Security of knots tied with ethibond, fiberwire, orthocord, or ultrabraid.

Omer A Ilahi1, Shiraz A Younas, David M Ho, Phillip C Noble.   

Abstract

BACKGROUND: The security of several popular arthroscopic knots to prolonged, incremental, cyclic loads is unknown, as is the security of knots tied with newer, superstrong sutures. HYPOTHESIS: Some arthroscopic knots are as secure as openly tied square knots, and knots tied with superstrong sutures are more secure than those tied with braided polyester. Some arthroscopic knots are significantly bulkier than openly tied square knots. STUDY
DESIGN: Controlled laboratory study.
METHODS: Five types of openly tied knots (3-throw square, 4-throw square, 5-throw square, 5-throw slip, open SAK [simple arthroscopic knot]), 6 complex arthroscopic knots backed with 3 reversed half-hitches with alternating posts (RHAPs) (SMC, Weston, taut-line hitch, Tennessee slider, Roeder, Duncan loop), and 2 stacked half-hitch (SHH) arthroscopic knots (surgeon's [S=S=S//xS//xS//xS], SAK [S=S//xSxS//xS]) were tied using No. 2 Ethibond around 2 aluminum rods, which were pulled apart with stepwise, incremental, cyclic loads to a maximum force of 120 N (2250 total cycles). Then, 5-throw square knots openly tied with No. 2 Fiberwire, Orthocord, or Ultrabraid were subjected to the stepwise, incremental, cyclic loading protocol extended to a 260-N load level. Before mechanical testing, the height (maximum diameter) of each knot was measured with digital calipers.
RESULTS: For Ethibond, the openly tied 3-throw square knots (56.2 +/- 21.4 N) and 5-throw slip knots (49.9 +/- 26.9 N) reached clinical failure (3 mm of laxity) at significantly lower loads (P < .05) than openly tied 5-throw square knots (90.8 +/- 6.5 N), whereas the openly tied SAK (82.3 +/- 9.4 N) and 4-throw square (84.3 +/- 11.6 N) and all arthroscopically tied knots reached 3 mm of laxity at statistically similar loads. Five-throw square knots openly tied with Fiberwire or Orthocord reached 3 mm of laxity at much higher loads (194.9 +/- 28.4 N and 168.4 +/- 8.6 N, respectively) than those tied using Ethibond (P < .001 for each comparison), but there was no significant difference in performance between Fiberwire knots and Orthocord knots. Although Ultrabraid square knots also were stronger than those tied with Ethibond (137.9 +/- 15.9 N, P < .005), they were not as secure as those tied with Orthocord or Fiberwire (P < .05). Compared with the 5-throw square knots, all arthroscopic knots were significantly bulkier. Especially bulky knots were the Duncan loop and the taut-line hitch. Orthocord square knots demonstrated bulkiness similar to Ethibond square knots, whereas Fiberwire and Ultrabraid square knots were significantly bulkier.
CONCLUSIONS: For braided suture, 5-throw knots optimize square knot security. Open or arthroscopic slip knots can achieve similar security with post switching and loop reversal. Fiberwire, Orthocord, or Ultrabraid openly tied square knots offer greater security than those tied with Ethibond. Arthroscopic knots vary in their bulkiness, but all are significantly bulkier than 5-throw openly tied square knots. Square knots openly tied with Fiberwire or Ultrabraid tend to be bulkier than if tied with Ethibond or Orthocord, which are similar to each other. CLINICAL RELEVANCE: The 5-throw openly tied square knot remains the gold standard, although the openly tied SAK offers similar security when tying in a hole. Arthroscopic knots, whether complex knots backed up by 3 RHAPs, the 6-throw surgeon's knot, or the 5-throw SAK, give security similar to the standard. Square knots tied with the newer sutures in open fashion are more secure than if tied with braided polyester. Using lower profile knots may be especially important when employing Fiberwire or Ultrabraid, as these sutures tend to result in bulkier knots than those tied with Ethibond or Orthocord.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18827068     DOI: 10.1177/0363546508323745

Source DB:  PubMed          Journal:  Am J Sports Med        ISSN: 0363-5465            Impact factor:   6.202


  20 in total

1.  Loop security and tensile properties of polyblend and traditional suture materials.

Authors:  Mehmet Türker; Onder Kılıçoğlu; Ahmet Salduz; Ergun Bozdağ; Emin Sünbüloğlu
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2010-06-10       Impact factor: 4.342

2.  Material properties of common suture materials in orthopaedic surgery.

Authors:  S Najibi; R Banglmeier; Jm Matta; M Tannast
Journal:  Iowa Orthop J       Date:  2010

3.  Effects of using a surgical clamp to hold tension while tying knots with commonly used orthopedic sutures.

Authors:  Leslie J Bisson; Andrew D Sobel; David Godfrey
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2011-09-16       Impact factor: 4.342

Review 4.  Biomechanical comparison of double grasping repair versus cross-locked cruciate flexor tendon repair.

Authors:  C Liam Dwyer; D Dean Dominy; Timothy E Cooney; Richard Englund; Leonard Gordon; John D Lubahn
Journal:  Hand (N Y)       Date:  2015-03

5.  Surgical knot tightening: how much pull is necessary?

Authors:  Stefan Neuhofer; Karl Wieser; Georg Lajtai; Daniel Müller; Christian Gerber; Dominik C Meyer
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-03-14       Impact factor: 4.342

6.  Optimization of the racking hitch knot: how many half hitches and which suture material provide the greatest security?

Authors:  James D Kelly; Suketu Vaishnav; Bradley M Saunders; Mark A Schrumpf
Journal:  Clin Orthop Relat Res       Date:  2014-02-11       Impact factor: 4.176

7.  Highly unstable cervical spine injury in an infant: a case report.

Authors:  Sue young Ha; Sun-Ho Lee; Eun-Sang Kim; Hyung Jin Shin; Whan Eoh
Journal:  Childs Nerv Syst       Date:  2014-08-21       Impact factor: 1.475

8.  Suture Anchor Versus Transosseous Tunnel Repair for Inferior Pole Patellar Fractures Treated With Partial Patellectomy and Tendon Advancement: A Biomechanical Study.

Authors:  Ryan O'Donnell; Nicholas J Lemme; Stephen Marcaccio; Devin F Walsh; Kalpit N Shah; Brett D Owens; Steven F DeFroda
Journal:  Orthop J Sports Med       Date:  2021-08-16

9.  Evaluation of polybutylate-coated braided polyester (ethibond) sutures for levator-advancement blepharoptosis repair.

Authors:  Michael Yulish; Joseph Pikkel
Journal:  ISRN Ophthalmol       Date:  2012-09-18

10.  Is the surgical knot tying technique associated with a risk for unnoticed glove perforation? An experimental study.

Authors:  Vincenzo Giordano; Hilton Augusto Koch; Juliano de Sousa Prado; Leonardo Schiavo de Morais; Rafael de Araújo Hara; Felipe Serrão de Souza; Ney Pecegueiro do Amaral
Journal:  Patient Saf Surg       Date:  2014-06-17
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.