Effective dose estimation for oncological and neurological PET/CT procedures.

Josep M Martí-Climent1,2, Elena Prieto3,4, Verónica Morán3, Lidia Sancho3, Macarena Rodríguez-Fraile3,4, Javier Arbizu3,4, María J García-Velloso3,4, José A Richter3,4.   

Abstract

BACKGROUND: The aim of this study was to retrospectively evaluate the patient effective dose (ED) for different PET/CT procedures performed with a variety of PET radiopharmaceutical compounds. PET/CT studies of 210 patients were reviewed including Torso (n = 123), Whole body (WB) (n = 36), Head and Neck Tumor (HNT) (n = 10), and Brain (n = 41) protocols with 18FDG (n = 170), 11C-CHOL (n = 10), 18FDOPA (n = 10), 11C-MET (n = 10), and 18F-florbetapir (n = 10). ED was calculated using conversion factors applied to the radiotracer activity and to the CT dose-length product.
RESULTS: Total ED (mean ± SD) for Torso-11C-CHOL, Torso-18FDG, WB-18FDG, and HNT-18FDG protocols were 13.5 ± 2.2, 16.5 ± 4.5, 20.0 ± 5.6, and 15.4 ± 2.8 mSv, respectively, where CT represented 77, 62, 69, and 63% of the protocol ED, respectively. For 18FDG, 18FDOPA, 11C-MET, and 18F-florbetapir brain PET/CT studies, ED values (mean ± SD) were 6.4 ± 0.6, 4.6 ± 0.4, 5.2 ± 0.5, and 9.1 ± 0.4 mSv, respectively, and the corresponding CT contributions were 11, 14, 23, and 26%, respectively. In 18FDG PET/CT, variations in scan length and arm position produced significant differences in CT ED (p < 0.01). For dual-time-point imaging, the CT ED (mean ± SD) for the delayed scan was 3.8 ± 1.5 mSv.
CONCLUSIONS: The mean ED for body and brain PET/CT protocols with different radiopharmaceuticals ranged between 4.6 and 20.0 mSv. The major contributor to total ED for body protocols is CT, whereas for brain studies, it is the PET radiopharmaceutical.

Entities:  

Keywords:  Brain; Effective dose; PET/CT; Radiopharmaceutical; Whole body

Year:  2017        PMID: 28439843      PMCID: PMC5403773          DOI: 10.1186/s13550-017-0272-5

Source DB:  PubMed          Journal:  EJNMMI Res            Impact factor:   3.138


Background

PET/CT is a dual-modality technique combining positron emission tomography (PET) with x-ray computed tomography (CT) [1]. Wide ranges of radiopharmaceuticals are used for studying different pathologies. The most common isotope is 18F as a label for the 2-[18F]fluoro-2-deoxy-D-glucose molecule (18FDG) [2]. Currently, PET/CT with 18FDG is a common diagnostic tool in oncology for staging and treatment evaluation [3, 4]. Thus, tumor imaging with 18FDG PET has been identified in the project “Study on European Population Doses from Medical Exposure” (DDM2) as the fourth procedure among the highest contributors to the collective effective dose in all DDM2 countries, producing a median contribution of 16.2% to total per capita effective dose in nuclear medicine procedures [5]. A PET scan has an average frequency of 0.8 per 1000 of population for European countries, and more than half the countries reported that the use of PET/CT for oncological imaging has increased. The reported administered activity for the different countries ranged from 240 to 433 MBq. The average effective dose of PET for tumor imaging was 6.7 mSv, with a wide range of variation (the ratio between maximum and minimum values was 1.7). Although the CT patient dose in PET/CT was not reported, on average, 32% of CT scanners were used as diagnostic studies. However, the centers that participated in the DDM2 study showed considerable variation from country to country: in France, CT scanners in hybrid systems were used only for attenuation correction, whereas, in Italy, they were used for diagnostic purposes. No data was included in the DDM2 study regarding brain PET/CT studies and PET tracers other than FDG [6], and dosimetric studies for those protocols are scarce. For “whole body” FDG PET/CT procedures, the relative contribution of the PET and CT effective dose has been reported to vary, with the CT component contributing between 54 and 81% of the total combined dose, depending on the CT parameters with the same scan length [7]. Furthermore, the term whole body protocol has been used to cover different body extensions: from vertex to mid-thighs [8], base of scull to upper thighs [7], and head to feet [9]. Dual-time-point imaging is a useful procedure for differentiating inflammatory and malignant processes and has been found to enhance the specificity of 18FDG PET imaging for diagnostic and prognostic purposes [10], with applications for both body [11] and brain protocols [12]. The dosimetric impact of the second PET/CT scan is produced by the CT part of the protocol. In order to estimate patient dose in PET/CT and to establish diagnostic reference levels (DRL), it is important to collect data on administered activity of PET tracer and dose-length product (DLP) or volume computed tomography dose index (CTDIvol) in CT. These parameters are established as a DRL in some countries [13]. The effective dose can then be calculated. Effective dose is defined by the International Commission on Radiological Protection as the tissue-weighted sum of the equivalent doses in all specified tissues and organs of the human body and represents the stochastic health risk to the whole body. The purpose of this study was to estimate patient effective dose of PET/CT procedures with the final motivation of patient dose optimization. To this end, DRL-related parameters were evaluated for different PET/CT procedures performed using a variety of PET radiopharmaceutical compounds, including oncological (covering different scan lengths) and neurological scans. The dosimetric impact of scan length, arm position, and dual time point imaging were studied. This dose evaluation is the initial step to subsequently achieving patient dose optimization.

