BACKGROUND: Delta ceramics may be the bearing of choice for younger and active patients due to its improved toughness and wear characteristics, provided there is no risk of fracture. However, ceramic fracture is the most serious complication related to this type of bearing. Although millions of Delta ceramics have been implanted worldwide, short to midterm results have been scarcely reported in the literature. The purpose of this study was to report the complication rate at short to midterm follow-up associated with the bearing surface used in a series of primary total hip arthroplasties with Delta ceramic-on-ceramic bearings performed in a single institution.
METHODS: A total of 939 cases (880 patients) undergoing primary total hip arthroplasty with fourth-generation Delta ceramic-on-ceramic bearings were retrospectively reviewed. They were followed for an average of 5.3 years (2-10 years).
RESULTS: One hip experienced a liner fracture, 2 cups presented early loosening due to friction between the acetabular screw and the backside of the liner, one femoral ball head had a fracture; one case of squeaking was reported, which is impending revision. Considering revision or impending revision in relationship with the bearing surface as the end point, the mean survival rate was 99.3% (confidence interval 95%, 98.3%-99.7%) at 2-10 years.
CONCLUSION: This study showed a low rate of ceramic fracture compared with others; however, it was much higher than the complication rate presented by the manufacturers. The complications observed were directly related to technical errors that surgeons should avoid when using this type of surface.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  ceramic fracture; ceramic-on-ceramic; complications; primary THA; squeaking

Mesh:

Year:  2016        PMID: 27618155     DOI: 10.1016/j.arth.2016.07.032

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  9 in total

1.  Third generation delta ceramic-on-ceramic bearing for total hip arthroplasty at mid-term follow-up.

Authors:  Alessandro Bistolfi; Riccardo Ferracini; Alessandro Aprato; Alessandro Massè; Walter Daghino; Sara Lea; Stefano Artiaco; Gwo-Chin Lee
Journal:  J Orthop       Date:  2020-08-28

2.  [Research progress of total hip arthroplasty for patients with sequelae of suppurative hip arthritis].

Authors:  Yongzhi Qin; Zongke Zhou
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2020-01-15

3.  Total Hip Arthroplasty Using Ceramic-on-ceramic Bearing Surfaces: Long-term Assessment of Squeaking Sounds.

Authors:  Min Wook Kim; Sang Min Kim; Young Yool Chung
Journal:  Hip Pelvis       Date:  2018-03-05

4.  Cobalt Intoxication Heart Failure after Revision Total Hip Replacement for Ceramic Head Fracture: A Case Report.

Authors:  Chul-Ho Kim; Young Hyun Choi; Mi Yeon Jeong; Jae Suk Chang; Pil Whan Yoon
Journal:  Hip Pelvis       Date:  2016-12-28

5.  Bearing surfaces in primary total hip arthroplasty.

Authors:  Luigi Zagra; Enrico Gallazzi
Journal:  EFORT Open Rev       Date:  2018-05-21

6.  Squeaking in fourth-generation ceramic-on-ceramic total hip replacement and the relationship with prosthesis brands: meta-analysis and systematic review.

Authors:  Chen-Chen Zhao; Guo-Xin Qu; Shi-Gui Yan; Xun-Zi Cai
Journal:  J Orthop Surg Res       Date:  2018-06-01       Impact factor: 2.359

7.  Total Hip Arthroplasty Using Hybrid Ceramic Bearing: A Minimum 10-Year Followup Study.

Authors:  Jun-Il Yoo; Yong-Chan Ha; Dong-Hyun Kim; Young-Kyun Lee; Kyung-Hoi Koo
Journal:  Indian J Orthop       Date:  2019 Sep-Oct       Impact factor: 1.251

8.  Mid-term outcomes of the R3™ delta ceramic acetabular system in total hip arthroplasty.

Authors:  Edward T Davis; Ville Remes; Petri Virolainen; Peter Gebuhr; Bart Van Backlé; Matthew P Revell; Branko Kopjar
Journal:  J Orthop Surg Res       Date:  2021-01-09       Impact factor: 2.359

9.  Long-Term Results of Delta Ceramic-on-Ceramic Total Hip Arthroplasty.

Authors:  Thomas J Blumenfeld; Joel Politi; Sean Coker; Tammy O'Dell; William Hamilton
Journal:  Arthroplast Today       Date:  2022-01-18
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.