Is metal-on-metal squeaking related to acetabular angle of inclination?

Thomas Bernasek1, David Fisher, David Dalury, Melissa Levering, Kirk Dimitris.   

Abstract

BACKGROUND: Postoperative audible squeaking has been well documented in ceramic-on-ceramic hip prostheses, and several metal-on-metal (MOM) THA designs, specifically those used for large-head resurfacing and MOM polyethylene sandwich designs, and are attributed to different implant- and patient-specific factors. Current literature does not identify the incidence of squeaking in modular MOM THA or possible etiologic factors. QUESTIONS/PURPOSES: Our purposes were to (1) identify the incidence of squeaking in modular MOM prostheses in THA; (2) determine whether males or females were more likely to have squeaking; and (3) determine whether the incidence of squeaking relates to acetabular inclination angle.
METHODS: We retrospectively reviewed the patient records and radiographs of 539 patients (542 hips) from three independent centers who underwent a MOM THA between February 2001 and December 2005. Demographic and implant factors were evaluated, including measurement of cup inclination angles. The minimum followup was 36 months (mean, 76 months; range, 36-119 months).
RESULTS: We identified squeaking in eight of the 542 hips (1.5%); five were in women and two were in men (one patient had bilateral squeaking). The time to onset of patient-reported audible squeaking averaged 23 months (range, 6-84 months). Squeaking was more likely to occur in women (six of eight hips). No hips with 45º or less acetabular inclination squeaked (291 hips); eight of 251 hips (3.2%) with inclination angles greater than 45º squeaked. Patients who reported squeaking had higher inclination angles than those who did not report squeaking.
CONCLUSIONS: Our observations suggest an increased frequency of squeaking in female patients and in patients with greater inclination angles with this MOM implant design.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21512812      PMCID: PMC3148366          DOI: 10.1007/s11999-011-1900-4

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  15 in total

1.  The influence of clearance on friction, lubrication and squeaking in large diameter metal-on-metal hip replacements.

Authors:  Claire L Brockett; Phil Harper; Sophie Williams; Graham H Isaac; Rob S Dwyer-Joyce; Zhongmin Jin; John Fisher
Journal:  J Mater Sci Mater Med       Date:  2007-11-08       Impact factor: 3.896

2.  In vivo comparison of hip mechanics for minimally invasive versus traditional total hip arthroplasty.

Authors:  Diana Glaser; Douglas A Dennis; Richard D Komistek; Todd M Miner
Journal:  Clin Biomech (Bristol, Avon)       Date:  2007-11-26       Impact factor: 2.063

3.  The role of stripe wear in causing acoustic emissions from alumina ceramic-on-ceramic bearings.

Authors:  Scott Taylor; Michael T Manley; Kate Sutton
Journal:  J Arthroplasty       Date:  2007-10       Impact factor: 4.757

Review 4.  The squeaking hip: a cause for concern-agrees.

Authors:  Amar S Ranawat; Chitranjan S Ranawat
Journal:  Orthopedics       Date:  2007-09       Impact factor: 1.390

5.  High incidence of squeaking in THAs with alumina ceramic-on-ceramic bearings.

Authors:  J C Keurentjes; R M Kuipers; D J Wever; B W Schreurs
Journal:  Clin Orthop Relat Res       Date:  2008-02-26       Impact factor: 4.176

6.  Early results of primary Birmingham hip resurfacings. An independent prospective study of the first 230 hips.

Authors:  D L Back; R Dalziel; D Young; A Shimmin
Journal:  J Bone Joint Surg Br       Date:  2005-03

7.  Osteolysis with a cementless second generation metal-on-metal cup in total hip replacement.

Authors:  Ian Holloway; William L Walter; Bernie Zicat; William K Walter
Journal:  Int Orthop       Date:  2008-11-05       Impact factor: 3.075

8.  Dislocations after total hip arthroplasty.

Authors:  R Y Woo; B F Morrey
Journal:  J Bone Joint Surg Am       Date:  1982-12       Impact factor: 5.284

9.  How do alternative bearing surfaces influence wear behavior?

Authors:  Ian C Clarke; Michael T Manley
Journal:  J Am Acad Orthop Surg       Date:  2008       Impact factor: 3.020

10.  Squeaking in ceramic-on-ceramic hips: the importance of acetabular component orientation.

Authors:  William L Walter; Gary C O'toole; William K Walter; Andrew Ellis; Bernard A Zicat
Journal:  J Arthroplasty       Date:  2007-03-28       Impact factor: 4.757

View more
  9 in total

1.  Hard-on-Hard Bearings Are Associated With Increased Noise Generation in Young Patients Undergoing Hip Arthroplasty.

Authors:  Denis Nam; Toby Barrack; Staci R Johnson; Ryan M Nunley; Robert L Barrack
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

2.  Does standing affect acetabular component inclination and version after THA?

Authors:  Gregory G Polkowski; Ryan M Nunley; Erin L Ruh; Brandon M Williams; Robert L Barrack
Journal:  Clin Orthop Relat Res       Date:  2012-11       Impact factor: 4.176

Review 3.  Review on squeaking hips.

Authors:  Yadin David Levy; Selin Munir; Shane Donohoo; William Lindsay Walter
Journal:  World J Orthop       Date:  2015-11-18

Review 4.  Metal-on-Metal Bearing: Is This the End of the Line? We Do Not Think So.

Authors:  Henri Migaud; Sophie Putman; Antoine Combes; Charles Berton; Donatien Bocquet; Laurent Vasseur; Julien Girard
Journal:  HSS J       Date:  2012-09-11

Review 5.  Ceramic-Ceramic Bearing: Too Unpredictable to Use it Regularly.

Authors:  Edwin P Su
Journal:  HSS J       Date:  2012-09-05

6.  Changes in Intra-pelvic Obliquity Angle 0-2 Years After Total Hip Arthroplasty and Its Effects on Leg Length Discrepancy: A Retrospective Study.

Authors:  Yin Zhang; Tao Cheng; Xian-Long Zhang
Journal:  Chin Med J (Engl)       Date:  2015-05-20       Impact factor: 2.628

7.  Squeaking in fourth-generation ceramic-on-ceramic total hip replacement and the relationship with prosthesis brands: meta-analysis and systematic review.

Authors:  Chen-Chen Zhao; Guo-Xin Qu; Shi-Gui Yan; Xun-Zi Cai
Journal:  J Orthop Surg Res       Date:  2018-06-01       Impact factor: 2.359

Review 8.  Hip Squeaking after Ceramic-on-ceramic Total Hip Arthroplasty.

Authors:  Guo-Liang Wu; Wei Zhu; Yan Zhao; Qi Ma; Xi-Sheng Weng
Journal:  Chin Med J (Engl)       Date:  2016-08-05       Impact factor: 2.628

9.  Do different types of bearings and noise from total hip arthroplasty influence hip-related pain, function, and quality of life postoperatively?

Authors:  Claus Varnum; Alma B Pedersen; Per Kjærsgaard-Andersen; Søren Overgaard
Journal:  Acta Orthop       Date:  2016-09-12       Impact factor: 3.717

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.