Michael Pearson1, David Eaton1, Tony Greener1.
Abstract
Machine Performance Check (MPC) is an automated Quality Control (QC) tool that is integrated into the TrueBeam and Halcyon linear accelerators (Linacs), utilizing the imaging systems to verify the Linac beam and geometry. This work compares the concordance of daily MPC results with conventional QC tests over a 3-year period for eight Linacs in order to assess the sensitivity of MPC in detecting faults. The MPC output measurements were compared with the monthly ionization chamber measurements for 6 and 10 MV photon beams and 6, 9, 12, 16, and 18 MeV electron beams. All 6 MV Beam and Geometry (6MVBG) MPC test failures were analyzed to determine the failure rate and the number of true and false negative results, using the conventional QC record as the reference. The concordance between conventional QC test failures and MPC test failures was investigated. The mean agreement across 1933 MPC output and monthly comparison chamber measurements for all beam energies was 0.2%, with 97.8% within 1.5%, and a maximum difference of 2.9%. Of the 5000-6000 MPC individual test parameter results for the 6MVBG test, the highest failure rate was BeamOutputChange (0.5%), then BeamCenterShift (0.3%), and was ≤ 0.1% for the remaining parameters. There were 50 true negative and 27 false negative out of tolerance MPC results, with false negatives resolved by repeating MPC or by independent measurement. The analysis of conventional QC failures demonstrated that MPC detected all failures, except occasions when MPC reported output within tolerance, a result of the MPC-chamber response variation. The variation in MPC output versus chamber measurement indicates MPC is appropriate for daily output constancy but not for the measurement of absolute output. The comparison of the 6MVBG results and conventional records provides evidence that MPC is a sensitive method of performing beam and mechanical checks in a clinical setting.Entities:
Keywords: automated QC; linac quality control (QC); linear accelerator; machine performance check (MPC); quality assurance
Mesh:
Year: 2020 PMID: 32790139 PMCID: PMC7484877 DOI: 10.1002/acm2.12950
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
All MPC and comparable Conventional QC test parameters and tolerances.
| MPC Group | MPC test parameter | Tolerance | Conventional QC test | Test Method and Equipment | Tolerance |
|---|---|---|---|---|---|
| Beam | BeamCenterShift | 0.5 mm | – | – | – |
| BeamOutputChange | 2% | Ionization Chamber Output | Photons: Farmer, 5 cm deep, 95 cm SSD, 10 × 10 field, T41014 stationary water phantomb. Electrons: Roos/NACP, dmax, 100 cm SSD, 10 × 10 field, Bart’s Solid Water WTEc. | 2% | |
| BeamUniformityChange | 2% | Flatness and Symmetry | Octavius 1500 Ionization chamber arrayb 20 × 20 field. Photons: 5 cm deep, 95 cm SSD. 6/9 MeV: 12 mm, 12/16/18 MeV: 22 mm deep, 100 cm SSD. | 3% | |
| Collimation | CollimationRotationOffset | 0.5° | Collimator Scale | Marked dot parallel to projected crosshair on gantry rotation. | 0.5° |
| Jaws | JawX1 | 1 mm | Radiation field size | Ball bearing plate aligned to projected crosshair at 100 cm SSD. MV image acquired. Inspection of position of each jaw edge with respect to images of 1 mm and 2 mm ball bearings on plate. Alternate 10 × 10, 20 × 20 fields monthly. | 2 mm |
