Transmission characteristics of a two dimensional antiscatter grid prototype for CBCT.

Cem Altunbas1, Brian Kavanagh1, Timur Alexeev1, Moyed Miften1.   

Abstract

AIM: High fraction of scattered radiation in cone-beam CT (CBCT) imaging degrades CT number accuracy and visualization of low contrast objects. To suppress scatter in CBCT projections, we developed a focused, two-dimensional antiscatter grid (2DASG) prototype. In this work, we report on the primary and scatter transmission characteristics of the 2DASG prototype aimed for linac mounted, offset detector geometry CBCT systems in radiation therapy, and compared its performance to a conventional one-dimensional ASG (1DASG).
METHODS: The 2DASG is an array of through-holes separated by 0.1 mm septa that was fabricated from tungsten using additive manufacturing techniques. Through-holes' focusing geometry was designed for offset detector CBCT in Varian TrueBeam system. Two types of ASGs were evaluated: (a) a conventional 1DASG with a grid ratio of 10, (b) the 2DASG prototype with a grid ratio of 8.2. To assess the scatter suppression performance of both ASGs, Scatter-to-primary ratio (SPR) and scatter transmission fraction (Ts) were measured using the beam stop method. Scatter and primary intensities were modulated by varying the phantom thickness between 10 and 40 cm. Additionally, the effect of air gap and bow tie (BT) filter on SPR and Ts were evaluated. Average primary transmission fraction (TP ) and pixel specific primary transmission were also measured for both ASGs. To assess the effect of transmission characteristics on projection image signal-to-noise ratio (SNR), SNR improvement factor was calculated. Improvement in contrast to noise ratio (CNR) was demonstrated using a low contrast object.
RESULTS: In comparison to 1DASG, 2DASG reduced SPRs by a factor of 3 to 6 across the range of phantom setups investigated. Ts values for 1D and 2DASGs were in the range of 21 to 29%, and 5 to 14% respectively. 2DASG continued to provide lower SPR and Ts at increased air gap and with BT filter. Tp of 1D and 2DASGs were 70.6% and 84.7% respectively. Due to the septal shadow of the 2DASG, its pixel specific primary transmission values varied between 32.5% and 99.1%. With respect to 1DASG, 2DASG provided up to factor of 1.7 more improvement in SNR across the SPR range investigated. Moreover, 2DASG provided improved visualization of low contrast objects with respect to 1DASG and NOASG setups.
CONCLUSIONS: When compared to a conventional 1DASG, 2DASG prototype provided noticeably lower SPR and Ts values, indicating its superior scatter suppression performance. 2DASG also provided 19% higher average primary transmission that was attributed to the absence of interseptal spacers and optimized grid geometry. Our results indicate that the combined effect of lower scatter and higher primary transmission provided by 2DASG may potentially translate into more accurate CT numbers and improved contrast resolution in CBCT images.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  CBCT; antiscatter grid; scatter

Mesh:

Year:  2017        PMID: 28513847      PMCID: PMC5553722          DOI: 10.1002/mp.12346

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  38 in total

1.  Cone-beam computed tomography with a flat-panel imager: magnitude and effects of x-ray scatter.

Authors:  J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  2001-02       Impact factor: 4.071

2.  Grids or air gaps for scatter reduction in digital radiography: a model calculation.

Authors:  U Neitzel
Journal:  Med Phys       Date:  1992 Mar-Apr       Impact factor: 4.071

3.  An efficient Monte Carlo-based algorithm for scatter correction in keV cone-beam CT.

Authors:  G Poludniowski; P M Evans; V N Hansen; S Webb
Journal:  Phys Med Biol       Date:  2009-06-02       Impact factor: 3.609

4.  The influence of bowtie filtration on cone-beam CT image quality.

Authors:  N Mail; D J Moseley; J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  2009-01       Impact factor: 4.071

5.  An Accurate Scatter Measurement and Correction Technique for Cone Beam Breast CT Imaging Using Scanning Sampled Measurement (SSM) Technique.