Methods

PET/CT studies were performed with a Biograph mCT TrueV scanner (Siemens Healthcare, Knoxville, TN, USA), which has a PET that covers a 21.8-cm axial field of view and a 64-detector CT scanner [14]. PET scan is performed in stop-and-go imaging mode, acquiring information in a variable number of bed positions along the patient. Each bed has an overlap of 9.5 cm with the contiguous bed. For each patient, data recorded included diagnostic protocol, patient demographic data (sex, age, weight, and height), CT exposure parameters (tube voltage (kV), effective tube current-time product (mAs), rotation time, pitch, slice collimation, CTDIvol, and DLP), administered radiopharmaceutical and its activity, and PET scan parameters (number of bed positions). The automatic exposure control (“CARE Dose” package by Siemens) was used in all exams to acquire CT scans according to the lowest possible patient dose, based on the reference tube current-time product (reference mAs) defined by the user. When a dual-time imaging protocol was prescribed, the CT exposure parameters were also recorded for the second PET/CT study. The PET/CT protocols, with 18FDG, [methyl-11C]-choline (11C-CHOL), l-[methyl-11C]-methionine (11C-MET), [18F]fluoro-l-dopa (18FDOPA), or 18F-florbetapir [15], were as follows: Torso imaging (Torso) covering from the base of the skull to mid-thigh (usually with arms above the head) Whole body (WB) examination from the top of the head to the feet, performed in two steps: the first step covered the head and torso (H&Torso) followed by PET/CT from the middle of the pelvis to the toes (Limbs) Head and neck tumor (HNT) protocol performed in two steps, the head and neck portion (H&N) with the arms down, and a scan from the apex of the lung to the mid-thigh (Trunk) with the arms up Brain examinations covering the head within one bed position Patient data were retrospectively collected for a month from hospital files and PACS. For some infrequent brain PET/CT studies (18FDOPA, 11C-MET, and 18F-florbetapir), the registration period had to be extended to obtain a sample of 10. Finally, studies of 210 patients were reviewed, including 170 PETs with 18FDG, 10 with 11C-CHOL, 10 with 18FDOPA, 10 with 11C-MET, and 10 with 18F-florbetapir. This study has been approved by the institutional review board (Comité Ético de Investigación Clínica de Navarra). The need for informed consent was waived, and data anonymization provided the privacy guarantee. For each patient, effective dose was evaluated using standard coefficients referring to a generic reference individual. Thus, effective dose (ED) from the CT exam was calculated using the conversion factors k (mSv/mGy cm) multiplied by the DLP [16], depending on the scanned region (Table 1). The PET ED was calculated by multiplying the injected activity by the Γ dose coefficient for each radiopharmaceutical. The Γ coefficients in mSv/MBq were 0.019 for 18FDG [17], 0.0084 for 11C-methionine [17], 0.025 for 18FDOPA [17], 0.0044 for 11C-choline [18], and 0.0186 for 11F-florbetapir [19]. The total ED associated with the combined PET/CT exam was evaluated as the sum of the PET and CT ED values.
Table 1

Dose conversion factors from DLP to ED

CT Scan in PET/CT protocolAnatomical area k = ED/DLP (mSv/mGy cm)Reference
BrainHead0.0024[34]
H&NHead/neck0.0090[34]
Dual timeChest0.0204[34]
Dual timeAbdomen0.0163[34]
Dual timePelvis0.0143[34]
Dual timeAbdomen/pelvis0.0171[34]
Trunk, TorsoChest/abdomen/pelvis0.0186[34]
H&TorsoWhole body0.0154[34]
LimbsLower extremities0.006 male[35]
0.0073 female
Dose conversion factors from DLP to ED Statistical analysis was performed using SPSS v20 (IBM Corporation). Results are expressed as mean ± standard deviation (minimum, maximum). Statistical differences between groups were evaluated using parametric or non-parametric tests depending on the normality of the data, and they were considered significant for p < 0.05.

