| Literature DB >> 30178521 |
Michael P Barnes1,2,3, Dennis Pomare1, Frederick W Menk3, Buiron Moraro4, Peter B Greer1,3.
Abstract
Alignment of the On-Board Imager (OBI) X-ray tube is important for ensuring imaging to treatment isocenter coincidence, which in turn is important for accurate Image Guided Radiotherapy (IGRT). Varian introduced a new X-ray tube alignment procedure for the TrueBeam linac in software version 2.5 MR2 as part of the machine performance check (MPC) application. This study evaluated the new procedure against conventional methods and examined the clinical significance of X-ray tube misalignment. Long term stability and short term repeatability of MPC tube alignment was assessed as well as sensitivity of the method to setup error. Standard quality assurance tests expected to be sensitive to tube misalignment were performed before and after tube alignment. These tests included: IsoCal verification; MPC kV imager offset; Winston-Lutz: kV imaging to treatment/radiation isocenter coincidence; CBCT image QA using the Catphan phantom; and OBI image geometric accuracy and center pixel alignment. Tube alignment measurements were performed with MPC, the two-plate method, and wire-on-faceplate method. The X-ray tube was then realigned by approximately 1.01 mm in the tangential plane based upon MPC and the tube alignment and standard quality assurance measurements were repeated. The time taken for each tube alignment method was estimated. The MPC method of tube alignment was found to be repeatable, insignificantly sensitive to phantom setup error and quick and simple to perform. The standard QA tests were generally insensitive to the tube alignment change, possibly because of the IsoCal correction. However, reduction in the magnitude of IsoCal correction and MPC kV imager offset was recorded after tube alignment. There was also apparent improvement in CBCT image uniformity. The MPC procedure is recommended for X-ray tube alignment.Entities:
Keywords: linac quality assurance; machine performance check (MPC)
Mesh:
Year: 2018 PMID: 30178521 PMCID: PMC6236821 DOI: 10.1002/acm2.12445
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Nearly 2 yrs’ data for MPC kV source offset in axial and tangential directions. Relevant maintenance events are also indicated.
MPC kV imager and source offset results over a 2 yr period presented between relevant maintenance events (mean ± 2 SD (mm))
| kV imager offset | kV source offset | ||
|---|---|---|---|
| Axial | Tangential | ||
| Initial to beam steer + IsoCal ( | 0.20 ± 0.10 | −0.61 ± 0.05 | 1.61 ± 0.18 |
| Beam steer + IsoCal to kVd isocenter cal ( | 0.22 ± 0.07 | −0.63 ± 0.05 | 1.48 ± 0.25 |
| kVd isocenter cal to X‐ray tube replacement ( | 0.35 ± 0.07 | −0.64 ± 0.08 | 1.49 ± 0.24 |
| X‐ray tube replacement to end ( | 0.17 ± 0.06 | 0.02 ± 0.13 | 1.06 ± 0.26 |
Figure 2Nearly 2 yr data for MPC kV imager offset. Relevant maintenance events are also indicated.
IsoCal verification results before and after X‐ray tube alignment
| Before | After | Difference | |
|---|---|---|---|
| In‐plane imager rotation kV (deg) | −0.02 | −0.018 | 0.002 |
| Max imager shift kV (mm) | 0.44 | 0.07 | −0.37 |
In‐house Winston Lutz results before and after X‐ray tube alignment (coordinates for patient head first and supine)
| Before | After | Difference | |
|---|---|---|---|
| 3D Isocenter offsets: CBCT vs treatment (mm) | |||
| Left–right | −0.02 | −0.16 | −0.14 |
| Anterior–posterior | 0.03 | 0.01 | −0.02 |
| Superior–inferior | −0.11 | −0.04 | 0.07 |
| 2D offsets: treatment field vs CBCT isocenter (mm) | |||
| Gantry maximum displacement | 0.40 | 0.38 | −0.02 |
| Gantry average displacement | 0.20 | 0.21 | 0.01 |
CBCT Catphan results before and after X‐ray tube alignment (mean ± 2 SD)
| Full fan | Half fan | |||
|---|---|---|---|---|
| Before | After | Before | After | |
| Image quality | ||||
| Low contrast (disks visible) | 1.0 ± 1.1 | 2.0 ± 1.1 | 6.0 ± 1.1 | 6.0 ± 1.1 |
| High contrast (line‐pair patterns discernible) | 7.0 ± 0.0 | 7.0 ± 0.0 | 5.0 ± 0.0 | 4.0 ± 0.0 |
| Uniformity (HU) | 17.0 ± 5.6 | 5.0 ± 5.6 | 10.0 ± 5.6 | 1.0 ± 5.6 |
| Hounsfield Units (HU) | ||||
| Air | −987 ± 2.7 | −992 ± 2.7 | −999 ± 2.7 | −999 ± 2.7 |
| Teflon | 983 ± 4.6 | 1017 ± 4.6 | 945 ± 4.6 | 957 ± 4.6 |
| Delrin | 350 ± 3.0 | 375 ± 3.0 | 350 ± 3.0 | 357 ± 3.0 |
| Acrylic | 117 ± 3.9 | 132 ± 3.9 | 118 ± 3.9 | 121 ± 3.9 |
| Polystyrene | −52 ± 2.0 | −41 ± 2.0 | −46 ± 2.0 | −42 ± 2.0 |
| LDPE | −109 ± 3.7 | −93 ± 3.7 | −105 ± 3.7 | −103 ± 3.7 |
| PMP | −192 ± 9.4 | −183 ± 9.4 | −195 ± 9.4 | −193 ± 9.4 |
| Spatial integrity | ||||
| Left–right (mm) | 50.1 ± 0.2 | 50.1 ± 0.2 | 50 ± 0.2 | 50 ± 0.2 |
| Anterior–posterior (mm) | 50 ± 0.2 | 49.8 ± 0.2 | 50.1 ± 0.2 | 49.9 ± 0.2 |
| Angle (deg) | 89.8 ± 0.2 | 90.2 ± 0.2 | 90 ± 0.2 | 90 ± 0.2 |
OBI geometric integrity and center pixel before and after X‐ray tube alignment
| Before | After | Difference | |
|---|---|---|---|
| Image scale | |||
| Left–right (mm) | 99.8 | 99.9 | 0.1 |
| Superior–inferior (mm) | 100 | 99.9 | −0.1 |
| Angle (deg) | 90.2 | 90 | −0.2 |
| Center pixel | |||
| Left–right (mm) | 0.1 | 0 | −0.1 |
| Superior–inferior (mm) | 0.1 | 0.1 | 0.0 |
Figure 3Images for the two‐plate method of X‐ray tube alignment performed before and after tube alignment and at both kV source head up and head down. Top Left = Before and head down, Top Right = Before and head up, Bottom Left = After and head down and Bottom Right = After and head up.
Figure 4Images for the wires on faceplate method of X‐ray tube alignment performed before and after tube alignment and at both kV source head up and head down. Top Left = Before and head down, Top Right = Before and head up, Bottom Left = After and head down and Bottom Right = After and head up.
Figure 5Screen capture of the MPC kV tube alignment procedure in the tangential tab.