| JawX2 | 1 mm | Radiation field size | 2 mm | ||
| JawY1 | 2 mm | Radiation field size | 2 mm | ||
| JawY2 | 2 mm | Radiation field size | 2 mm | ||
| MLC | MLCLeafXXa | 1 mm | Picket Fence | Inspection of pickets in portal image. | 1 mm |
| MLCMaxOffsetA | 1 mm | Picket Fence | 1 mm | ||
| MLCMaxOffsetB | 1 mm | Picket Fence | 1 mm | ||
| MLCMeanOffsetA | 1 mm | Picket Fence | 1 mm | ||
| MLCMeanOffsetB | 1 mm | Picket Fence | 1 mm | ||
| Couch | CouchLat | 0.7 mm | Couch lateral scales | Move couch known distances with ruler. | 1 mm |
| CouchLng | 0.7 mm | Couch longitudinal scales | Move couch known distances with ruler. | 1 mm | |
| CouchRtn | 0.4° | Couch rotation scales | Couch 0°: Marked dot parallel to projected crosshair for couch long movements. Other angles: align plate to crosshair on couch rotation. | 0.5° | |
| CouchVrt | 1.2 mm | Couch vertical scales | Move couch known distances with ruler. | 1 mm | |
| RotationInducedCouchShift | 0.75 mm | – | – | – | |
| Gantry | GantryAbsolute | 0.30 | Gantry rotation scales | Gantry leveled with spirit level abutting the Interface Mount at each cardinal angle. | 0.5° |
| GantryRelative | 0.30 | Gantry rotation scales | 0.5° | ||
| Isocenter | IsoCenterKVOffset | 0.5 mm | Isoverification kV center | Varian Isoverification | 0.5 mm |
| IsoCenterMVOffset | 0.5 mm | Isoverification MV center | 0.5 mm | ||
| IsoCenterSize | 0.5 mm | – | – | – |
“‐“ indicates that the equivalent conventional QC test was not performed. aEach MLC leaf (XX) is reported individually, for example, leaf 11 is MLCLeaf11. bPTW, Freiburg, Germany. cSt Bartholomew's Hospital, London, UK.
Fig. 1The % difference in 6 MV MPC and ion chamber outputs measured monthly on all eight Linacs, February 2017–November 2019.
The % difference between MPC and ion chamber outputs measured monthly on all eight Linacs.
| Beam | Number of measurements | Mean % difference | Min % difference | Max % difference | % within 1% | % within 1.5% | % within 2% | SD |
|---|---|---|---|---|---|---|---|---|
| 6 MV | 306 | 0.3 | −1.2 | 2.0 | 87.3 | 98.7 | 100.0 | 0.6 |
| 10 MV | 308 | 0.3 | −1.5 | 2.0 | 89.9 | 97.7 | 99.7 | 0.5 |
| 6 MeV | 272 | 0.1 | −1.7 | 2.7 | 85.3 | 95.2 | 98.5 | 0.7 |
| 9 MeV | 265 | 0.1 | −1.5 | 1.8 | 90.2 | 98.1 | 100.0 | 0.6 |
| 12 MeV | 267 | 0.0 | −2.9 | 1.6 | 91.4 | 98.1 | 99.6 | 0.6 |
| 16 MeV | 259 | 0.1 | −1.4 | 1.9 | 91.1 | 98.5 | 100.0 | 0.6 |
| 18 MeV | 256 | 0.1 | −1.6 | 1.6 | 89.8 | 98.4 | 100.0 | 0.6 |
| Total | Mean | Min | Max | Total | Total | Total | ||
| 1933 | 0.2 | −2.9 | 2.7 | 89.2 | 97.8 | 99.7 |
Fig. 2Distribution of MPC measurement of 6 MV output, measured daily on all eight Linacs, February 2017–November 2019.
Fig. 3The % difference in 6 MV MPC and ion chamber outputs: measured monthly on all eight Linacs, February 2017–November 2019, corrected for re‐baselining of MPC. The % difference is normalized to zero for the first measurement on each Linac. % difference in 6 MV absorbed dose to water calibration factor: for Farmer chambers 403, 405, 424 determined from annual cross calibrations with the secondary standard. The % difference is normalized to zero for the first cross calibration in November 2016.
