M J Fokkert1, P R van Dijk2, M A Edens3, S Abbes1, D de Jong3, R J Slingerland1, H J G Bilo4.
Abstract
OBJECTIVE: To evaluate the performance of the FreeStyle Libre Flash continuous glucose monitoring (FSL-CGM) system against established central laboratory methods. RESEARCH DESIGN AND METHODS: 20 subjects (8 type 1 diabetes mellitus, 12 type 2 diabetes mellitus) were analyzed. FSL-CGM sensor measurements (inserted in arm and abdomen) were compared with capillary blood glucose results analyzed with StatStrip as semigold standard. The glucose response after a standardized oral glucose load was measured by FSL-CGM and capillary samples analyzed by perchloric acid hexokinase (PCA-HK) method, StatStrip and FSL test strip (FSLC), and a commonly used CGM system (iPro2).Entities:
Keywords: Blood Glucose Self-Monitoring; Continuous Glucose Monitoring; Endocrinology Diabetes; Glucose Control
Year: 2017 PMID: 28243449 PMCID: PMC5316912 DOI: 10.1136/bmjdrc-2016-000320
Source DB: PubMed Journal: BMJ Open Diabetes Res Care ISSN: 2052-4897
Baseline characteristics
| Type 1 diabetes | Type 2 diabetes | |
|---|---|---|
| N=8 | N=12 | |
| Age (years) | 63 [24–74] | 56 [39–65] |
| Female sex (%) | 3 (38) | 3 (25) |
| BMI (kg/m2) | 23.2 [21.6–24.5] | 28.3 [26.4–30.7] |
| Fat percentage (%) | 27.1 [23.6–33.0] | 28.9 [25–35.1] |
| Insulin therapy (n): CSII/MDI/BD | 4/4/0 | 3/8/1 |
| Oral blood glucose-lowering agents (%) | ||
| Metformin | – | 10 |
| SU | – | 1 |
| Other | – | 0 |
| HbA1c (mmol/mol) | 59 [46–70] | 55 [42–82] |
| Presence of microvascular complications | ||
| Nephropathy (n) | 0 | 1 |
| Neuropathy (n) | 2 | 4 |
| Retinopathy (n) | 2 | 1 |
| Diabetes duration (years) | 14 [7–24] | 13 [4–25] |
Data are presented as n (%), mean (SD) or median [range].
BD, twice daily mix insulin; BMI, body mass index; CSII, continuous subcutaneous insulin infusion; HbA1c, glycated hemoglobin; MDI, multiple daily injections; SU, sulfonylurea derivate.
Figure 1Clarke Error Grids (ISO15197; 2013) of StatStrip and coincident FSLTab-CGM arm (upper left panel), FSL-CGM abdomen (lower left panel) and FSLC readings (right panel) in the combined population of type 1 and 2 diabetes. Number and percentages of paired readings within error grid zones A, B, C D and E are presented. FSL-CGM, FreeStyle Libre Flash continuous glucose monitoring.
Glucose measurement data at various concentrations (mean+SD; and median with 25–75 IQRs; paired t-tests)
| FSLA (arm) | FSLC (capillary) | StatStrip | FSLA vs FSLC | FSLA vs StatStrip | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| N | Mean (SD) | Median (25–75) | N | Mean (SD) | Median (25–75) | N | Mean (SD) | Median (25–75) | MAD | MARD | p Value | MAD | MARD | p Value | |
| ≤70 | 56 | 60 (10) | 65 (62, 69) | 45 | 68 (16) | 70 (59, 75) | 56 | 70 (16) | 69 (62, 79) | 11 | 19 | <0.0001 | 13 | 24 | 0.0059 |
| 71–80 | 39 | 76 (2) | 76 (74, 78) | 39 | 73 (26) | 78 (69, 85) | 39 | 83 (16) | 83 (74, 94) | 10 | 13 | 0.0071 | 13 | 17 | 0.0264 |
| 81–90 | 53 | 86 (3) | 85 (83, 88) | 53 | 85 (23) | 90 (83, 96) | 53 | 95 (12) | 96 (87, 101) | 9 | 11 | <0.0001 | 12 | 14 | 0.0067 |
| 91–100 | 48 | 97 (3) | 97 (94, 99) | 48 | 88 (33) | 97 (87, 111) | 48 | 106 (15) | 105 (96, 117) | 11 | 12 | <0.0001 | 14 | 14 | 0.0001 |
| 101–110 | 44 | 107 (3) | 106 (105, 110) | 39 | 113 (24) | 106 (101, 121) | 44 | 117 (12) | 115 (109, 126) | 13 | 13 | <0.0001 | 13 | 12 | 0.1083 |
| 111–125 | 77 | 117 (4) | 115 (112, 121) | 77 | 116 (39) | 126 (112, 133) | 77 | 127 (15) | 126 (117, 136) | 15 | 13 | <0.0001 | 15 | 12 | 0.2237 |
| 126–150 | 124 | 138 (7) | 139 (132, 144) | 124 | 123 (49) | 137 (119, 149) | 124 | 143 (18) | 142 (130, 155) | 13 | 9 | 0.0106 | 14 | 10 | 0.0032 |
| 151–180 | 105 | 166 (8) | 166 (159, 173) | 105 | 141 (61) | 160 (144, 177) | 105 | 169 (23) | 169 (156, 184) | 16 | 9 | 0.1751 | 19 | 10 | <0.0001 |
| >180 | 169 | 235 (49) | 220 (196, 259) | 199 | 178 (92) | 202 (159, 236) | 199 | 215 (45) | 211 (186, 239) | 23 | 10 | <0.0001 | 25 | 10 | 0.1298 |
MAD, mean absolute difference (mg/dL); MARD, mean absolute relative difference (%).
Figure 2The 75 g glucose loading test. Comparison of FSL-CGM versus PCA-HK, StatStrip, FSLC test strip, and iPro2; and PCA-HK versus FSLC test strip, and iPro2. Glucose concentrations (mg/dL; median, interquartile Q1 and Q3) at time points 0, 15, 30, 45, 60, 75 and 90 min in the combined population of type 1 and 2 diabetes. *p Value ≤0.05; **p Value ≤0.01. FSL-CGM, FreeStyle Libre Flash continuous glucose monitoring; PCA-HK, perchloric acid hexokinase.