Results

The main technical CT parameters used in each protocol are shown in Table 2. Tube voltage was set to 120 kV, and collimation was fixed to 16 × 1.2 mm. The same parameters were used for the Torso, Trunk, H&Torso, H&N and Limbs protocols. CT parameters were adapted to each brain protocol to provide different image quality with different diagnostic value.
Table 2

Main technical parameters used for the CT protocols

Diagnostic protocolCT partRotation time(s)PitchReference current(mAs)
Torso-11C-CHOLTorso0.51120
Torso-18FDGTorso0.51120
WB-18FDGH&Torso0.51120
Limbs0.51120
HNT-18FDGTrunk0.51120
H&N0.51120
Brain-18FDGBrain10.55120
Brain-18FDOPABrain10.55100
Brain-11C-METBrain10.8200
Brain-18F-florbetapirBrain10.55400
Main technical parameters used for the CT protocols A total of 210 adult patients (130 men) were included in the study. Patient demographic characteristics (mean ± SD) were age, 61 ± 13 (21, 95) years old; weight, 73 ± 14 (40, 125) kg; and height, 167 ± 9 (140, 190) cm. The number of patients per diagnostic procedure with the corresponding PET scan parameters (administered activity and number of acquired bed positions) and CT exposure parameters (effective tube current, CTDIvol, and DLP) is shown in Table 3.
Table 3

PET scan parameters and CT exposure parameters for each protocol

Diagnostic protocolNo. of patientsActivity(MBq)Number of bed positionsMedian, %CT partEffective tube current-time product(mAs)CTDIvol (mGy)DLP(mGy cm)
Torso-11C-CHOL10695 ± 446, 80%Torso89 ± 17 (57, 113)6.5 ± 1.3 (4.2, 8.3)562 ± 120 (317, 729)
Torso-18FDG113323 ± 636, 66%Torso87 ± 27 (35, 227)6.3 ± 1.9 (3.8, 17.0)557 ± 208 (287, 1876)
WB-18FDG36321 ± 717, 44%H&Torso109 ± 35 (55, 192)7.9 ± 2.6 (4.0, 14.0)776 ± 294 (336, 1049)
7, 53%Limbs44 ± 14 (26, 94)3.3 ± 1.0 (1.9, 6.9)307 ± 104 (170, 693)
HNT-18FDG10299 ± 435, 70%Trunk74 ± 16 (54, 106)5.4 ± 1.2 (3.9, 7.8)429 ± 109 (300, 684)
2, 100%H&N67 ± 14 (44, 46)4.9 ± 1.0 (3.2, 6.3)187 ± 38 (123, 237)
Brain-18FDG11296 ± 331, 100%Brain84 ± 7 (74, 96)13.1 ± 1.0 (11.6, 15.0)302 ± 30 (247, 349)
Brain-18FDOPA10159 ± 191, 100%Brain72 ± 5 (65, 80)11.6 ± 1.4 (10.2, 15.0)271 ± 32 (237, 349)
Brain-11C-MET10453 ± 961, 100%Brain137 ± 12 (115, 155)20.4 ± 1.9 (17.9, 24.1)508 ± 45 (419, 564)
Brain-18F-florbetapir10378 ± 131, 100%Brain268 ± 25 (239, 321)39.0 ± 3.8 (37.3, 50.0)973 ± 89 (870, 1167)
PET scan parameters and CT exposure parameters for each protocol Table 4 summarizes the effective dose produced by the different PET/CT diagnostic protocols. Total mean ED for the different protocols ranged from 4.6 to 20.0 mSv. For Torso, WB, and HNT protocols, effective doses for CT were greater than for PET, whereas for brain protocols, the PET study produced a higher ED than the CT component. CT ED for Torso-11C-CHOL, Torso-18FDG, WB-18FDG, and HNT-18FDG diagnostic protocols accounted for 77, 62, 69, and 63% of the total protocol ED, respectively, and the mean PET ED was 3.1 mSv for the11C-CHOL, ranging between 5.7 and 6.1 mSv for the 18FDG protocols. The most common exploration (oncological Torso-18FDG) produced 6.1 mSv due to PET radiotracer and 10.3 mSv due to CT.
Table 4

