Literature DB >> 21199666

Comparison of the QIAsymphony automated nucleic acid extraction and PCR setup platforms with NucliSens easyMAG and Corbett CAS-1200 liquid handling station for the detection of enteric pathogens in fecal samples.

E van Zanten1, G J Wisselink, W de Boer, S Stoll, R Alvarez, A M D Kooistra-Smid.   

Abstract

In this study we compared the QIAsymphony Sample Preparation and Assay Setup modules (Qiagen), which provide automated nucleic acid extraction and PCR setup, to the NucliSens easyMAG (bioMérieux) and CAS-1200 liquid handling station (Corbett) for a molecular screening approach of enteric pathogens in fecal samples using multiplex real-time PCR. The relative DNA recovery of both platforms, within- and between-run reproducibility and a prospective study, including 510 clinical fecal samples, were performed. The results demonstrated that the QIAsymphony Sample Preparation and Assay Setup modules were highly reproducible and achieve equal performance, quantitative and qualitative, when compared with the NucliSens easyMAG and CAS-1200 systems for the molecular screening analysis of enteric pathogens in fecal samples.
Copyright © 2010 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21199666     DOI: 10.1016/j.mimet.2010.12.024

Source DB:  PubMed          Journal:  J Microbiol Methods        ISSN: 0167-7012            Impact factor:   2.363


  4 in total

Review 1.  Molecular testing for clinical diagnosis and epidemiological investigations of intestinal parasitic infections.

Authors:  Jaco J Verweij; C Rune Stensvold
Journal:  Clin Microbiol Rev       Date:  2014-04       Impact factor: 26.132

2.  Comparison of eMAG™ versus NucliSENS® EasyMAG® performance on clinical specimens.

Authors:  Magali Garcia; Céline Chessa; Anne Bourgoin; Geneviève Giraudeau; Chloé Plouzeau; Gérard Agius; Nicolas Lévêque; Agnès Beby-Defaux
Journal:  J Clin Virol       Date:  2017-01-29       Impact factor: 3.168

3.  Comparison of DNA Extraction Methods and Real-Time PCR Assays for the Detection of Blastocystis sp. in Stool Specimens.

Authors:  Céline Nourrisson; Julie Brunet; Pierre Flori; Maxime Moniot; Virginie Bonnin; Frédéric Delbac; Philippe Poirier
Journal:  Microorganisms       Date:  2020-11-11

4.  Comparative evaluation of the QIAsymphony RGQ system with the easyMAG/R-gene combination for the quantitation of cytomegalovirus DNA load in whole blood.

Authors:  Sylvie Pillet; Thomas Bourlet; Bruno Pozzetto
Journal:  Virol J       Date:  2012-10-09       Impact factor: 4.099

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.