Abstract
BACKGROUND: This study aimed to evaluate the prediction accuracy of postoperative refractions using partial coherence interferometry (IOL-Master) and applanation ultrasound (AL-3000) assisted with corneal topography (TMS-4) in eyes that had undergone myopic laser-assisted in situ keratomileusis (LASIK).Entities:
Keywords: Biometry; Cataract surgery; Intraocular lens power calculation; Lasik
Mesh:
Year: 2017 PMID: 28390411 PMCID: PMC5385019 DOI: 10.1186/s12886-017-0439-x
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Results of 10 post-LASIK eyes
| Method | ME | SD | Range | MAE | MedAE | 95% CI | ± 0.5 D | ± 1.0 D | ± 1.5 D |
|---|---|---|---|---|---|---|---|---|---|
| IOL-Master | |||||||||
| Haigis-L | 0.01 | 0.721a | – 1.15–1.23 | 0.604 | 0.465 | 0.361–0.847 | 6 | 8 | 10 |
| Koch–Maloney (Haigis formula) | – 0.15 | 0.695 | – 1.00–1.19 | 0.557 | 0.510 | 0.312–0.842 | 5 | 9 | 10 |
| Shammas c.d. (Haigis formula) | 0.09 | 0.695 | – 0.80–1.41 | 0.557 | 0.470 | 0.283–0.831 | 6 | 9 | 10 |
| Shammas c.d. (Shammas-PL formula) | – 0.02 | 0.698a | – 1.19–0.99 | 0.558 | 0.519 | 0.334–0.822 | 5 | 9 | 10 |
| AL-3000 assisted with TMS-4 | |||||||||
| Haigis-L | 0.45 | 0.782 | – 0.94–1.37 | 0.774 | 0.876 | 0.480–1.067 | 4 | 7 | 10 |
| Koch–Maloney (Haigis formula) | 0.25 | 0.741 | – 0.88–1.24 | 0.627 | 0.781 | 0.324–0.931 | 4 | 8 | 10 |
| Shammas c.d. (Haigis formula) | 0.49 | 0.743 | – 0.67–1.47 | 0.734 | 0.708 | 0.396–1.072 | 3 | 6 | 10 |
| Shammas c.d. (Shammas-PL formula) | 0.03 | 0.778 | – 1.36–1.13 | 0.653 | 0.583 | 0.392–0.914 | 4 | 8 | 10 |
ME mean error, SD standard deviation, MAE mean absolute error, MedAE median absolute error, 95% CI, 95% confidence interval of mean absolute error; ±0.5 D, ±1.0 D, and ±1.5 D = number of refractions within 0.5 D, 1.0 D, and 1.5 D of prediction
aWorse than Haigis with measurements using IOL-Master in all 133 virgin eyes (P < 0.05)
Fig. 1Box plot of prediction errors for each method. Green color represents measurements from IOL-Master. Yellow color represents measurements from AL-3000 and TMS-4
Intraocular lens constants used for formulas
| Formula | ||||
|---|---|---|---|---|
| Haigis | Shammas-PL | |||
| IOL model | a0 | a1 | a2 | C |
| IOL-Master | ||||
| 920H | 0.154 | 0.166 | 0.166 | 4.612 |
| Adapt-AO | −1.418 | 0.629 | 0.176 | 4.710 |
| SN60WF | 1.948 | 0.376 | 0.076 | 4.883 |
| ZCB00 | −3.031 | 0.331 | 0.309 | 5.403 |
| AL-3000 assisted with TMS-4 | ||||
| 920H | −1.068 | 0.209 | 0.191 | 4.176 |
| Adapt-AO | −0.319 | 0.563 | 0.106 | 4.114 |
| SN60WF | 3.091 | 0.319 | 0.009 | 4.432 |
| ZCB00 | 0.632 | 0.614 | 0.095 | 4.931 |