Stefan Pleus1, Ulrike Kamecke1, Manuela Link1, Cornelia Haug1, Guido Freckmann1.
Abstract
The flash glucose monitoring system FreeStyle Libre (Abbott Diabetes Care Ltd., Witney, UK) measures interstitial glucose concentrations and continuously stores measurement values every 15 minutes. To obtain a current glucose reading, users have to scan the sensor with the reader. In a clinical trial, 5% of the scanned data showed relative differences of more than ±10% compared with continuously stored data points (median -0.5%). Such differences might impact results of studies using this system. It should be indicated whether scanned or continuously stored data were used for analyses. Health care professionals might have to differentiate between data reports from clinical software and the scanned data their patients are provided with. Additional information on these differences and their potential impact on therapeutic decisions would be helpful.Entities:
Keywords: FreeStyle Libre; continuous glucose monitoring; flash glucose monitoring; interstitial glucose
Mesh:
Substances:
Year: 2017 PMID: 28990438 PMCID: PMC5851226 DOI: 10.1177/1932296817733095