A radiographic and tomographic imaging system integrated into a medical linear accelerator for localization of bone and soft-tissue targets.

D A Jaffray1, D G Drake, M Moreau, A A Martinez, J W Wong.   

Abstract

PURPOSE: Dose escalation in conformal radiation therapy requires accurate field placement. Electronic portal imaging devices are used to verify field placement but are limited by the low subject contrast of bony anatomy at megavoltage (MV) energies, the large imaging dose, and the small size of the radiation fields. In this article, we describe the in-house modification of a medical linear accelerator to provide radiographic and tomographic localization of bone and soft-tissue targets in the reference frame of the accelerator. This system separates the verification of beam delivery (machine settings, field shaping) from patient and target localization.
MATERIALS AND METHODS: A kilovoltage (kV) x-ray source is mounted on the drum assembly of an Elekta SL-20 medical linear accelerator, maintaining the same isocenter as the treatment beam with the central axis at 90 degrees to the treatment beam axis. The x-ray tube is powered by a high-frequency generator and can be retracted to the drum-face. Two CCD-based fluoroscopic imaging systems are mounted on the accelerator to collect MV and kV radiographic images. The system is also capable of cone-beam tomographic imaging at both MV and kV energies. The gain stages of the two imaging systems have been modeled to assess imaging performance. The contrast-resolution of the kV and MV systems was measured using a contrast-detail (C-D) phantom. The dosimetric advantage of using the kV imaging system over the MV system for the detection of bone-like objects is quantified for a specific imaging geometry using a C-D phantom. Accurate guidance of the treatment beam requires registration of the imaging and treatment coordinate systems. The mechanical characteristics of the treatment and imaging gantries are examined to determine a localizing precision assuming an unambiguous object. MV and kV radiographs of patients receiving radiation therapy are acquired to demonstrate the radiographic performance of the system. The tomographic performance is demonstrated on phantoms using both the MV and the kV imaging system, and the visibility of soft-tissue targets is assessed. RESULTS AND DISCUSSION: Characterization of the gains in the two systems demonstrates that the MV system is x-ray quantum noise-limited at very low spatial frequencies; this is not the case for the kV system. The estimates of gain used in the model are validated by measurements of the total gain in each system. Contrast-detail measurements demonstrate that the MV system is capable of detecting subject contrasts of less than 0.1% (at 6 and 18 MV). A comparison of the kV and MV contrast-detail performance indicates that equivalent bony object detection can be achieved with the kV system at significantly lower doses (factors of 40 and 90 lower than for 6 and 18 MV, respectively). The tomographic performance of the system is promising; soft-tissue visibility is demonstrated at relatively low imaging doses (3 cGy) using four laboratory rats.
CONCLUSIONS: We have integrated a kV radiographic and tomographic imaging system with a medical linear accelerator to allow localization of bone and soft-tissue structures in the reference frame of the accelerator. Modeling and experiments have demonstrated the feasibility of acquiring high-quality radiographic and tomographic images at acceptable imaging doses. Full integration of the kV and MV imaging systems with the treatment machine will allow on-line radiographic and tomographic guidance of field placement.

Entities:  

Mesh:

Year:  1999        PMID: 10524434     DOI: 10.1016/s0360-3016(99)00118-2

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  54 in total

Review 1.  Current progress in adaptive radiation therapy for head and neck cancer.

Authors:  David L Schwartz
Journal:  Curr Oncol Rep       Date:  2012-04       Impact factor: 5.075

2.  Electronic portal imaging vs kilovoltage imaging in fiducial marker image-guided radiotherapy for prostate cancer: an analysis of set-up uncertainties.

Authors:  S Gill; J Thomas; C Fox; T Kron; A Thompson; S Chander; S Williams; K H Tai; G Duchesne; F Foroudi
Journal:  Br J Radiol       Date:  2011-10-05       Impact factor: 3.039

3.  Accuracy estimation for projection-to-volume targeting during rotational therapy: a feasibility study.

Authors:  Yong Long; Jeffrey A Fessler; James M Balter
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

4.  A pseudoinverse deformation vector field generator and its applications.

Authors:  C Yan; H Zhong; M Murphy; E Weiss; J V Siebers
Journal:  Med Phys       Date:  2010-03       Impact factor: 4.071

5.  Monte Carlo investigations of megavoltage cone-beam CT using thick, segmented scintillating detectors for soft tissue visualization.

Authors:  Yi Wang; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao; Amit Sawant; Hong Du
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

6.  Dosimetric properties and commissioning of cone-beam CT image beam line with a carbon target.

Authors:  Y Dzierma; F G Nuesken; N P Licht; C Ruebe
Journal:  Strahlenther Onkol       Date:  2013-05-30       Impact factor: 3.621

7.  Progressive cone beam CT dose control in image-guided radiation therapy.

Authors:  Hao Yan; Xin Zhen; Laura Cerviño; Steve B Jiang; Xun Jia
Journal:  Med Phys       Date:  2013-06       Impact factor: 4.071

8.  Cone beam breast CT with a high pitch (75 μm), thick (500 μm) scintillator CMOS flat panel detector: visibility of simulated microcalcifications.

Authors:  Youtao Shen; Yuncheng Zhong; Chao-Jen Lai; Tianpeng Wang; Chris C Shaw
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

Review 9.  Anatomic, functional and molecular imaging in lung cancer precision radiation therapy: treatment response assessment and radiation therapy personalization.

Authors:  Michael MacManus; Sarah Everitt; Tanja Schimek-Jasch; X Allen Li; Ursula Nestle; Feng-Ming Spring Kong
Journal:  Transl Lung Cancer Res       Date:  2017-12

10.  Comparative analysis of image guidance in two institutions for prostate cancer patients.

Authors:  Tomasz Piotrowski; Slav Yartsev; George Rodrigues; Tomasz Bajon
Journal:  Rep Pract Oncol Radiother       Date:  2014-01-02
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.