Two-dimensional antiscatter grid: A novel scatter rejection device for Cone-beam computed tomography.

Timur Alexeev1, Brian Kavanagh1, Moyed Miften1, Cem Altunbas1.   

Abstract

PURPOSE: Scattered radiation remains to be a major cause of image quality degradation in Flat Panel Detector (FPD)-based Cone-beam computed tomography (CBCT). We have been investigating a novel two-dimensional antiscatter grid (2D-ASG) concept to reduce scatter intensity, and hence improve CBCT image quality. We present the first CBCT imaging experiments performed with the 2D-ASG prototype, and demonstrate its efficacy in improving CBCT image quality.
METHODS: A 2D-ASG prototype with septa focused to x-ray source was additively manufactured from tungsten and mounted on a Varian TrueBeam CBCT system. CBCT projections of phantoms were acquired with an offset detector geometry using TrueBeam's "developer" mode. To minimize the effect of gantry flex, projections were gain corrected on angle-specific bases. CBCT images were reconstructed using a filtered backprojection algorithm and image quality improvement was quantified by measuring contrast-to-noise ratio (CNR) and CT number accuracy in images acquired with no antiscatter grid (NO-ASG), conventional one dimensional antiscatter grid (1D-ASG), and the 2D-ASG prototype.
RESULTS: A significant improvement in contrast resolution was achieved using our 2D-ASG prototype compared to results of 1D-ASG and NO-ASG acquisitions. Compared to NO-ASG and 1D-ASG experiments, the CNR of material inserts improved by as much as 86% and 54% respectively. Using 2D-ASG, CT number underestimation in water equivalent material section of the phantom was reduced by up to 325 HU when compared to NO-ASG and up to 179 HU when compared to 1D-ASG.
CONCLUSION: We successfully performed the first CBCT imaging experiments with a 2D-ASG prototype. 2D-ASG provided significantly higher CT number accuracy, higher CNR, and diminished scatter-induced image artifacts in qualitative evaluations. We strongly believe that utilization of a 2D-ASG may potentially lead to better soft tissue visualization in CBCT and may enable novel clinical applications that require high CT number accuracy.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  CBCT; anti-scatter; grid; two-dimensional

Mesh:

Year:  2018        PMID: 29235120      PMCID: PMC5807157          DOI: 10.1002/mp.12724

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  10 in total

1.  Cone-beam computed tomography with a flat-panel imager: magnitude and effects of x-ray scatter.

Authors:  J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  2001-02       Impact factor: 4.071

2.  X-ray micro-CT with a displaced detector array.

Authors:  Ge Wang
Journal:  Med Phys       Date:  2002-07       Impact factor: 4.071

3.  Monte Carlo evaluation of scatter mitigation strategies in cone-beam CT.

Authors:  Dimitrios Lazos; Jeffrey F Williamson
Journal:  Med Phys       Date:  2010-10       Impact factor: 4.071

4.  Efficiency of antiscatter grids for flat-detector CT.

Authors:  Yiannis Kyriakou; Willi Kalender
Journal:  Phys Med Biol       Date:  2007-10-02       Impact factor: 3.609

5.  Cone-beam CT from width-truncated projections.

Authors:  P S Cho; A D Rudd; R H Johnson
Journal:  Comput Med Imaging Graph       Date:  1996 Jan-Feb       Impact factor: 4.790

6.  Monte Carlo study of the effects of system geometry and antiscatter grids on cone-beam CT scatter distributions.

Authors:  A Sisniega; W Zbijewski; A Badal; I S Kyprianou; J W Stayman; J J Vaquero; J H Siewerdsen
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

7.  The effects of scatter in x-ray computed tomography.

Authors:  P M Joseph; R D Spital
Journal:  Med Phys       Date:  1982 Jul-Aug       Impact factor: 4.071

8.  Scattered radiation in fan beam imaging systems.

Authors:  P C Johns; M Yaffe
Journal:  Med Phys       Date:  1982 Mar-Apr       Impact factor: 4.071

9.  Compton scatter effects in CT reconstructions.

Authors:  G H Glover
Journal:  Med Phys       Date:  1982 Nov-Dec       Impact factor: 4.071

10.  Transmission characteristics of a two dimensional antiscatter grid prototype for CBCT.

Authors:  Cem Altunbas; Brian Kavanagh; Timur Alexeev; Moyed Miften
Journal:  Med Phys       Date:  2017-06-16       Impact factor: 4.071

  10 in total
  8 in total

1.  A novel total variation based ring artifact suppression method for CBCT imaging with two-dimensional antiscatter grids.

Authors:  Timur Alexeev; Brian Kavanagh; Moyed Miften; Cem Altunbas
Journal:  Med Phys       Date:  2019-03-20       Impact factor: 4.071

2.  Simultaneous scatter rejection and correction method using 2D antiscatter grids for CBCT.

Authors:  Zhelin Yu; Yeonok Park; Cem Altunbas
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2020-03-16

3.  Concurrent kilovoltage CBCT imaging and megavoltage beam delivery: suppression of cross-scatter with 2D antiscatter grids and grid-based scatter sampling.

Authors:  Farhang Bayat; Mohamed Elsayed Eldib; Brian Kavanagh; Moyed Miften; Cem Altunbas
Journal:  Phys Med Biol       Date:  2022-08-09       Impact factor: 4.174

4.  Megavoltage cross-scatter rejection and correction using 2D antiscatter grids in kilovoltage CBCT imaging.

Authors:  Farhang Bayat; Mohamed Elsayed Eldib; Cem Altunbas
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2022-04-04

5.  Effect of grid geometry on the transmission properties of 2D grids for flat detectors in CBCT.

Authors:  Cem Altunbas; Timur Alexeev; Moyed Miften; Brian Kavanagh
Journal:  Phys Med Biol       Date:  2019-11-15       Impact factor: 3.609

6.  Evaluation of scatter rejection and correction performance of 2D antiscatter grids in cone beam computed tomography.

Authors:  Yeonok Park; Timur Alexeev; Brian Miller; Moyed Miften; Cem Altunbas
Journal:  Med Phys       Date:  2021-03-04       Impact factor: 4.071

7.  A unified scatter rejection and correction method for cone beam computed tomography.

Authors:  Cem Altunbas; Yeonok Park; Zhelin Yu; Anant Gopal
Journal:  Med Phys       Date:  2021-02-06       Impact factor: 4.071

8.  Determination of intra-oral surface areas by cone-beam computed tomography analysis and their relation with anthrometric measurements of the head.

Authors:  Z Assy; C Klop; H S Brand; R C Hoogeveen; J H Koolstra; F J Bikker
Journal:  Surg Radiol Anat       Date:  2020-07-11       Impact factor: 1.246

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.