| Literature DB >> 25132320 |
J Kropff1, D Bruttomesso, W Doll, A Farret, S Galasso, Y M Luijf, J K Mader, J Place, F Boscari, T R Pieber, E Renard, J H DeVries.
Abstract
AIMS: To assess the accuracy and reliability of the two most widely used continuous glucose monitoring (CGM) systems.Entities:
Keywords: CSII; type 1 diabetes
Mesh:
Substances:
Year: 2014 PMID: 25132320 PMCID: PMC4409843 DOI: 10.1111/dom.12378
Source DB: PubMed Journal: Diabetes Obes Metab ISSN: 1462-8902 Impact factor: 6.577
Glucose sensor accuracy assessed for the Dexcom G4 Platinum and Medtronic Paradigm Veo Enlite systems in patients with type 1 diabetes under clinical research centre and home conditions
| Variable | DG4P | ENL | p | CI Δ/IQR Δ |
|---|---|---|---|---|
| Number of sensors used (n) | 24 | 26 | — | — |
| Number of CGM–reference pairs: CRC (n) | 532 | 530 | — | — |
| Number of CGM–reference pairs: home (n) | 987 | 839 | — | —— |
| Number of CGM–reference pairs <3.9 mmol/l: CRC (n) | 117 | 117 | — | — |
| Number of CGM–reference pairs <3.9 mmol/l: home (n) | 56 | 47 | — | — |
| Available CGM readings (% of maximum) | 99.2 | 98.1 | <0.0001 | 1.1–1.2 |
| MARD CRC (%) | 13.6 | 16.6 | 0.0002 | 1.7–4.3 |
| MARD home (%) | 12.2 | 19.9 | <0.0001 | 5.8–8.7 |
| MAD CRC (mmol/l) | 0.73 | 0.91 | <0.0001 | 0.1–0.3 |
| MAD home (mmol/l) | 1.03 | 1.68 | <0.0001 | 0.5–0.7 |
| MARD <3.9 mmol/l CRC (%) | 17.6 | 24.6 | 0.0058 | 3.1–10.7 |
| MARD <3.9 mmol/l home (%) | 21.2 | 36.5 | 0.0142 | 3.6–25.4 |
| MARD 3.9–10 mmol/l CRC (%) | 13.0 | 14.2 | 0.08 | 0.03–2.6 |
| MARD 3.9–10 mmol/l home (%) | 11.7 | 19.5 | <0.0001 | 5.5–9.1 |
| MARD ≥10 mmol/l CRC (%) | 6.2 | 17.1 | <0.0001 | 6.9–14.8 |
| MARD ≥10 mmol/l home (%) | 11.6 | 18.0 | <0.0001 | 3.5–8.1 |
| Median ARD CRC (%) | 10.9 | 13.9 | 0.0002 | 17.7 |
| Median ARD home (%) | 8.9 | 14.4 | <0.0001 | 18.8 |
| Median ARD <3.9 mmol/l CRC (%) | 16.8 | 22.0 | 0.0058 | 23.5 |
| Median ARD <3.9 mmol/l home (%) | 12.7 | 23.1 | 0.0142 | 37.2 |
| ARD ≤20% CRC (%) | 77.1 | 67.2 | 0.0003 | 4.5–15.3 |
| ARD ≤20% home (%) | 82.2 | 63.5 | <0.0001 | 14.6–22.7 |
| ARD ≤20% or 1.1 mmol/l CRC (%) | 86.1 | 77.5 | 0.0003 | 3.9–13.2 |
| ARD ≤20% or 1.1 mmol/l home (%) | 83.9 | 65.2 | <0.0001 | 14.7–22.7 |
| ARD ≤15% or 0.8 mmol/l CRC (%) | 75.2 | 60.8 | <0.0001 | 8.8–20.1 |
| ARD ≤15% or 0.8 mmol/l home (%) | 74.0 | 53.6 | <0.0001 | 15.9–24.7 |
| CGM >3.9 mmol/l, when reference <3.1 mmol/lhome (%) | 33.3 | 43.75 | 0.5327 | −22.3 to 43.2 |
| CGM <3.1 mmol/l, when reference >3.9 mmol/l home (%) | 0.5 | 1.3 | 0.1062 | −0.3 to 1.5 |
| CGM >3.9 mmol/l, when reference <3.1 mmol/l CRC (%) | 8 | 28 | 0.0657 | |
| CGM <3.1 mmol/l, when reference >3.9 mmol/l CRC (%) | 0 | 0.72 | 0.0820 | |
| CEG zone A CRC (%) | 79.9 | 72.3 | 0.0036 | 2.5–12.8 |
| CEG zone A home (%) | 83.0 | 64.6 | <0.0001 | 14.4–22.4 |
