| Literature DB >> 28332342 |
Michael P Barnes1,2,3, Peter B Greer1,3.
Abstract
Machine Performance Check (MPC) is an automated and integrated image-based tool for verification of beam and geometric performance of the TrueBeam linac. The aims of the study were to evaluate the performance of the MPC geometric tests relevant to OBI/CBCT IGRT geometric accuracy. This included evaluation of the MPC isocenter and couch tests. Evaluation was performed by comparing MPC to QA tests performed routinely in the department over a 4-month period. The MPC isocenter tests were compared against an in-house developed Winston-Lutz test and the couch compared against routine mechanical QA type procedures. In all cases the results from the routine QA procedure was presented in a form directly comparable to MPC to allow a like-to-like comparison. The sensitivity of MPC was also tested by deliberately miscalibrating the appropriate linac parameter. The MPC isocenter size and MPC kV imager offset were found to agree with Winston-Lutz to within 0.2 mm and 0.22 mm, respectively. The MPC couch tests agreed with routine QA to within 0.12 mm and 0.15°. The MPC isocenter size and kV imager offset parameters were found to be affected by a change in beam focal spot position with the kV imager offset more sensitive. The MPC couch tests were all unaffected by an offset in the couch calibration but the three axes that utilized two point calibrations were sensitive to a miscalibration of the size in the span of the calibration. All MPC tests were unaffected by a deliberate misalignment of the MPC phantom and roll of the order of one degree.Entities:
Keywords: linac quality assurance; machine performance check (MPC)
Mesh:
Year: 2017 PMID: 28332342 PMCID: PMC5689847 DOI: 10.1002/acm2.12064
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Short term repeatability of the MPC isocenter and couch geometric tests based upon five successive measurements
| Test | Standard Deviation |
|---|---|
| Couch | |
| Lateral | 0.04 (mm) |
| Longitudinal | 0.02 (mm) |
| Vertical | 0.02 (mm) |
| Pitch | 0.01 (Degrees) |
| Roll | 0.00 (Degrees) |
| Rotation | 0.01 (Degrees) |
| Isocenter | |
| kV offset | 0.05 (mm) |
| size | 0.02 (mm) |
Figure 1Comparison between isocenter sizes as measured using MPC and in‐house Winston–Lutz.
Figure 2MPC kV imager offset and kV imager offset and in‐house Winston–Lutz distance between imaging and radiation isocenter.
Figure 3MPC isocenter size and kV imager offsets leading up to and following the focal spot adjustment
MPC isocenter size and kV imager offset results before and after the focal spot position adjustment. [mean (mm) ± 1 Standard deviation]
| MPC Parameter | Before adjustment (n = 20) | After adjustment (n = 20) |
|---|---|---|
| Isocenter size | 0.35 ± 0.02 | 0.37 ± 0.02 |
| kV Imager offsett | 0.14 ± 0.03 | 0.22 ± 0.04 |
Comparison between MPC (n = 89) and mechanical QA (n = 7) couch tests. Difference from nominal. Mean ± 1 SD
| Couch | MPC (mm/°) | Mechanical QA (mm/°) |
|---|---|---|
| Lateral | −0.16 ± 0.04 | −0.04 ± 0.04 |
| Longitudinal | 0.02 ± 0.02 | 0.02 ± 0.04 |
| Vertical | −0.02 ± 0.03 | −0.08 ± 0.40 |
| Rotation | −0.16 ± 0.01 | −0.01 ± 0.07 |
| Pitch | −0.01 ± 0.01 | 0.00 ± 0.09 |
| Roll | −0.01 ± 0.01 | −0.02 ± 0.10 |
MPC measured changes in couch lateral, longitudinal, and vertical with deliberate miscalibration of the calibration span
| Expected change (mm) | MPC change (mm) | Difference (mm) (expected – MPC) | |
|---|---|---|---|
|
| |||
| Larger span | −0.75 | −0.76 | 0.01 |
| Smaller span | 0.75 | 0.71 | 0.04 |
|
| |||
| Larger span | 0.75 | 0.76 | −0.01 |
| Smaller span | −0.75 | −0.79 | 0.04 |
|
| |||
| Larger span | 2.0 | 1.99 | 0.01 |
| Smaller span | −2.0 | −1.96 | −0.04 |