Estimating plasma glucose with the FreeStyle Libre Pro continuous glucose monitor during oral glucose tolerance tests in youth without diabetes.

Nejla Ghane1,2, Miranda M Broadney1,3, Elisabeth K Davis1, Robert W Trenschel1, Shavonne M Collins4, Sheila M Brady1,3, Jack A Yanovski1.   

Abstract

Few studies have assessed the accuracy of the FreeStyle Libre Pro (FLP) continuous glucose monitor for estimating plasma glucose (PG) in non-diabetic children.
OBJECTIVE: Determine the accuracy of FLP compared to PG during OGTT in healthy children.
SUBJECTS: Children (7-11.99 years) with healthy weight and overweight/obesity (n = 33; 52% male).
METHODS: Participants wore the FLP before and during a 2-hour OGTT; PG was measured at 30 minutes intervals. Potential systematic- and magnitude-related biases for FLP vs PG were examined.
RESULTS: FLP 15-minute averages and PG were correlated at most timepoints during OGTT (r2 = 0.35-0.69, P's < .001 for time point 30-120 minutes) and for PG area under the curve (AUC) (r2 = 0.65, P < .0001). There were no systematic biases as assessed by Bland-Altman analyses for FLP AUC or for FLP at each OGTT timepoint. However, for fasting glucose, a significant magnitude bias was noted (r2 = 0.38, P < .001), such that lower PG was underestimated, and higher PG was overestimated by FLP readings; further, there was poor correlation between fasting PG and FLP (r2 = 0.06, P = .22). BMIz was also associated with FLP accuracy: FLP overestimated PG in children with low BMIz and underestimated PG in those with overweight/obesity for OGTT AUC and OGTT PG at baseline, 60, and 120 minutes (all P's ≤ .015). No adverse events occurred with FLP.
CONCLUSIONS: Among children without diabetes, the FLP was well tolerated and correlated with post-OGTT glucose, but had magnitude bias affecting fasting glucose and appeared to underestimate plasma glucose in those with overweight/obesity. These results suggest potential limitations for the utility of the FLP for research.
© 2019 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  biosensing techniques; blood glucose self-monitoring; body mass index; child; glucose tolerance test

Mesh:

Substances:

Year:  2019        PMID: 31433542      PMCID: PMC6821586          DOI: 10.1111/pedi.12910

Source DB:  PubMed          Journal:  Pediatr Diabetes        ISSN: 1399-543X            Impact factor:   4.866


  30 in total

1.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

2.  Evaluation of a real-time blood glucose monitor in children with diabetic ketoacidosis.

Authors:  J A Coss-Bu; L S Jefferson; S Stone-McCord; C N Ou; C Watrin; R Sachdeva; K C Copeland
Journal:  Diabetes Res Clin Pract       Date:  1999-06       Impact factor: 5.602

3.  Accuracy and Precision of Measured Blood Sugar Values by Three Glucometers Compared to the Standard Technique.

Authors:  Shadi Khakpour Kermani; Alireza Khatony; Rostam Jalali; Mansur Rezaei; Alireza Abdi
Journal:  J Clin Diagn Res       Date:  2017-04-01

4.  Characterizing glucose exposure for individuals with normal glucose tolerance using continuous glucose monitoring and ambulatory glucose profile analysis.

Authors:  Roger S Mazze; Ellie Strock; David Wesley; Sarah Borgman; Blaine Morgan; Richard Bergenstal; Robert Cuddihy
Journal:  Diabetes Technol Ther       Date:  2008-06       Impact factor: 6.118

5.  Continuous glucose monitoring system in free-living healthy subjects: results from a pilot study.

Authors:  Giuseppe Derosa; Sibilla A T Salvadeo; Roberto Mereu; Angela D'Angelo; Leonardina Ciccarelli; Mario N Piccinni; Ilaria Ferrari; Alessia Gravina; Pamela Maffioli; Carmine Tinelli
Journal:  Diabetes Technol Ther       Date:  2009-03       Impact factor: 6.118

6.  Evaluation of the FreeStyle® Libre Flash Glucose Monitoring System in Children and Adolescents with Type 1 Diabetes.

Authors:  Guy G Massa; Inge Gys; Anniek Op 't Eyndt; Esmiralda Bevilacqua; Anne Wijnands; Peter Declercq; Renate Zeevaert
Journal:  Horm Res Paediatr       Date:  2018-03-27       Impact factor: 2.852

7.  The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System.

Authors:  Timothy Bailey; Bruce W Bode; Mark P Christiansen; Leslie J Klaff; Shridhara Alva
Journal:  Diabetes Technol Ther       Date:  2015-07-14       Impact factor: 6.118

8.  Performance of the FreeStyle Libre Flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus.

Authors:  M J Fokkert; P R van Dijk; M A Edens; S Abbes; D de Jong; R J Slingerland; H J G Bilo
Journal:  BMJ Open Diabetes Res Care       Date:  2017-02-17

9.  The use and efficacy of continuous glucose monitoring in type 1 diabetes treated with insulin pump therapy: a randomised controlled trial.

Authors:  T Battelino; I Conget; B Olsen; I Schütz-Fuhrmann; E Hommel; R Hoogma; U Schierloh; N Sulli; J Bolinder
Journal:  Diabetologia       Date:  2012-09-11       Impact factor: 10.122

10.  Reference values for continuous glucose monitoring in Chinese subjects.

Authors:  Jian Zhou; Hong Li; Xingwu Ran; Wenying Yang; Qiang Li; Yongde Peng; Yanbing Li; Xin Gao; Xiaojun Luan; Weiqing Wang; Weiping Jia
Journal:  Diabetes Care       Date:  2009-04-23       Impact factor: 17.152

View more
  2 in total

Review 1.  Use of continuous glucose monitoring in obesity research: A scoping review.

Authors:  Elizabeth Hegedus; Sarah-Jeanne Salvy; Choo Phei Wee; Monica Naguib; Jennifer K Raymond; D Steven Fox; Alaina P Vidmar
Journal:  Obes Res Clin Pract       Date:  2021-09-02       Impact factor: 5.214

2.  The Accuracy and Precision of the Continuously Stored Data from Flash Glucose Monitoring System in Type 2 Diabetes Patients during Standard Meal Tolerance Test.

Authors:  Rengna Yan; Huiqin Li; Xiaocen Kong; Xiaofang Zhai; Maoyuan Chen; Yixuan Sun; Lei Ye; Xiaofei Su; Jianhua Ma
Journal:  Int J Endocrinol       Date:  2020-01-04       Impact factor: 3.257

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.