Effective dose for PET/CT protocols with PET and CT contributions*

Diagnostic protocolPET ED(mSv)CT partCT ED(mSv)Total ED(mSv)
Torso-11C-CHOL3.1 ± 0.2 (2.8, 3.3)Torso10.4 ± 2.2 (5.9, 13.6)13.5 ± 2.2 (9.2, 16.6)
Torso-18FDG6.1 ± 1.2 (3.5, 9.5)Torso10.3 ± 3.8 (5.3, 34.9)16.5 ± 4.5 (9.7, 42.1)
WB-18FDG6.1 ± 1.3 (4.6, 10.6)H&Torso11.9 ± 4.5 (5.2, 21.7)20.0 ± 5.6 (1.8, 32.7)
Limbs2.0 ± 0.8 (1.1, 5.1)
HNT-18FDG5.7 ± 0.8 (4.7, 7.3)Trunk8.0 ± 2.0 (5.6, 12.7)15.4 ± 2.8 (12.3, 22.1)
H&N1.7 ± 0.3 (1.1, 2.1)
Brain-18FDG5.6 ± 0.6 (4.3, 6.5)Brain0.73 ± 0.07 (0.59, 0.84)6.4 ± 0.6 (5.0, 7.2)
Brain-18FDOPA4.0 ± 0.5 (3.1, 4.5)Brain0.65 ± 0.08 (0.57, 0.84)4.6 ± 0.4 (3.9, 5.2)
Brain-11C-MET4.0 ± 0.8 (2.9, 5.2)Brain1.22 ± 0.1 (1.0, 1.4)5.2 ± 0.5 (4.1, 6.5)
Brain-18F-florbetapir7.0 ± 0.2 (6.6, 7.4)Brain2.33 ± 0.2 (2.1, 2.8)9.1 ± 0.4 (8.8, 10.2)

*Values are presented as mean ± SD (minimum, maximum)

Effective dose for PET/CT protocols with PET and CT contributions* *Values are presented as mean ± SD (minimum, maximum) CT ED in brain protocols was statistically different (Kruskal-Wallis test, p < 0.01); thus, for 18F-florbetapir studies, it was 3.6 times higher than for Brain-18FDOPA studies. CT effective dose was 11, 14, 23, and 26% of the total dose received by the patients in 18FDG, 18FDOPA, 11C-MET, and 18F-florbetapir brain PET/CT studies, respectively. PET ED was also statistically different among radiotracers (ANOVA p < 0.001). In particular, 18FDOPA and 11C-MET PET scans provided lower doses than 18FDG and 18F-florbetapir PET scans (Posthoc-Bonferroni test, alpha = 0.05, p < 0.01). When arm position was considered for the Torso protocol, statistically significant differences (Mann-Whitney U test, p < 0.01) were found in mAs, CTDIvol, DLP, and CT ED between patients with arms above the head and patients with arms alongside the body. For these groups, the ED due to the CT examination (mean ± SD) was 9.7 ± 6.0 (5.3, 18.0) mSv and 17.1 ± 7.8 (8.0, 34.9) mSv, respectively. In 18FDG PET/CT scans with different scan length (HNT, Torso, and WB protocols), no significant differences in radiotracer activity and PET effective dose were found, but the CT effective dose for the CT portion that included torso or trunk was statistically different (Kruskal-Wallis test, p < 0.01). Although CT parameters were the same, scan length was different. Thus, the CT effective dose increased according to the scanned length, with 8.0, 10.4, and 11.9 mSv for the trunk, torso, and H&Torso parts of the HNT, Torso, and WB protocols, with mean scanned patient length of 761, 839, and 926 mm, respectively. Dual time PET/CT was performed in 32 patients. Only one corresponded to a brain-18FDG protocol, and in 31 patients (18% of the 169 body studies), a delayed PET/CT was performed for different parts of the torso. The frequency of the number of bed positions explored was 25, 66, and 9% for one, two, and three bed positions, respectively. The CT ED (mean ± SD) for the delayed imaging was 3.8 ± 1.5 (0.8, 6.9) mSv.

