Literature DB >> 22014618

Comparative evaluation of the QIAGEN QIAsymphony® SP system and bioMérieux NucliSens easyMAG automated extraction platforms in a clinical virology laboratory.

Adele V Lee1, Claire Atkinson, Rohini J Manuel, Duncan A Clark.   

Abstract

BACKGROUND: The QIAGEN QIAsymphony(®) SP is an automated system that can process a variety of different sample types for nucleic acid extraction.
OBJECTIVES: To compare this system against the bioMérieux NucliSens easyMAG using a range of common clinical sample types. STUDY
DESIGN: Nucleic acid extracts were tested on 6 in-house real time viral PCR assays: quantitative cytomegalovirus (CMV); respiratory and enteric multiplex (10 viruses); influenza A H1N1 2009 specific H1 and N1 assays (sH1/N1); herpes simplex virus (HSV) 1, HSV 2 and varicella zoster virus (VZV) multiplex; norovirus genogroups I and II (Noro GI/II) multiplex.
RESULTS: Extraction of the clinical sample by either QIAsymphony(®) SP or NucliSens easyMAG gave similar results for each PCR; CMV viral loads, 52 plasma samples had a mean difference (easyMAG-QIAsymphony(®)) of 0.002 log(10)copies/ml (s.d. 0.536), 52 whole blood samples had a mean difference of -0.232 log(10)copies/ml (s.d. 0.490). Concordance for the qualitative assays were; 64/67 (95.5%) for the respiratory and enteric multiplex, all 28 (100%) for the sH1, sN1 and influenza A matrix multiplex, 33/34 (97%) for the HSV1/HSV2/VZV multiplex and all 15 (100%) for the Noro GI/II. Inter- and intra-run variation, determined for a 10-fold dilution series of CMV (5.20-3.20 log(10)copies/ml), was less than 0.63 log(10)copies/ml.
CONCLUSIONS: Our evaluation found the performance of the QIAsymphony(®) SP comparable to the NucliSens easyMAG for a range of sample types commonly extracted in a clinical virology laboratory. In total, 331/343 (96.5%) PCR results were concordant on samples extracted by both platforms.
Copyright © 2011 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22014618     DOI: 10.1016/j.jcv.2011.08.016

Source DB:  PubMed          Journal:  J Clin Virol        ISSN: 1386-6532            Impact factor:   3.168


  7 in total

1.  Automated, simple, and efficient influenza RNA extraction from clinical respiratory swabs using TruTip and epMotion.

Authors:  Sara B Griesemer; Rebecca Holmberg; Christopher G Cooney; Nitu Thakore; Alissa Gindlesperger; Christopher Knickerbocker; Darrell P Chandler; Kirsten St George
Journal:  J Clin Virol       Date:  2013-07-21       Impact factor: 3.168

2.  Fully automated quantification of cytomegalovirus (CMV) in whole blood with the new sensitive Abbott RealTime CMV assay in the era of the CMV international standard.

Authors:  Nathalie Schnepf; Catherine Scieux; Matthieu Resche-Riggon; Linda Feghoul; Alienor Xhaard; Sébastien Gallien; Jean-Michel Molina; Gérard Socié; Denis Viglietti; François Simon; Marie-Christine Mazeron; Jérôme Legoff
Journal:  J Clin Microbiol       Date:  2013-04-24       Impact factor: 5.948

3.  Performance of the MagNA Pure 96 system for cytomegalovirus nucleic acid amplification testing in clinical samples.

Authors:  A Edelmann; U Eichenlaub; S Lepek; D H Krüger; J Hofmann
Journal:  J Clin Microbiol       Date:  2013-03-06       Impact factor: 5.948

4.  Automated Nucleic Acid Extraction Systems for Detecting Cytomegalovirus and Epstein-Barr Virus Using Real-Time PCR: A Comparison Study Between the QIAsymphony RGQ and QIAcube Systems.

Authors:  Hanah Kim; Mina Hur; Ji Young Kim; Hee Won Moon; Yeo Min Yun; Hyun Chan Cho
Journal:  Ann Lab Med       Date:  2017-03       Impact factor: 3.464

5.  Comparison of eMAG™ versus NucliSENS® EasyMAG® performance on clinical specimens.

Authors:  Magali Garcia; Céline Chessa; Anne Bourgoin; Geneviève Giraudeau; Chloé Plouzeau; Gérard Agius; Nicolas Lévêque; Agnès Beby-Defaux
Journal:  J Clin Virol       Date:  2017-01-29       Impact factor: 3.168

6.  Comparative evaluation of the QIAsymphony RGQ system with the easyMAG/R-gene combination for the quantitation of cytomegalovirus DNA load in whole blood.

Authors:  Sylvie Pillet; Thomas Bourlet; Bruno Pozzetto
Journal:  Virol J       Date:  2012-10-09       Impact factor: 4.099

7.  Laboratory detection of respiratory viruses by automated techniques.

Authors:  Mercedes Pérez-Ruiz; Irene Pedrosa-Corral; Sara Sanbonmatsu-Gámez; María Navarro-Marí
Journal:  Open Virol J       Date:  2012-11-30
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.