:  Michael B Cross; Denis Nam; David J Mayman
Journal:  HSS J       Date:  2012-09-13

5.  Hard-on-Hard Bearings Are Associated With Increased Noise Generation in Young Patients Undergoing Hip Arthroplasty.

Authors:  Denis Nam; Toby Barrack; Staci R Johnson; Ryan M Nunley; Robert L Barrack
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

Review 6.  [Evidence-based update in hip arthroplasty].

Authors:  H Gollwitzer; L Gerdesmeyer; R Gradinger; R von Eisenhart-Rothe
Journal:  Orthopade       Date:  2011-06       Impact factor: 1.087

7.  Comparative effectiveness of ceramic-on-ceramic implants in stemmed hip replacement: a multinational study of six national and regional registries.

Authors:  Art Sedrakyan; Stephen Graves; Barbara Bordini; Miquel Pons; Leif Havelin; Susan Mehle; Elizabeth Paxton; Thomas Barber; Guy Cafri
Journal:  J Bone Joint Surg Am       Date:  2014-12-17       Impact factor: 5.284

8.  CORR Insights(®): do alumina matrix composite bearings decrease hip noises and bearing fractures at a minimum of 5 years after THA?

Authors:  Laurent Sedel
Journal:  Clin Orthop Relat Res       Date:  2015-07-23       Impact factor: 4.176

9.  Squeaking and microcracks in a delta-delta ceramic coupling: pin-on-disc study.

Authors:  Kiyokazu Fukui; Ayumi Kaneuji; Tadami Matsumoto; Kazuhiro Shintani
Journal:  Eur J Orthop Surg Traumatol       Date:  2016-03-12

10.  Alumina-on-Polyethylene Bearing Surfaces in Total Hip Arthroplasty.

Authors:  Yup Lee Jung; Shin-Yoon Kim
Journal:  Open Orthop J       Date:  2010-02-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.