Literature DB >> 21274762

A squeaky reputation: the problem may be design-dependent.

Javad Parvizi1, Bahar Adeli, Justin C Wong, Camilo Restrepo, Richard H Rothman.   

Abstract

BACKGROUND: Squeaking is reportedly a complication in patients having ceramic-on-ceramic total hip implants. The etiology remains unknown and multifactorial with recent studies suggesting a relationship between the audible squeak and implant design. When we evaluated our ceramic-on-ceramic cohort, we noticed squeaking primarily in patients receiving an acetabular system designed with an elevated titanium rim. OBJECTIVES/PURPOSES: We therefore (1) determined the incidence of squeaking among four different ceramic-on-ceramic bearing surfaces used for THA at our institution; (2) evaluated the association between different acetabular designs and the incidence of squeaking; and (3) assessed other potential variables associated with squeaking.
METHODS: We retrospectively reviewed 1507 patients having a ceramic-on-ceramic THA between 2002 and 2009; we separately analyzed those receiving an acetabular system with and without an elevated titanium rim. Data were collected through phone-administered questionnaires and retrospective reviews of patient charts for intraoperative findings, followup reports, demographic information, and radiographic findings.
RESULTS: Squeaking occurred in 92 of the 1507 patients (6%). All 92 patients with squeaking received an elevated rim design (1291 patients) or an incidence of 7% with that design. We found no association between squeaking and any other examined factors.
CONCLUSION: Our findings complement the theory from in vitro studies suggesting that neck impingement on the elevated titanium rim is the probable cause of the increased frequency of squeaking with this design.

Entities:  

Mesh:

Year:  2011        PMID: 21274762      PMCID: PMC3094607          DOI: 10.1007/s11999-011-1777-2

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  48 in total

1.  Total hip arthroplasty using a ceramic prosthesis. Pierre Boutin (1924-1989).

Authors:  P Boutin
Journal:  Clin Orthop Relat Res       Date:  2000-10       Impact factor: 4.176

2.  High incidence of squeaking in THAs with alumina ceramic-on-ceramic bearings.

Authors:  J C Keurentjes; R M Kuipers; D J Wever; B W Schreurs
Journal:  Clin Orthop Relat Res       Date:  2008-02-26       Impact factor: 4.176

Review 3.  Bearings of the future for total hip arthroplasty.

Authors:  Michael T Manley; Kate Sutton
Journal:  J Arthroplasty       Date:  2008-08-12       Impact factor: 4.757

4.  Squeaking hips.

Authors:  William L Walter; Tim S Waters; Mark Gillies; Shane Donohoo; Steven M Kurtz; Amar S Ranawat; William J Hozack; Michael A Tuke
Journal:  J Bone Joint Surg Am       Date:  2008-11       Impact factor: 5.284

5.  Clicking and squeaking: in vivo correlation of sound and separation for different bearing surfaces.

Authors:  Diana Glaser; Richard D Komistek; Harold E Cates; Mohamed R Mahfouz
Journal:  J Bone Joint Surg Am       Date:  2008-11       Impact factor: 5.284

6.  Squeaking in total hip replacement: a cause for concern.

Authors:  José A Rodríguez; Alejandro Gonzalez DelaValle; Nicole McCook
Journal:  Orthopedics       Date:  2008-09       Impact factor: 1.390

7.  Revision hip arthroplasty for ceramic-on-ceramic squeaking hips does not compromise the results.

Authors:  Wadih Y Matar; Camilo Restrepo; Javad Parvizi; Steven M Kurtz; William J Hozack
Journal:  J Arthroplasty       Date:  2010-07-15       Impact factor: 4.757

8.  Ceramic-on-ceramic total hip arthroplasty.

Authors:  Joseph W Greene; Arthur L Malkani; Frank R Kolisek; Nenette M Jessup; Dale L Baker
Journal:  J Arthroplasty       Date:  2009-07-04       Impact factor: 4.757

9.  Ceramic-on-ceramic total hip arthroplasty: update.

Authors:  William N Capello; James A D'Antonio; Judy R Feinberg; Michael T Manley; Marybeth Naughton
Journal:  J Arthroplasty       Date:  2008-10       Impact factor: 4.757

10.  The noisy ceramic hip: is component malpositioning the cause?

Authors:  Camilo Restrepo; Javad Parvizi; Steve M Kurtz; Peter F Sharkey; William J Hozack; Richard H Rothman
Journal:  J Arthroplasty       Date:  2008-08       Impact factor: 4.757

View more
  16 in total

1.  Squeaking: Current knowledge and how to avoid it.

Authors:  Arjuna M Imbuldeniya; Simon J Pearce; William L Walter; Bernard A Zicat; William K Walter
Journal:  Curr Rev Musculoskelet Med       Date:  2013-12

2.  Clinical faceoff: Ceramic-on-ceramic THA: do the advantages outweigh the limitations?

Authors:  Laurent Sedel; William L Walter; Rocco P Pitto
Journal:  Clin Orthop Relat Res       Date:  2014-05-13       Impact factor: 4.176

3.  Meta-analysis of the Incidence and Risk Factors for Squeaking after Primary Ceramic-on-ceramic Total Hip Arthroplasty in Asian Patients.

Authors:  Tae-Hun Lee; Young-Wan Moon; Seung-Jae Lim; Youn-Soo Park
Journal:  Hip Pelvis       Date:  2014-06-30

4.  [Which hip articulation bearing for which patient? : Tribology of the future].

Authors:  M M Morlock; N Bishop; C Kaddick
Journal:  Orthopade       Date:  2011-12       Impact factor: 1.087

Review 5.  Review on squeaking hips.

Authors:  Yadin David Levy; Selin Munir; Shane Donohoo; William Lindsay Walter
Journal:  World J Orthop       Date:  2015-11-18

Review 6.  Ceramic-Ceramic Bearing: Too Unpredictable to Use it Regularly.

Authors:  Edwin P Su
Journal:  HSS J       Date:  2012-09-05

7.  Noise from total hip replacements: a case-controlled study.

Authors:  M C Wyatt; S Jesani; C Frampton; P Devane; J G Horne
Journal:  Bone Joint Res       Date:  2014-06       Impact factor: 5.853

8.  Extraction of a well fixed but fractured ceramic acetabular liner.

Authors:  David Ferguson; Robert Metcalf
Journal:  Arthroplast Today       Date:  2015-03-20

9.  Squeaking Is Common and Increases Over Time Among Patients With Long-term Follow-up After Ceramic-on-ceramic THA.

Authors:  Kevin Taniguchi; Michael Quacinella; Brian Barlow
Journal:  Clin Orthop Relat Res       Date:  2021-04-01       Impact factor: 4.176

Review 10.  Hip Squeaking after Ceramic-on-ceramic Total Hip Arthroplasty.

Authors:  Guo-Liang Wu; Wei Zhu; Yan Zhao; Qi Ma; Xi-Sheng Weng
Journal:  Chin Med J (Engl)       Date:  2016-08-05       Impact factor: 2.628

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.