| Literature DB >> 28210644 |
Steven Struhl1, Theodore S Wolfson1, Frederick Kummer2.
Abstract
BACKGROUND: Although numerous suture-button fixation techniques for acromioclavicular (AC) joint reconstruction have been validated with biomechanical testing in the superior direction, clinical reports continue to demonstrate high rates of construct slippage and breakage.Entities:
Keywords: Endobutton; acromioclavicular joint; anatomic reconstruction; biomechanics; coracoclavicular ligament
Year: 2016 PMID: 28210644 PMCID: PMC5298557 DOI: 10.1177/2325967116674668
Source DB: PubMed Journal: Orthop J Sports Med ISSN: 2325-9671
Cadaver Demographics and Specimen Characteristics
| Specimen | Right | Left | DE Size, mm | Age, y | Sex | BMI, kg/m2 |
|---|---|---|---|---|---|---|
| 1 | DE | DB | 40 | 65 | Male | 17.6 |
| 2 | DE | DB | 35 | 45 | Male | 27.0 |
| 3 | DE | DB | 45 | 54 | Male | 24.1 |
| 4 | DB | DE | 35 | 79 | Male | 29.8 |
| 5 | DB | DE | 30 | 83 | Male | 14.0 |
| 6 | DB | DE | 30 | 80 | Male | 22.3 |
BMI, body mass index; DB, dog bone button; DE, double Endobutton.
Figure 1.Illustrations depicting 2 acromioclavicular joint reconstruction techniques: (A) continuous-loop double Endobutton construct and (B) dog bone button fixation construct.
Figure 2.(A) Testing rig. Position of the scapula in the vise, vise angles, and position were adjusted so that scapula was anatomically positioned and vertically oriented. The ring stand position and pulley were adjusted to apply a static perpendicular superior load to the clavicle anterior to the fixation construct. Two marking pins can be seen. (B) Artist depiction of testing rig. AP, anterior-posterior plane; SI, superior-inferior plane.
Cyclical Loading Results for 100, 1000, and 5000 Cycles
| Group | Axial Displacement, mm | Superior Displacement, mm | ||||
|---|---|---|---|---|---|---|
| 100 | 1000 | 5000 | 100 | 1000 | 5000 | |
| DB group | 0.3 ± 0.1 | 1 ± 0.5 | 1.7 ± 0.6 | 0.2 ± 0.1 | 0.5 ± 0.3 | 1.1 ± 0.8 |
| DE group | 0.4 ± 0.1 | 0.8 ± 0.1 | 1.2 ± 0.2 | 0.3 ± 0.1 | 0.5 ± 0.2 | 0.7 ± 0.2 |
|
| .75 | .63 | .19 | .01 | .69 | .32 |
Data are expressed as mean ± SD. DB, dog bone button; DE, double Endobutton.
Statistically significant.
Load-to-Failure Results
| Load at Failure, N | Stiffness, N/mm | |
|---|---|---|
| DB group | 552 ± 205 | 21.1 ± 6.5 |
| DE group | 558 ± 70 | 17.9 ± 3.4 |
|
| .96 | .42 |
Data are expressed as mean ± SD. DB, dog bone button; DE, double Endobutton.