Literature DB >> 33378551

A unified scatter rejection and correction method for cone beam computed tomography.

Cem Altunbas1, Yeonok Park1, Zhelin Yu2, Anant Gopal3.   

Abstract

PURPOSE: Scattered radiation is a major cause of image quality degradation in flat panel detector-based cone beam CT (CBCT). While recently introduced 2D antiscatter grids reject the majority of scatter fluence, the small percentage of scatter fluence still transmitted to the detector remains a major challenge for implementation of quantitative imaging techniques such as dual energy imaging in CBCT. Additionally, this residual scatter is also a major source of grid-induced artifacts, which impedes implementation of 2D grids in CBCT. We therefore present a new method to achieve both robust scatter rejection and residual scatter correction using a 2D antiscatter grid; in doing so, we expand the role of 2D grids from mere scatter rejection devices to scatter measurement devices.
METHOD: In our method, the radiopaque septa of the 2D grid emulate a micro array of beam-stops placed on the detector which introduce spatially periodic septal shadows. By selecting sufficiently thin grid septa, the primary intensity can be reduced while preserving the uniformity of scatter intensity. This enables us to correlate the modulated pixel signal intensity in septal shadows with local scatter intensity. Our method then exploits this correlation to measure and remove residual scatter intensity from projections. No assumptions are made about the object being imaged. We refer to this as Grid-based Scatter Sampling (GSS). In this work, we evaluate the principle of signal modulation with grid septa, the accuracy of scatter estimates, and the effect of the GSS method on image quality using simulations and measurements. We also implement the GSS method experimentally using a 2D grid prototype.
RESULTS: Our results demonstrate that the GSS method increased CT number accuracy and reduced image artifacts associated with scatter. With 2D grid and residual scatter correction, HU nonuniformity was reduced from 65 HU to 30 HU in pelvis sized phantoms, and HU variations due to change in phantom size were reduced from 59 HU to 20 HU, when compared to use of only a 2D grid. With residual scatter correction via GSS method, grid-induced ring artifacts were suppressed, leading to a 41% reduction in noise. The shape of the modulation transfer function (MTF) was preserved before and after suppression of ring artifacts.
CONCLUSIONS: Our grid-based scatter sampling method enables utilization of a 2D grid as a scatter measurement and correction device. This method significantly improves quantitative accuracy in CBCT, further reducing the image quality gap between CBCT and multi-detector CT. By correcting residual scatter with the proposed method, grid-induced line artifacts in projections and associated ring artifacts in CBCT images were also suppressed with no compromise of spatial resolution.
© 2020 American Association of Physicists in Medicine.

Entities:  

Keywords:  Antiscatter grid; flat panel detectors; quantitative CBCT; scatter correction

Mesh:

Year:  2021        PMID: 33378551      PMCID: PMC7965329          DOI: 10.1002/mp.14681

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  27 in total

1.  A ghost story: spatio-temporal response characteristics of an indirect-detection flat-panel imager.

Authors:  J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  1999-08       Impact factor: 4.071

2.  X-ray micro-CT with a displaced detector array.

Authors:  Ge Wang
Journal:  Med Phys       Date:  2002-07       Impact factor: 4.071

3.  Scatter correction method for X-ray CT using primary modulation: theory and preliminary results.

Authors:  Lei Zhu; N Robert Bennett; Rebecca Fahrig
Journal:  IEEE Trans Med Imaging       Date:  2006-12       Impact factor: 10.048

4.  Physical evaluation of prototype high-performance anti-scatter grids: potential for improved digital radiographic image quality.

Authors:  Kenneth A Fetterly; Beth A Schueler
Journal:  Phys Med Biol       Date:  2008-12-19       Impact factor: 3.609

5.  Feasibility study of a synchronized-moving-grid (SMOG) system to improve image quality in cone-beam computed tomography (CBCT).

Authors:  Lei Ren; Fang-Fang Yin; Indrin J Chetty; David A Jaffray; Jian-Yue Jin
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

6.  An Accurate Scatter Measurement and Correction Technique for Cone Beam Breast CT Imaging Using Scanning Sampled Measurement (SSM) Technique.

Authors:  Xinming Liu; Chris C Shaw; Tianpeng Wang; Lingyun Chen; Mustafa C Altunbas; S Cheenu Kappadath
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2006-02-28

7.  A simple approach to measure computed tomography (CT) modulation transfer function (MTF) and noise-power spectrum (NPS) using the American College of Radiology (ACR) accreditation phantom.

Authors:  Saul N Friedman; George S K Fung; Jeffrey H Siewerdsen; Benjamin M W Tsui
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

8.  Dual-energy cone-beam CT with a flat-panel detector: effect of reconstruction algorithm on material classification.

Authors:  W Zbijewski; G J Gang; J Xu; A S Wang; J W Stayman; K Taguchi; J A Carrino; J H Siewerdsen
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

9.  Digital radiographic imaging system with multiple-slit scanning x-ray beam: preliminary report.

Authors:  K Doi; H Fujita; K Ohara; K Ono; H Matsui; M L Giger; H P Chan
Journal:  Radiology       Date:  1986-11       Impact factor: 11.105

10.  Acuros CTS: A fast, linear Boltzmann transport equation solver for computed tomography scatter - Part II: System modeling, scatter correction, and optimization.

Authors:  Adam Wang; Alexander Maslowski; Philippe Messmer; Mathias Lehmann; Adam Strzelecki; Elaine Yu; Pascal Paysan; Marcus Brehm; Peter Munro; Josh Star-Lack; Dieter Seghers
Journal:  Med Phys       Date:  2018-03-23       Impact factor: 4.071

View more
  3 in total

1.  Concurrent kilovoltage CBCT imaging and megavoltage beam delivery: suppression of cross-scatter with 2D antiscatter grids and grid-based scatter sampling.

Authors:  Farhang Bayat; Mohamed Elsayed Eldib; Brian Kavanagh; Moyed Miften; Cem Altunbas
Journal:  Phys Med Biol       Date:  2022-08-09       Impact factor: 4.174

2.  Megavoltage cross-scatter rejection and correction using 2D antiscatter grids in kilovoltage CBCT imaging.

Authors:  Farhang Bayat; Mohamed Elsayed Eldib; Cem Altunbas
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2022-04-04

3.  Evaluation of scatter rejection and correction performance of 2D antiscatter grids in cone beam computed tomography.

Authors:  Yeonok Park; Timur Alexeev; Brian Miller; Moyed Miften; Cem Altunbas
Journal:  Med Phys       Date:  2021-03-04       Impact factor: 4.071

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.