Frederico Santos Lages1, Dhelfeson Willya Douglas-de-Oliveira1, Guilherme Siqueira Ibelli2, Fatimah Assaf2, Thallita Pereira Queiroz2, Fernando Oliveira Costa1.
Abstract
Resonance frequency analysis (RFA) has become the main tool used to assess the osseointegration of dental implants. The objective of this study was to verify the relationship between the ISQ values with different prosthetic abutments and with the implant platform. The hypothesis was that ISQ values changes according to the abutment height. Twelve patients were included, whose contribution to the study was 31 dental implants (external hexagon connection implants, 4.1x10 mm). The temporary implant-supported crown and prosthetic components were removed and the following smartpegs were inserted, one at a time: type 1, in the implant platform (G1); type A3, in the microunit component with 1mm transmucosal height (G2) and type A3, in the microunit component with 5mm transmucosal height (G3). In all the smartpegs, RFA measurements were taken on mesial, distal, buccal and lingual surfaces. All evaluations were performed by a single calibrated examiner (ICC = 0.989). Data were analyzed by Friedman and Spearman correlation tests and log-linear marginal regression (p<0.05). The mean age of participants was 52.83 (± 3.77) years. There was statistically significant difference (p<0.001) among the mean ISQ of G1 (88.27 ±5.70); G2 (72.75 ±4.73) and G3 (66.33 ±3.67). There was statistically significant negative correlation between the ISQ and the measurement distance (rs:-0.852; p<0.001; R2:0.553). Measurement distance was significantly associated (p<0.001) with ISQ value in the log-linear regression. The abutment height has a significant impact on resonance frequency analysis measurements. The higher the transmucosal abutment height, the lower the implant stability quotient value. Clinically, the ISQ measured on the abutment cannot be compared with values measured on the implant platform.Entities:
Mesh:
Substances:
Year: 2017 PMID: 28742164 PMCID: PMC5526494 DOI: 10.1371/journal.pone.0181873
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1Resonance frequency analysis by Osstell.
A) Implant platform B) 1mm microunit C) 5mm microunit.
ISQ comparisons among groups.
| Group | Mean (SD) | Friedman test | Post-hoc test |
|---|---|---|---|
| Platform (G1) | 88.27 (5.70) | G1XG2: <0.001 | |
| 1mm (G2) | 72.75 (4.73) | <0.001 | G1XG3: <0.001 |
| 5mm (G3) | 66.33 (3.67) | G2XG3: <0.001 |
Log-linear regression of the variable that predicts the ISQ value.
| Variable | β | Exp(β) | C.I—95% | P-value |
|---|---|---|---|---|
| Intercept | 4.46 | - | - | - |
| √Distance | -0.13 | 0.88 | [0.87; 0.89] | <0.001 |