Summary of all 6 MV Beam and Geometry MPC failures, November 2016–July 2019. Failures are classed FN (False Negative) or TN (True Negative). Failure rate is the total number of out of tolerance MPC results/total number of MPC measurements.
| MPC Failure |
MPC Tolerance | Linac | Date | Number of MPC results> tolerance | Max Value of MPC parameter | FN/TN | Cause of MPC Failure | Total MPC measurements | Total number> tolerance | Failure Rate % | Total TN | Total FN |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| BeamOutputChange | 2% | Q1 | 12/11/2017 | 1 | 2.04% | FN | Repeated MPC and in tolerance (repeat 1.91%) | 6230 | 33 | 0.5 | 18 | 15 |
| 12/27/2017 | 1 | 2.15% | TN | Output drift | ||||||||
| Q2 | 10/11/2018 | 2 | 2.08% | FN | DailyQA3 — MPC difference | |||||||
| 10/19/2018 | 2 | −2.35% | TN | Output drift | ||||||||
| 10/23/2018 | 3 | −2.06% | FN | Farmer — MPC difference | ||||||||
| 10/30/2018 | 1 | 2.78% | TN | Output drift | ||||||||
| 12/27/2018 | 1 | −2.88% | TN | Output drift | ||||||||
| G1 | 12/31/2018 | 2 | −2.16% | FN | Farmer — MPC difference | |||||||
| 01/02/2019 | 1 | −2.82% | TN | Output drift | ||||||||
| 01/29/2019 | 4 | 2.10% | TN | Output drift | ||||||||
| 02/01/2019 | 1 | 2.52% | TN | Output drift | ||||||||
| 02/14/2019 | 1 | −3.56% | TN | Output drift | ||||||||
| 02/25/2019 | 1 | −2.25% | FN | DailyQA3 — MPC difference | ||||||||
| G2 | 12/18/2017 | 1 | −2.73% | TN | Output drift | |||||||
| 09/25/2018 | 1 | −2.11% | TN | Output drift | ||||||||
| 10/10/2018 | 1 | −2.05% | FN | Farmer — MPC difference | ||||||||
| 01/02/2019 | 2 | −2.37% | TN | Output drift | ||||||||
| 01/14/2019 | 1 | 2.60% | TN | Output drift | ||||||||
| 02/25/2019 | 1 | −2.12% | TN | Output drift | ||||||||
| G3 | 08/10/2017 | 1 | −31.61% | FN | Repeated MPC and in tolerance (repeat −0.37%) | |||||||
| G4 | 07/06/2017 | 4 | 2.56% | FN | Farmer — MPC difference | |||||||
| BeamUniformityChange | 2% | G3 | 08/10/2017 | 1 | 2.71% | FN | Repeated MPC and in tolerance (repeat 1.08%) | 6249 | 6 | 0.1 | 5 | 1 |
| G4 | 06/12/2019 | 5 | 4.42% | TN | Mistake in retuning symmetry | |||||||
| BeamCenterShift | 0.5 mm | G2 | 03/22/2019 | 9 | 0.7 mm | TN | Steering error when ion chamber replaced | 6246 | 17 | 0.3 | 11 | 6 |
| G4 | 06/13/2019 | 2 | 0.53 mm | TN | Mistake in retuning symmetry | |||||||
| G5 | 07/31/2018 | 6 | 0.9 mm | FN | MPC Rebaselined, passed Conventional Radiation field size test | |||||||
| CouchLat | 0.7 mm | G1 | 05/16/2018 | 1 | 1.09 mm | FN | Repeated MPC and in tolerance (repeat −0.01 mm) | 5012 | 2 | 0.0 | 0 | 2 |
| G5 | 11/29/2018 | 1 | −0.77 mm | FN | Repeated MPC and in tolerance (repeat 0.04 mm) | |||||||
| RotationInducedCouchShift | 0.75 mm | G1 | 05/16/2018 | 1 | 1.16 mm | FN | Repeated MPC and in tolerance (repeat 0.12 mm) | 5012 | 2 | 0.0 | 0 | 2 |
