A three-part system for refining intraocular lens power calculations.

J T Holladay1, T C Prager, T Y Chandler, K H Musgrove, J W Lewis, R S Ruiz.   

Abstract

A three-part system that determines the correct power for an intraocular lens (IOL) to achieve a desired postoperative refraction is presented. The three components are (1) data screening criteria to identify improbable axial length and keratometry measurements, (2) a new IOL calculation formula that exceeds the current accuracy of other formulas for short, medium, and long eyes, and (3) a personalized "surgeon factor" that adjusts for any consistent bias in the surgeon's results, from any source, based on a reverse solution of the new formula; the reverse solution uses the postoperative stabilized refraction, the dioptric power of the implanted IOL, and the preoperative corneal and axial length measurements to calculate the personalized surgeon factor. The improved accuracy of the new formula was proven by performing IOL power calculations on 2,000 eyes from 12 surgeons and comparing the results to seven other currently used formulas.

Entities:  

Mesh:

Year:  1988        PMID: 3339543     DOI: 10.1016/s0886-3350(88)80059-2

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  101 in total

1.  Predicting the refractive outcome after cataract surgery: the comparison of different IOLs and SRK-II v SRK-T.

Authors:  M J Elder
Journal:  Br J Ophthalmol       Date:  2002-06       Impact factor: 4.638

2.  IOLMaster biometry: refractive results of 100 consecutive cases.

Authors:  H Eleftheriadis
Journal:  Br J Ophthalmol       Date:  2003-08       Impact factor: 4.638

3.  Calculation of the power of anterior chamber implants.

Authors:  J R Villada; P S Raj; T Akingbehin
Journal:  Br J Ophthalmol       Date:  1992-05       Impact factor: 4.638

4.  Comparison of the biometric formulas used for applanation A-scan ultrasound biometry.

Authors:  Fatih Özcura; Serdar Aktaş; Hacı Murat Sağdık; Mehmet Tetikoğlu
Journal:  Int Ophthalmol       Date:  2016-02-08       Impact factor: 2.031

5.  Comparison of formulas and methods for high myopia patients requiring intraocular lens powers less than six diopters.

Authors:  Harry S Geggel
Journal:  Int Ophthalmol       Date:  2017-06-20       Impact factor: 2.031

6.  Comparison of intraocular lens power calculation methods in eyes that have undergone laser-assisted in-situ keratomileusis.

Authors:  Li Wang; Marc A Booth; Douglas D Koch
Journal:  Trans Am Ophthalmol Soc       Date:  2004

7.  Paediatric secondary intraocular lens estimation from the aphakic refraction alone: comparison with a standard biometric technique.

Authors:  A O Khan; A AlGaeed
Journal:  Br J Ophthalmol       Date:  2006-08-17       Impact factor: 4.638

8.  Intraocular lens power calculation after myopic and hyperopic laser vision correction using optical coherence tomography.

Authors:  Maolong Tang; Li Wang; Douglas D Koch; Yan Li; David Huang
Journal:  Saudi J Ophthalmol       Date:  2012-01

9.  [Determination of toric intraocular lenses].

Authors:  A Langenbucher; Anja Viestenz; N Szentmáry; W Behrens-Baumann; A Viestenz
Journal:  Ophthalmologe       Date:  2008-07       Impact factor: 1.059

10.  [Biometry in cataract camps. Experiences from north Kenya].

Authors:  S Briesen; H Roberts; J Karimurio; M Kollmann
Journal:  Ophthalmologe       Date:  2010-04       Impact factor: 1.059

View more