Discussion

In this study, we evaluated DRL-related parameters and estimated patient effective dose for different PET/CT procedures performed using a variety of PET radiopharmaceutical compounds, including oncological 18FDG scans (covering different patient lengths) and brain studies. This is the first 18FDG PET/CT study to evaluate the patient effective dose considering three different scanning lengths: torso scan (from base of the skull to mid-thigh), WB scan (from top of the head to the feet), and HNT examinations (from the vertex to the mid-thigh in two steps or scans). This analysis with three different scan lengths may provide an understanding of the differences among previously published results of PET/CT ED where different scan lengths have been considered (Table 5). Only Avramova-Cholakova et al. [9] have reported the impact of including limbs in ED. With regard to CT, in this study, the objective of the CT study was anatomical localization in most cases; the CT exposure parameters were fixed for each protocol, and no contrast enhancement was used in any case. However, as different protocols covered different patient lengths, a variety of conditions have been evaluated. Previously published studies have analyzed differences between adapted CT protocols for normal and obese patients or between contrast-enhanced CT [20] and CT adapted for high-quality or interspaced high-speed CT [21]. Others demonstrated an increase in CT dose when moving from a standard protocol to a diagnostic protocol [22] or from low-dose protocols to contrast-enhanced protocols [23]. Willowson et al. [8] showed that PET ED decreases from 6.3 to 6.0 mSv when ED is scaled to individual patient weight.
Table 5

Summary of published data on PET/CT parameters and patient effective dose for 18FDG whole body oncological studies