| G5 | 11/29/2018 | 1 | 0.92 mm | FN | Repeated MPC and in tolerance (repeat 0.08 mm) | |||||||
| MLCMaxOffset | 1 mm | G2 | 07/10/2017 | 3 | 1.67 mm | TN | Resolved by re‐initializing MLC | 5944 | 7 | 0.1 | 7 | 0 |
| G5 | 09/27/2018 | 4 | 1.05 mm | TN | Resolved by re‐initializing MLC | |||||||
| IsoCenterKVOffset | 0.5 mm | Q1 | 06/27/2019 | 1 | 0.59 mm | TN | X‐ray tube realigned | 5147 | 7 | 0.1 | 6 | 1 |
| Q2 | 11/08/2017 | 1 | 0.55 mm | FN | Repeated MPC and in tolerance (repeat 0.2 mm) | |||||||
| G2 | 06/07/2019 | 2 | 0.69 mm | TN | Post steering adjustment, Isocal performed and corrected | |||||||
| G5 | 01/31/2019 | 1 | 0.57 mm | TN | X‐ray tube realigned | |||||||
| G6 | 05/18/2018 | 2 | 1.07 mm | TN | X‐ray tube realigned | |||||||
| IsoCenterMVOffset | 0.5 mm | G2 | 03/22/2019 | 3 | 0.92 mm | TN | Post steering adjustment, Isocal performed and corrected | 5147 | 3 | 0.1 | 3 | 0 |
All Conventional QC test failures equating to parameters tested by MPC, November 2016–July 2019.
| Conventional QC Test | Tolerance | Linac | Date | Cause of QC Test Failure | Detected by MPC |
|---|---|---|---|---|---|
| Ionization chamber Output | 2% | Q1 | 27/12/2017 | Output drift | Yes |
| Q2 | 28/04/2018 | Output drift | Noa | ||
| Q2 | 30/10/2018 | Output drift | Yes | ||
| Q2 | 27/12/2018 | Output drift | Yes | ||
| G1 | 16/03/2018 | Output drift | Noa | ||
| G1 | 09/05/2019 | Output drift | Noa | ||
| G2 | 03/01/2019 | Output drift | Yes | ||
| G2 | 14/01/2019 | Output drift | Yes | ||
| G4 | 23/11/2017 | Output drift | Noa | ||
| G4 | 09/08/2018 | Output drift | Noa | ||
| G5 | 28/05/2019 | Output drift | Noa | ||
| Radiation Field Size | 2 mm | G2 | 17/03/2017 | QC test passes on following month | Nob |
| 03/05/2019 | During period with Steering error | Yes | |||
| G5 | 27/03/2017 | QC test passes on following month | Nob | ||
| G6 | 20/02/2017 | QC test passes on following month | Nob | ||
| 25/04/2017 | QC test passes on following month | Nob | |||
| Isoverification kV center | 0.5 mm | G2 | 07/06/2019 | Replacement ion chamber, Steering adjustment | Yes |
| Isoverification MV center | 0.5 mm | G2 | 07/06/2019 | Replacement ion chamber, Steering adjustment | Yes |
| 6 MV Symmetry PTW Array | 3% | G4 | 13/06/2019 | Mistake in retuning symmetry | Yes |
MPC detects all the failures, with the exceptions of: aOccasions where the ion chamber measurement is out of tolerance, but MPC reports in tolerance due to the MPC–ion chamber output tolerance of 1.5%. bOccasions where it is probable that the conventional QC falsely reported the incorrect field size (likely due to misalignment of the phantom), since the repeat conventional QC measurement is in tolerance.
Estimates of sensitivity, specificity, and NPV for the 6 MV Beam and Geometry test groups.
| MPC Test Group | Sensitivity | Specificity | NPV |
|---|---|---|---|
| 6 MV Beam | 93% | 54% | 61% |
| 6 MV Geometry | 98% | 100% | 76% |