Magali Garcia1, Céline Chessa1, Anne Bourgoin2, Geneviève Giraudeau2, Chloé Plouzeau3, Gérard Agius2, Nicolas Lévêque4, Agnès Beby-Defaux5.
Abstract
BACKGROUND: eMAG™ (bioMerieux) is a new nucleic acid extraction platform based on magnetic silica technology, like its predecessor, NucliSENS® easyMAG® (bioMerieux). Using the same reagents and disposables, eMAG™ adds further automation, allowing simultaneous extraction of 48 samples directly from primary tubes, and distribution of nucleic acid extracts on PCR strips or in tubes at the end of the extraction process.Entities:
Keywords: Clinical specimens; Evaluation; Nucleic acid extraction; eMAG™; easyMAG(®)
Mesh:
Substances:
Year: 2017 PMID: 28160729 PMCID: PMC7185493 DOI: 10.1016/j.jcv.2017.01.004
Source DB: PubMed Journal: J Clin Virol ISSN: 1386-6532 Impact factor: 3.168
Qualitative comparison and agreement percentages between NucliSENS® easyMAG® and eMAG™ nucleic acid extraction from respiratory, whole blood, plasma and urine samples.
| easyMAG® extraction | |||||
|---|---|---|---|---|---|
| Positive | Negative | Agreement | |||
| eMAG™ extraction | Respiratory | Positive | 415 | 8 | |
| Negative | 10 | 8 | 95.9% | ||
| Whole Blood | Positive | 52 | 5 | ||
| Negative | 5 | 3 | 84.6% | ||
| Plasma | Positive | 21 | 3 | ||
| Negative | 0 | 1 | 88% | ||
| Urine | Positive | 24 | 0 | ||
| Negative | 0 | 1 | 100% | ||
Fig. 1Analysis of the cycle threshold of respiratory virus (A) and cell control (B) detection after Nuclisens easyMAG® and eMAG™ extractions. (Left) Deming regression analysis of Ct values of samples positive with both extraction methods. The Deming regression curve equation and the correlation coefficient are indicated. The thin dotted line represents hypothetical identical performance between the two methods (Y = X), and the bold dotted line shows the actual correlation between the two methods; (right) Bland-Altman difference scatter plot for samples positive with both extraction methods.
Fig. 2Comparison of CMV, EBV and HHV6 viral load in 50 whole blood specimens after extraction with Nuclisens easyMAG® and eMAG™. (Left) Deming regression analysis of viral load (log10 copies/mL) in samples positive with both extraction methods. The Deming regression curve equation and the correlation coefficient are indicated. The solid line represents hypothetical identical performance between the two methods (Y = X), while the dotted line shows the actual correlation; (right) Bland-Altman difference scatter plot for samples positive by both extraction methods.
Fig. 3Comparison of BKV viral load in 25 plasma and 25 urine samples after extraction with Nuclisens easyMAG® and eMAG™. (Left) Deming regression analysis of viral load (log10 copies/mL) in samples positive with both extraction methods. The Deming regression curve equation and the correlation coefficient are indicated. The solid line represents hypothetical identical performance between the 2 methods (Y = X), while the dotted line shows the actual correlation; (right) Bland-Altman difference scatter plot for samples positive with both extraction methods.