Literature DB >> 17274715

Evaluation of two different resonance frequency devices to detect implant stability: a clinical trial.

Pilar Valderrama1, Thomas W Oates, Archie A Jones, James Simpson, John D Schoolfield, David L Cochran.   

Abstract

BACKGROUND: Resonance frequency analysis (RFA) provides a non-invasive assessment of implant stability. The established RFA device uses electronic technology, whereas a recently developed device uses magnetic technology. The goal of this clinical trial was to evaluate the ability of the magnetic RFA device to detect changes in stability during early healing following implant placement and to determine whether the implant stability quotient (ISQ) values obtained correlated with those made with the electronic device.
METHODS: RFA assessments were performed using electronic- and magnetic-based devices on 34 non-submerged titanium dental implants in 17 patients. Each patient received two implants in the posterior maxilla or mandible. Implant stability was measured at placement and weekly until week 6, when implants received provisional crowns, and at 12 weeks, when definitive crowns were cemented. During each visit, measurements were taken three times and averaged to obtain a single representative ISQ for each device.
RESULTS: At placement, the mean ISQ obtained with the electronic device was 61.9 (95% confidence interval [CI], 59.4 to 64.3); it increased to 63.2 (95% CI, 61.2 to 65.2) at 12 weeks. With the magnetic device, the mean ISQs were 70.6 (95% CI, 68.4 to 72.8) and 75.9 (95% CI, 74.2 to 77.7), respectively. Both devices indicated a pattern of decreased mean stability from 1 to 3 weeks post-placement, small fluctuations in mean ISQ from 3 to 6 weeks, and significantly increased mean stability from 6 to 12 weeks. For the complete set of implant measures across all weeks, the paired electronic and magnetic ISQ values correlated significantly (r = 0.52; P <0.001).
CONCLUSIONS: This study demonstrates that changes in implant stability measured with the newer magnetic device correlate well with those found with the electronic device. Both devices confirmed the initial decreases in implant stability that occur following placement and identified an increase in stability during the first 6 weeks of functional loading.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17274715     DOI: 10.1902/jop.2007.060143

Source DB:  PubMed          Journal:  J Periodontol        ISSN: 0022-3492            Impact factor:   6.993


  27 in total

Review 1.  Biomechanical behaviours of the bone-implant interface: a review.

Authors:  Xing Gao; Manon Fraulob; Guillaume Haïat
Journal:  J R Soc Interface       Date:  2019-07-31       Impact factor: 4.118

2.  Efficacy of autogenous teeth for the reconstruction of alveolar ridge deficiencies: a systematic review.

Authors:  Ausra Ramanauskaite; D Sahin; R Sader; J Becker; F Schwarz
Journal:  Clin Oral Investig       Date:  2019-03-11       Impact factor: 3.573

3.  The effects of elevated hemoglobin A(1c) in patients with type 2 diabetes mellitus on dental implants: Survival and stability at one year.

Authors:  Thomas W Oates; Patrick Galloway; Peggy Alexander; Adriana Vargas Green; Guy Huynh-Ba; Jocelyn Feine; C Alex McMahan
Journal:  J Am Dent Assoc       Date:  2014-12       Impact factor: 3.634

4.  Ultrasonic Evaluation of the Bone-Implant Interface.

Authors:  Yoann Hériveaux; Vu-Hieu Nguyen; Guillaume Haïat
Journal:  Adv Exp Med Biol       Date:  2022       Impact factor: 2.622

5.  Intra- and inter-operator concordance of the resonance frequency analysis. A cross-sectional and prospective clinical study.

Authors:  Raquel Antequera-Diaz; Maria-Pilar Quesada-García; Cristina Vallecillo; Marta Vallecillo-Rivas; Esther Muñoz-Soto; Maria-Victoria Olmedo-Gaya
Journal:  Clin Oral Investig       Date:  2022-07-08       Impact factor: 3.573

6.  A comparison of the implant stability among various implant systems: clinical study.

Authors:  Jae-Min Kim; Sun-Jong Kim; Inho Han; Sang-Wan Shin; Jae-Jun Ryu
Journal:  J Adv Prosthodont       Date:  2009-03-31       Impact factor: 1.904

7.  Engineered protein coatings to improve the osseointegration of dental and orthopaedic implants.

Authors:  Jordan Raphel; Johan Karlsson; Silvia Galli; Ann Wennerberg; Christopher Lindsay; Matthew G Haugh; Jukka Pajarinen; Stuart B Goodman; Ryo Jimbo; Martin Andersson; Sarah C Heilshorn
Journal:  Biomaterials       Date:  2016-01-06       Impact factor: 12.479

8.  Resonance frequency analysis-reliability in third generation instruments: Osstell mentor®.

Authors:  Mariano Herrero-Climent; Matteo Albertini; Jose-Vicente Rios-Santos; Pedro Lázaro-Calvo; Ana Fernández-Palacín; Pedro Bullon
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2012-09-01

9.  A comparative study on the accuracy of the devices for measuring the implant stability.

Authors:  In-Ho Cho; Young-Il Lee; Young-Mi Kim
Journal:  J Adv Prosthodont       Date:  2009-11-30       Impact factor: 1.904

10.  Evaluation of stability changes in tapered and parallel wall implants: a human clinical trial.

Authors:  Ar Rokn; Aar Rasouli Ghahroudi; A Mesgarzadeh; Aa Miremadi; S Yaghoobi
Journal:  J Dent (Tehran)       Date:  2011-12-20