Available CGM readings = percentage of calculated maximum number of CGM readings. YSI = reference value measured with the YSI blood glucose analyzer. Self monitoring of blood glucose = reference value measured with finger prick. 20%/1.1 mmol/l = below 4.4 mmol/l (80 mg/dl) reference glucose, deviation is defined as absolute deviation of ≤1.1 mmol/l (20 mg/dl), above 4.4 mmol/l (80 mg/dl) reference glucose, deviation is defined as ≤20%. 15%/0.8 mmol/l = below 5.6 mmol/l (100 mg/dl) reference glucose, deviation is defined as absolute deviation of ≤0.8 mmol/l (15 mg/dl), above 5.6 mmol/l (100 mg/dl) reference glucose, deviation is defined as ≤15% ARD (ISO 15197:2013). CGM >3.9 mmol/l when reference <3.1 mmol/l home = percentage of CGM values above 3.9 mmol/l (70 mg/dl) when reference is below 3.1 mmol/l (55 mg/dl) (a measure for false negative sensor alarms) 7. CGM <3.1 mmol/l when reference >3.9 mmol/l home = percentage of CGM values below 3.1 mmol/l (55 mg/dl) when reference is above 3.9 mmol/l (70 mg/dl) (a measure for false-positive sensor alarms). ARD, absolute relative difference; CGM, continuous glucose monitoring; CRC, clinical research centre; CEG, Clarke error grid; CI Δ, 95% confidence interval of difference; DG4P, Dexcom G4 Platinum; ENL, Medtronic Paradigm Veo Enlite; MARD, mean absolute relative difference; IQR Δ, interquartile range of difference, given for median ARD; MAD, mean absolute difference.
Confidence interval of difference was not calculated because of too few cases per cell, respectively 2 and 0 for DG4P.
Figure 1Bland–Altman plots for the Dexcom®G4 Platinum (DG4P) and Medtronic Paradigm Veo Enlite system (ENL). Panel A: clinical research centre (CRC) phase and panel B: home phase. The x-axis represents the average reference and continuous glucose monitoring (CGM) glucose measurements, the y-axis represents the difference (CGM – reference) versus the average of values measured by CGM expressed as a percentage. The long dashed line is drawn at the mean difference; dotted lines are drawn at the mean difference ± 1.96 times the standard deviation of the mean difference.
Figure 2Clarke error grid analysis combined for the Dexcom®G4 Platinum (DG4P; grey dots) and the Medtronic Paradigm Veo Enlite (ENL; black dots) during the clinical research centre (CRC) phase (A) and the home phase (B). CRC phase: based on a total of 532 DG4P – YSI pairs, the DG4P had 79.9% of values in zone A, 15.0% in zone B, 0% in zone C, 5.1% in zone D and 0% in zone E. Based on a total of 530 ENL–YSI (reference value measured with the YSI blood glucose analyser) pairs, the ENL had 72.3% of values in zone A, 20.0% in zone B, 0% in zone C, 7.7% in zone D and 0% in zone E. Home phase: based on a total of 987 DG4P – self-monitoring of blood glucose (SMBG) pairs, the DG4P had 83.0% of values in zone A, 15.0% in zone B, 0.3% in zone C, 1.6% in zone D and 0.1% in zone E. Based on a total of 839 ENL – SMBG pairs, the ENL had 64.6% of values in zone A, 29.8% in zone B, 1.6% in zone C, 3.6% in zone D and 0.5% in zone E.
Figure 3(A) Cumulative distribution of individual continuous glucose monitoring (CGM) performance (in the clinical research centre). (B) Cumulative distribution of individual CGM performance (at home). DG4P, Dexcom®G4 Platinum system; ENL, Medtronic Paradigm Veo Enlite system; MARD, mean absolute relative difference.
Figure 4Average continuous glucose monitoring accuracy per day (mean absolute relative difference (MARD) ± 95% confidence interval) for the Dexcom®G4 Platinum (DG4P) and the Medtronic Paradigm Veo Enlite (ENL) systems.