ReferenceScan length as per this studyStudy fromStudy toComment or CT descriptionTube current mAs or mA and sAECCTDIvol (mGy)DLP (mGy cm)ED_CT (mSv)Activity (MBq)ED_PET (mSv)
[7]TorsoBase of the skullUpper-thighImpact of AEC100–300 mA and 0.5 sYes7.23706.23
250 mA and 0.5 sNo18.6
[36]TorsoBase of the skullUpper-thighTAI patients112 mAsYes14.4 ± 2.8312 ± 434.4 ± 0.6
[22]TorsoBase of the skullMid-thighsStandard CT39 mAsYes438 ± 855.0 ± 1.0450 ± 329.0 ± 1.6
Diagnostic CT120 mAs1050 ± 34215.4 ± 5.0
[30]H&TorsoHeadMid-thighsInitial CTYes6.4 ± 2.4537 ± 2228.1 ± 3.3
Optimized CT4.3 ± 1.6368 ± 1455.5 ± 2.1
[20]H&TorsoTop of the skullMid-thighsMost frequent CT110 mA and 0.8 s4.34347.5336 ± 495.6
Special CT250 mA and 0.5 s6.061610.95
Special CT300 mA and 0.5 s7.273213.2
[8]H&TorsoVertexMid-thighs80 mAsYes8.2304 ± 456.3 ± 1.2
[21]H&TorsoHeadPelvisHigh-quality CT80 mA and 0.8 s19.0
High-speed CT8.8
[9]H&TorsoHeadMid-thighsCenter 184 mA and 0.5 sYes5.9 ± 1591 ± 1307. 8258 ± 564.9
Center 2Yes5.4 ± 2575 ± 2487.7313 ± 735.9
WBHeadFeetCenter 189 mA & 0.5 sYes6.2 ± 1.41112 ± 2859.4258 ± 564.9
Center 2Yes3.7 ± 1620 ± 1665.9313 ± 735.9
[23]SymphysisLow-dose CT1–2.91.3–4.5300–3705.7–7
Contrast-enhanced CT9.5–14.114.1–18.6
This studyTorsoBase of the skullMid-thighs120 mAsYes6.355710.33236.1 ± 1.2
HNTVertexMid-thighs4.9 + 5.4187 + 4299.7 ± 2.2299 ± 435.7 ± 0.8
WBHeadFeet7.9 + 3.3776 + 30714.0 ± 4.9321 ± 716.1 ± 1.3
Summary of published data on PET/CT parameters and patient effective dose for 18FDG whole body oncological studies CTDIvol 75th percentile of whole-body oncological PET/CT protocols submitted to the Society of Nuclear Medicine and Molecular Imaging Clinical Trials Network ranged from 9.7 to 10.2 mGy [24]. The wide ranges of CT acquisition parameters suggested that CTDIvol reference levels may help to optimize CT protocols. In a French national survey for whole-body PET/CT [25], the average 18FDG administered activity, CTDIvol, and DLP were 310 MBq, 6.6 mGy, and 628 mGy cm, respectively. The ED whole-body CT (from neck to thighs) was 8.6 mSv. A Japanese national survey [26] also observed the different definition of whole-body protocol; the most commonly used extension was from the top of the head to the thigh. CT and 18FDG PET ED for male (female) patients were 10.1 (9.7) and 4.5 (3.7) mSv, respectively. For 18FDG PET tumor imaging in our patient cohort, the mean PET effective dose, directly proportional to the administered activity, was approximately 6 mSv, within the range of the published data 4.9 to 9 mSv (from 258 to 450 MBq of 18FDG). In the European Union, only seven countries have DRL, ranging from 200 to 400 MBq [13], with a variability of 100% in the DRL and of 43% in activity (Table 5). The CT ED depends on the scan operating parameters, scanning length, and patient area. For ED evaluation, this dependence is reflected in CTDIvol and DLP parameters as well as in the dose conversion factor. Of the evaluated protocols, the H&Torso part of the WB protocol had the highest mean DLP value, and the k factor was slightly lower than the value for Trunk and Torso scans. The Trunk CT part of the HNT-18FDG protocol also had a lower DLP than the Torso CT scans. CT ED for Torso-18FDG and HNT-18FDG protocols were similar (10 mSv) and lower than for WB-18FDG scan (14 mSv). These values, which demonstrate dose dependence on the PET/CT protocol used, lie within the range of published data considering the PET/CT protocol (Table 5). For the WB protocol, ED may be potentially reduced if instead of performing two scans (one for H&Torso and another one for Limbs) with an overlap zone, a single CT study (without overlaping) would be performed. The impact on the CT effective dose due to the position of the arms in a PET/CT study and the impact of the second PET/CT in a dual-time-point imaging scan have never been reported. In this study, for the Torso-18FDG protocol, we showed that ED increases by approximately a factor of two when the arms are alongside the body, compared to when the arms are positioned above the head. Additionally, for dual-time-point imaging, total ED was increased by 3.8 mSv in the 18% of the patients with body protocols. Few references exist on patient dosimetry in brain studies, other than those reporting radiopharmaceutical biodistribution and organ dosimetry. This is probably because these protocols are less frequent and produce a lower dose in comparison with oncological 18FDG studies. In 2014, only Ireland had a DRL for 18FDG brain studies (290 MBq) [13]. Patient dosimetry for neurological PET/CT protocols with 18FDOPA, 18FDG, 11C-MET, and 18F-florbetapir have been evaluated. The purpose of the CT scan was different in each protocol, and the reference tube current was adjusted according to the required image quality for that specific purpose (e.g., anatomical localization, or diagnosis). In particular, the tube currents for the 11C-MET and 18F-florbetapir protocols were adjusted for diagnostic purposes. Hence, the mean CT ED in a Brain-18F-florbetapir study (2.33 mSv) was 3.6 times higher than the CT ED in a Brain-18FDOPA study (0.65 mSv). Thus, for amyloid brain imaging, it would be desirable to distinguish the relative accumulation of radiotracer binding between gray and white matter, as well as possible pitfalls due to bone deposits and atrophy. CT is also important in neurooncological studies using 11C-MET PET/CT studies, where the CT can help by adding structural information. CT doses reported in this study are similar to those in the study by Kaushik et al. [20], where brain CT parameters were adapted for obese patients and contrast-enhanced studies. Effective doses due to the radiopharmaceutical compounds in the brain PET/CT protocols have been found to contribute between 77 and 88% of the total dose. Kaushik et al. evaluated a PET effective dose of 3.5 and 5.8 mSv for 18FDOPA and 18FDG brain PET/CT studies [20], respectively, similar to that found in this study. The PET dose for 18F-florbetapir corresponds to the dose reported for 370 MBq activity administration to a 70-kg patient [19]. Finally, further studies are needed in the field of patient dose optimization for each PET/CT protocol. The effective dose due to the PET scan is directly proportional to the administered dose. Hence, reducing the injected activity will provide a direct reduction in the patient dose. In addition, there will be a positive impact on the operating costs of the PET/CT facility and in the medical staff dosimetry performing the examinations. The new generation of PET scanners, which incorporate improved reconstruction methods with point spread function and time of flight information [27] and image-noise modeling [28], should be used to reduce the injected activity in the patient [29]. CT protocol optimization results in a reduction in the mean CT radiation dose and analysis of the image quality can show no clinically relevant degradation of the low-dose studies [30]. Typically, PET/CT tomographs acquire the PET scan in a stop-and-go mode, imaging a fixed number of bed positions with overlapping. Hence, the total length is usually more extense than the area of interest. In PET acquisition with a continuous table motion [31], the total scan length is not forced to be a specific number of bed positions and can be adjusted to any area of interest to be explored, hence reducing total CT scanned length and patient dose. Recent advances in CT such as iterative reconstruction [32] and adaptive kV [33] have shown a reduction in patient CT dose, additional to the dose reduction achieved with the tube current modulation used in this study. The main limitation of this study is the simplified approach used to estimate effective dose, using the same conversion coefficient independently of weight, age, or sex of the patient. Therefore, the estimated doses refer to a generic reference individual. For a more realistic analysis of the actual PET doses received by each patient, additional information about the individual biokinetics and anatomical and physiological properties are required. However, this simple approach is suitable for the aim of this study, which was to estimate and compare population doses for different PET/CT procedures; this approach is also used elsewhere to obtain population PET and CT doses [8]. Another limitation is that all PET coefficients were obtained from the ICRP, except for those not included in the ICPR documents. Those coefficients were obtained from other publications on individual biokinetics and dosimetric studies, where the procedure for calculating the dose may not be fully equivalent to the one used by ICRP.

