A novel total variation based ring artifact suppression method for CBCT imaging with two-dimensional antiscatter grids.

Timur Alexeev1, Brian Kavanagh1, Moyed Miften1, Cem Altunbas1.   

Abstract

PURPOSE: Two-dimensional antiscatter grids (2DASG) for cone beam computed tomography (CBCT) is a new area of research to reduce scatter intensity, and consequently improve CBCT image quality. One of the challenges in implementation of 2DASGs is their septa shadows that are impinged on the projections. If these artifacts are not corrected, they may lead to ring artifacts in CBCT images. In this work, we present a novel method to suppress ring artifacts in FPD-based CBCT images.
METHODS: Briefly, our method first detects the locations of 2DASG's septa shadows in projections and then, reduces projection pixel values in septa shadows iteratively until a residual-based convergence criterion is met. To suppress the 2DASG's septa shadows, we developed a total variation minimization (TVM) formulation, referred to as adaptive-diffusive total variation minimization (adTVM), where the diffusivity regularization parameter was adaptively adjusted during each iteration based on the magnitude of the local pixel gradients. To test our method, we have acquired CBCT scans of phantoms using three 2DASG prototypes with different grid geometries. Projections were acquired with a linac mounted CBCT system, operated in offset detector geometry. These projections were then corrected in the following steps: first, projections were corrected using a gantry angle-specific gain correction map; next, projections were corrected by applying our adTVM method. CBCT images were reconstructed using FDK filtered backprojection algorithm. To evaluate adTVM's performance, pixel value statistics and contrast-to-noise ratio (CNR) were compared in CBCT images corrected with and without our adTVM method.
RESULTS: Without our adTVM method, all three 2DASG prototypes introduced ring artifacts with varying intensities in CBCT images. With our method, significant reduction in ring artifacts was observed in all test cases. Standard deviation of CT numbers was reduced by 7-74% in uniform density phantom CBCT images, CNR was increased by 8-67%, and CT number accuracy of contrast objects embedded in the phantom was preserved.
CONCLUSION: We propose a new method to suppress ring artifacts caused by the 2DASG's septa shadows in CBCT images. Our initial investigations indicated that adTVM method could substantially reduce such ring artifacts while preserving CT number accuracy and maintaining good spatial resolution. Therefore, our method may potentially play an important role in enabling the implementation of 2DASGs in flat panel detector based CBCT systems.
© 2019 American Association of Physicists in Medicine.

Entities:  

Keywords:  antiscatter grids; cone beam computed tomography; image enhancement/restoration(noise and artifact reduction); x-ray imaging and computed tomography

Mesh:

Year:  2019        PMID: 30802970      PMCID: PMC6559363          DOI: 10.1002/mp.13456

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  11 in total

1.  Cone-beam computed tomography with a flat-panel imager: magnitude and effects of x-ray scatter.

Authors:  J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  2001-02       Impact factor: 4.071

2.  Cone beam volume CT image artifacts caused by defective cells in x-ray flat panel imagers and the artifact removal using a wavelet-analysis-based algorithm.

Authors:  X Tang; R Ning; R Yu; D Conover
Journal:  Med Phys       Date:  2001-05       Impact factor: 4.071

3.  Reduction of ring artefacts in high resolution micro-CT reconstructions.

Authors:  Jan Sijbers; Andrei Postnov
Journal:  Phys Med Biol       Date:  2004-07-21       Impact factor: 3.609

4.  Noise removal using fourth-order partial differential equation with applications to medical magnetic resonance images in space and time.

Authors:  Marius Lysaker; Arvid Lundervold; Xue-Cheng Tai
Journal:  IEEE Trans Image Process       Date:  2003       Impact factor: 10.856

5.  Removal of ring artifacts in computed tomographic imaging using iterative center weighted median filter.

Authors:  Fazle Sadi; Soo Yeol Lee; Md Kamrul Hasan
Journal:  Comput Biol Med       Date:  2009-12-23       Impact factor: 4.589

6.  Compensation of ring artefacts in synchrotron tomographic images.

Authors:  Mirko Boin; Astrid Haibel
Journal:  Opt Express       Date:  2006-12-11       Impact factor: 3.894

7.  An augmented Lagrangian method for total variation video restoration.

Authors:  Stanley H Chan; Ramsin Khoshabeh; Kristofor B Gibson; Philip E Gill; Truong Q Nguyen
Journal:  IEEE Trans Image Process       Date:  2011-05-31       Impact factor: 10.856

8.  A simple approach to measure computed tomography (CT) modulation transfer function (MTF) and noise-power spectrum (NPS) using the American College of Radiology (ACR) accreditation phantom.

Authors:  Saul N Friedman; George S K Fung; Jeffrey H Siewerdsen; Benjamin M W Tsui
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

9.  Design and evaluation of a grid reciprocation scheme for use in digital breast tomosynthesis.

Authors:  Tushita Patel; Helen Sporkin; Heather Peppard; Mark B Williams
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-29

10.  Transmission characteristics of a two dimensional antiscatter grid prototype for CBCT.

Authors:  Cem Altunbas; Brian Kavanagh; Timur Alexeev; Moyed Miften
Journal:  Med Phys       Date:  2017-06-16       Impact factor: 4.071

View more
  3 in total

1.  An Investigation of Slot-scanning for Mammography and Breast CT.

Authors:  Andrew F T Leong; Grace J Gang; Alejandro Sisniega; Wenying Wang; Jesse Wu; Shabbir Bambot; J Webster Stayman
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2020-03-16

2.  Effect of grid geometry on the transmission properties of 2D grids for flat detectors in CBCT.

Authors:  Cem Altunbas; Timur Alexeev; Moyed Miften; Brian Kavanagh
Journal:  Phys Med Biol       Date:  2019-11-15       Impact factor: 3.609

3.  Evaluation of scatter rejection and correction performance of 2D antiscatter grids in cone beam computed tomography.

Authors:  Yeonok Park; Timur Alexeev; Brian Miller; Moyed Miften; Cem Altunbas
Journal:  Med Phys       Date:  2021-03-04       Impact factor: 4.071

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.