Hood Thabit1, Lalantha Leelarathna1, Malgorzata E Wilinska1, Daniella Elleri1, Janet M Allen1, Alexandra Lubina-Solomon2, Emma Walkinshaw2, Marietta Stadler3, Pratik Choudhary3, Julia K Mader4, Sibylle Dellweg5, Carsten Benesch5, Thomas R Pieber4, Sabine Arnolds5, Simon R Heller2, Stephanie A Amiel3, David Dunger1, Mark L Evans1, Roman Hovorka1.
Abstract
OBJECTIVES: Closed-loop (CL) systems modulate insulin delivery based on glucose levels measured by a continuous glucose monitor (CGM). Accuracy of the CGM affects CL performance and safety. We evaluated the accuracy of the Freestyle Navigator(®) II CGM (Abbott Diabetes Care, Alameda, CA) during three unsupervised, randomized, open-label, crossover home CL studies.Entities:
Mesh:
Substances:
Year: 2015 PMID: 26241693 PMCID: PMC4649721 DOI: 10.1089/dia.2015.0062
Source DB: PubMed Journal: Diabetes Technol Ther ISSN: 1520-9156 Impact factor: 6.118
Baseline Demographics of Participants from the Three Closed-Loop Home Studies
| Age (years) | 39.5±11.6 | 15.4±2.0 |
| Gender (male/female) | 23/18 | 10/6 |
| Body mass index (kg/m2) | 26.1±3.4 | 21.7±2.2 |
| Hemoglobin A1c (%) | 7.9±0.8 | 8.1±0.9 |
| Duration of diabetes (years) | 24.8±11.6 | 7.2±4.3 |
Sensor Accuracy for the Whole Glucose Range and Stratified According to Capillary Glucose During the Three Closed-Loop Home Studies
| Number of glucose sensor–capillary pairs | 10,647 | 6,223 | 3,410 | 1,014 |
| Mean capillary glucose (mmol/L) | 8.6±3.9 | 6.9±1.7 | 13.3±2.8 | 3.2±0.5 |
| Mean AD | 1.0±1.0 | 0.9±0.8 | 1.3±1.2 | 0.9±0.8 |
| Median AD | 0.8 (0.4, 1.4) | 0.7 (0.3, 1.2) | 1.0 (0.4, 1.8) | 0.7 (0.4, 1.2) |
| Mean ARD | 14.2±15.5 | 13.9±13.3 | 9.8±8.8 | 30.6±28.7 |
| Median ARD (%) | 10.0 (4.5, 18.4) | 10.4 (4.7, 18.6) | 7.8 (3.4, 13.7) | 22.8 (11.2, 41.2) |
| Bias | 0.04 (1.44) | 0.08 (1.3) | −0.25 (1.7) | 0.7 (0.9) |
| ISO criteria (%)[ | 70.0 | 67.5 | 78.7 | 55.7 |
Data are mean±SD or median (interquartile range) values as indicated.
International Organization for Standardization (ISO) 15197:2013 15/15% limits.
AD, absolute difference; ARD, absolute relative difference.

Bland–Altman plot of sensor and capillary glucose levels. The solid black line represents the mean difference between the sensor and capillary glucose values; the dashed lines indicate 1.96×SD of the difference. CGM, continuous glucose monitor.

Clarke error grid of sensor and capillary glucose levels. CGM, continuous glucose monitor.