Conclusions

The mean effective dose for body and brain PET/CT protocols with different radiopharmaceuticals (18F-FDG, 11C-choline, 11C-methionine, 18FDOPA, and 18F-florbetapir) ranged between 4.6 and 20.0 mSv. In body protocols, CT is the major contributor to the total effective dose. Moreover, the increase in total scan length, the position of the arms alongside the body instead of above the head, and the delayed PET/CT study in a dual-time protocol significantly increased the contribution of the CT effective dose to the total effective dose. Hence, inclusion of the head and legs in a body protocol, the positioning of the arms alongside the body, and the acquisition of delayed images should be justified for each patient in order to optimize the PET/CT protocol dose and to maintain the dose as low as reasonably achievable without compromising diagnostic purpose. For brain protocols, although the main contribution to the total effective dose was due to the PET radiopharmaceutical, CT can increase the dose by a factor of about 4 if it is used for diagnostic purposes.
  32 in total

1.  A retrospective evaluation of radiation dose associated with low dose FDG protocols in whole-body PET/CT.

Authors:  Kathy P Willowson; Elizabeth A Bailey; Dale L Bailey
Journal:  Australas Phys Eng Sci Med       Date:  2011-12-10       Impact factor: 1.430

2.  Diagnostic Reference Levels of CT Radiation Dose in Whole-Body PET/CT.

Authors:  Ngoneh Jallow; Paul Christian; John Sunderland; Michael Graham; John M Hoffman; Jonathon A Nye
Journal:  J Nucl Med       Date:  2015-11-12       Impact factor: 10.057

3.  Patient doses from PET-CT procedures.

Authors:  S Avramova-Cholakova; S Ivanova; E Petrova; M Garcheva; J Vassileva
Journal:  Radiat Prot Dosimetry       Date:  2015-04-05       Impact factor: 0.972

4.  Methods of CT Dose Estimation in Whole-Body ¹⁸F-FDG PET/CT.

Authors:  Yusuke Inoue; Kazunori Nagahara; Yoshihito Tanaka; Hiroki Miyatake; Hirofumi Hata; Toshimasa Hara
Journal:  J Nucl Med       Date:  2015-04-03       Impact factor: 10.057

5.  Estimates of effective dose for CT scans of the lower extremities.

Authors:  Natalia Saltybaeva; Mary Ellen Jafari; Martin Hupfer; Willi A Kalender
Journal:  Radiology       Date:  2014-06-16       Impact factor: 11.105

6.  Reducing Radiation Dose at Chest CT: Comparison Among Model-based Type Iterative Reconstruction, Hybrid Iterative Reconstruction, and Filtered Back Projection.

Authors:  Constance de Margerie-Mellon; Cédric de Bazelaire; Claire Montlahuc; Jérôme Lambert; Antoine Martineau; Philippe Coulon; Eric de Kerviler; Catherine Beigelman
Journal:  Acad Radiol       Date:  2016-06-23       Impact factor: 3.173

7.  Whole-body PET/CT scanning: estimation of radiation dose and cancer risk.

Authors:  Bingsheng Huang; Martin Wai-Ming Law; Pek-Lan Khong
Journal:  Radiology       Date:  2009-02-27       Impact factor: 11.105

8.  Radiation dosimetry of 18F-FDG PET/CT: incorporating exam-specific parameters in dose estimates.

Authors:  Brian Quinn; Zak Dauer; Neeta Pandit-Taskar; Heiko Schoder; Lawrence T Dauer
Journal:  BMC Med Imaging       Date:  2016-06-18       Impact factor: 1.930

9.  Radiation dose reduction via sinogram affirmed iterative reconstruction and automatic tube voltage modulation (CARE kV) in abdominal CT.