Authors:  Xinming Liu; Chris C Shaw; Tianpeng Wang; Lingyun Chen; Mustafa C Altunbas; S Cheenu Kappadath
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2006-02-28

6.  Deformable planning CT to cone-beam CT image registration in head-and-neck cancer.

Authors:  Jidong Hou; Mariana Guerrero; Wenjuan Chen; Warren D D'Souza
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

7.  Cone beam computerized tomography: the effect of calibration of the Hounsfield unit number to electron density on dose calculation accuracy for adaptive radiation therapy.

Authors:  Joan Hatton; Boyd McCurdy; Peter B Greer
Journal:  Phys Med Biol       Date:  2009-07-10       Impact factor: 3.609

8.  Improved image quality of cone beam CT scans for radiotherapy image guidance using fiber-interspaced antiscatter grid.

Authors:  Uros Stankovic; Marcel van Herk; Lennert S Ploeger; Jan-Jakob Sonke
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

9.  Compensators for dose and scatter management in cone-beam computed tomography.

Authors:  S A Graham; D J Moseley; J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  2007-07       Impact factor: 4.071

10.  Effects on image quality of a 2D antiscatter grid in x-ray digital breast tomosynthesis: Initial experience using the dual modality (x-ray and molecular) breast tomosynthesis scanner.

Authors:  Tushita Patel; Heather Peppard; Mark B Williams
Journal:  Med Phys       Date:  2016-04       Impact factor: 4.071

View more
  9 in total

1.  A novel total variation based ring artifact suppression method for CBCT imaging with two-dimensional antiscatter grids.

Authors:  Timur Alexeev; Brian Kavanagh; Moyed Miften; Cem Altunbas
Journal:  Med Phys       Date:  2019-03-20       Impact factor: 4.071

2.  Two-dimensional antiscatter grid: A novel scatter rejection device for Cone-beam computed tomography.

Authors:  Timur Alexeev; Brian Kavanagh; Moyed Miften; Cem Altunbas
Journal:  Med Phys       Date:  2018-01-08       Impact factor: 4.071

3.  Simultaneous scatter rejection and correction method using 2D antiscatter grids for CBCT.

Authors:  Zhelin Yu; Yeonok Park; Cem Altunbas
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2020-03-16

4.  Concurrent kilovoltage CBCT imaging and megavoltage beam delivery: suppression of cross-scatter with 2D antiscatter grids and grid-based scatter sampling.

Authors:  Farhang Bayat; Mohamed Elsayed Eldib; Brian Kavanagh; Moyed Miften; Cem Altunbas
Journal:  Phys Med Biol       Date:  2022-08-09       Impact factor: 4.174

5.  Megavoltage cross-scatter rejection and correction using 2D antiscatter grids in kilovoltage CBCT imaging.

Authors:  Farhang Bayat; Mohamed Elsayed Eldib; Cem Altunbas
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2022-04-04

6.  Effect of grid geometry on the transmission properties of 2D grids for flat detectors in CBCT.

Authors:  Cem Altunbas; Timur Alexeev; Moyed Miften; Brian Kavanagh
Journal:  Phys Med Biol       Date:  2019-11-15       Impact factor: 3.609

7.  Evaluation of scatter rejection and correction performance of 2D antiscatter grids in cone beam computed tomography.

Authors:  Yeonok Park; Timur Alexeev; Brian Miller; Moyed Miften; Cem Altunbas
Journal:  Med Phys       Date:  2021-03-04       Impact factor: 4.071

8.  A unified scatter rejection and correction method for cone beam computed tomography.

Authors:  Cem Altunbas; Yeonok Park; Zhelin Yu; Anant Gopal
Journal:  Med Phys       Date:  2021-02-06       Impact factor: 4.071

9.  Shading artifact correction in breast CT using an interleaved deep learning segmentation and maximum-likelihood polynomial fitting approach.

Authors:  Peymon Ghazi; Andrew M Hernandez; Craig Abbey; Kai Yang; John M Boone
Journal:  Med Phys       Date:  2019-06-23       Impact factor: 4.071

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.