Alberto Chisari1, Anna Maria Pistritto2, Raffaele Piccolo3, Alessio La Manna4, Gian Battista Danzi5.
Abstract
The Ultimaster coronary stent system (Terumo Corporation, Tokyo, Japan) represents a new iteration in drug-eluting stent (DES) technology that has recently received the Conformité Européenne (CE) mark approval for clinical use. The Ultimaster is a thin-strut, cobalt chromium, biodegradable-polymer, sirolimus-eluting coronary stent. The high elasticity of the biodegradable-polymer (PDLLA-PCL) and the abluminal gradient coating technology are additional novel features of this coronary device. The Ultimaster DES has undergone extensive clinical evaluation in two studies: The CENTURY I and II trials. Results from these two landmark studies suggested an excellent efficacy and safety profile of the Ultimaster DES across several lesion and patient subsets, with similar clinical outcomes to contemporary, new-generation DES. The aim of this review is to summarize the rationale behind this novel DES technology and to provide an update of available evidence about the clinical performance of the Ultimaster DES.Entities:
Keywords: Ultimaster stent; biodegradable polymer drug eluting stent; coronary stent; new-generation DES; sirolimus-eluting stent
Mesh:
Substances:
Year: 2016 PMID: 27608017 PMCID: PMC5037768 DOI: 10.3390/ijms17091490
Source DB: PubMed Journal: Int J Mol Sci ISSN: 1422-0067 Impact factor: 5.923
Figure 1Gradient coating of Ultimaster coronary stent: Gradient coating consists in the absence of drug polymer on the hinges of the stent in order to reduce the risk of polymer damage that can occur with circumferential coating. The gradient coating is achieved by the multi-layer coating process, protected by Terumo’s patents.
Figure 2Process of polymer degradation after stent implantation: Bioresorbable polymer is metabolized into carbon dioxide and water and excreted via urine and feces within 3–4 months. Sirolimus is released along with polymer bioabsorption. After that time the Ultimaster drug-eluting stent is expected to behave as a thin strut bare metal stent. Blue spots indicate % drug release over time; red circles indicate % amount of polymer remaining on stent surface.