Authors:  Hyun Joo Shin; Yong Eun Chung; Young Han Lee; Jin-Young Choi; Mi-Suk Park; Myeong-Jin Kim; Ki Whang Kim
Journal:  Korean J Radiol       Date:  2013-11-05       Impact factor: 3.500

10.  Performance evaluation of the Biograph mCT Flow PET/CT system according to the NEMA NU2-2012 standard.

Authors:  Ivo Rausch; Jacobo Cal-González; David Dapra; Hans Jürgen Gallowitsch; Peter Lind; Thomas Beyer; Gregory Minear
Journal:  EJNMMI Phys       Date:  2015-10-26
View more
  14 in total

1.  Dynamic PET imaging with ultra-low-activity of 18F-FDG: unleashing the potential of total-body PET.

Authors:  Xiaoli Lan; Kevin Fan; Ke Li; Weibo Cai
Journal:  Eur J Nucl Med Mol Imaging       Date:  2021-01-30       Impact factor: 9.236

2.  CT dose modulation using automatic exposure control in whole-body PET/CT: effects of scout imaging direction and arm positioning.

Authors:  Yusuke Inoue; Kazunori Nagahara; Hiroko Kudo; Hiroyasu Itoh
Journal:  Am J Nucl Med Mol Imaging       Date:  2018-04-25

3.  Conversion from dose length product to effective dose for the CT component of whole-body PET/CT.

Authors:  Yusuke Inoue; Yu Ohkubo; Kazunori Nagahara; Ayuka Uehara; Makoto Takano
Journal:  Ann Nucl Med       Date:  2022-01-24       Impact factor: 2.668

4.  Optimizing Immuno-PET Imaging of Tumor PD-L1 Expression: Pharmacokinetic, Biodistribution, and Dosimetric Comparisons of 89Zr-Labeled Anti-PD-L1 Antibody Formats.

Authors:  Alizée Bouleau; Hervé Nozach; Steven Dubois; Dimitri Kereselidze; Céline Chevaleyre; Cheng-I Wang; Michael J Evans; Vincent Lebon; Bernard Maillère; Charles Truillet
Journal:  J Nucl Med       Date:  2021-12-21       Impact factor: 11.082

5.  Maximizing the use of batch production of 18F-FDOPA for imaging of brain tumors to increase availability of hybrid PET/MR imaging in clinical setting.

Authors:  Mariam Aboian; Ramon Barajas; Julia Shatalov; Vahid Ravanfar; Emma Bahroos; Elizabeth Tong; Jennie W Taylor; N Oberheim Bush; Patricia Sneed; Youngho Seo; Soonmee Cha; Miguel Hernandez-Pampaloni
Journal:  Neurooncol Pract       Date:  2020-10-14

6.  89Zr-nimotuzumab for immunoPET imaging of epidermal growth factor receptor I.

Authors:  Rufael Chekol; Viswas Raja Solomon; Elahe Alizadeh; Wendy Bernhard; Darrell Fisher; Wayne Hill; Kris Barreto; John Francis DeCoteau; Angel Casaco Parada; Clarence Ronald Geyer; Humphrey Fonge
Journal:  Oncotarget       Date:  2018-03-30

7.  Application of 82 Sr/82 Rb generator in neurooncology.

Authors:  Nikolay A Kostenikov; Boris L Zhuikov; Valeriy M Chudakov; Yuriy R Iliuschenko; Sergey V Shatik; Vadim V Zaitsev; Dmitriy S Sysoev; Andrey A Stanzhevskiy
Journal:  Brain Behav       Date:  2019-02-06       Impact factor: 2.708

8.  The Use of Positron Emission Tomography in Coronavirus Disease 2019 Cases.

Authors:  Joseph C Lee; John K Blazak
Journal:  J Thorac Oncol       Date:  2020-05-05       Impact factor: 15.609

9.  A Nordic survey of CT doses in hybrid PET/CT and SPECT/CT examinations.

Authors:  Natalie A Bebbington; Bryan T Haddock; Henrik Bertilsson; Eero Hippeläinen; Ellen M Husby; Virpi I Tunninen; Marcus Söderberg
Journal:  EJNMMI Phys       Date:  2019-12-16