Literature DB >> 14744976

Localized hypersensitivity and late coronary thrombosis secondary to a sirolimus-eluting stent: should we be cautious?

Renu Virmani1, Giulio Guagliumi, Andrew Farb, Giuseppe Musumeci, Niccolo Grieco, Teresio Motta, Laurian Mihalcsik, Maurizio Tespili, Orazio Valsecchi, Frank D Kolodgie.   

Abstract

BACKGROUND: The US Food and Drug Administration recently issued a warning of subacute thrombosis and hypersensitivity reactions to sirolimus-eluting stents (Cypher). The cause and incidence of these events have not been determined. METHODS AND
RESULTS: We present findings of a 58-year-old man who died of late stent thrombosis 18 months after receiving 2 Cypher stents for unstable angina. Although angiographic and intravascular ultrasound results at 8 months demonstrated the absence of neointimal formation, vessel enlargement was present. An autopsy showed aneurysmal dilation of the stented arterial segments with a severe localized hypersensitivity reaction consisting predominantly of T lymphocytes and eosinophils.
CONCLUSIONS: The known pharmacokinetic elution profile of Cypher stents and the presence of polymer fragments surrounded by giant cells and eosinophils suggest that a reaction to the polymer may have caused late stent thrombosis. Careful long-term follow-up of patients with vessel enlargement after Cypher stent placement is recommended.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 14744976     DOI: 10.1161/01.CIR.0000116202.41966.D4

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  244 in total

1.  Comparison of vascular remodeling in patients treated with sirolimus-versus zotarolimus-eluting stent following acute myocardial infarction.

Authors:  Ki-Woon Kang; Young-Guk Ko; Dong-Ho Shin; Jung-Sun Kim; Byeong-Keuk Kim; Donghoon Choi; Yangsoo Jang; Myeong-Ki Hong
Journal:  Clin Cardiol       Date:  2011-12-12       Impact factor: 2.882

2.  [Drug-coated stents. Where do we stand in 2004?].

Authors:  B Scheller
Journal:  Z Kardiol       Date:  2004-09

3.  Evaluation of in-stent neointimal tissue components using integrated backscatter intravascular ultrasound: comparison of drug-eluting stents and bare-metal stents.

Authors:  Yoshitaka Muraoka; Shinjo Sonoda; Kuninobu Kashiyama; Fumihiko Kamezaki; Yuki Tsuda; Masaru Araki; Masahiro Okazaki; Yutaka Otsuji
Journal:  Int J Cardiovasc Imaging       Date:  2011-12-18       Impact factor: 2.357

Review 4.  Late stent thrombosis: the last remaining obstacle in coronary interventional therapy.

Authors:  Piera Capranzano; George Dangas
Journal:  Curr Cardiol Rep       Date:  2012-08       Impact factor: 2.931

5.  Drug-eluting stents.

Authors:  Xiaodong Ma; Tim Wu; Michael P Robich; Xingwei Wang; Hao Wu; Bryan Buchholz; Stephen McCarthy
Journal:  Int J Clin Exp Med       Date:  2010-07-15

Review 6.  Emerging applications of nanotechnology for the diagnosis and management of vulnerable atherosclerotic plaques.

Authors:  Shann S Yu; Ryan A Ortega; Brendan W Reagan; John A McPherson; Hak-Joon Sung; Todd D Giorgio
Journal:  Wiley Interdiscip Rev Nanomed Nanobiotechnol       Date:  2011-08-10

7.  Effect of a novel peptide, WKYMVm- and sirolimus-coated stent on re-endothelialization and anti-restenosis.

Authors:  Eun-Jae Jang; In-Ho Bae; Dae Sung Park; So-Youn Lee; Kyung Seob Lim; Jun-Kyu Park; Jae-Won Shim; Doo Sun Sim; Myung Ho Jeong
Journal:  J Mater Sci Mater Med       Date:  2015-10-05       Impact factor: 3.896

8.  Mechanism of in-stent restenosis after second-generation drug-eluting stents (DES): is it different from bare-metal stents and first-generation DES?

Authors:  Shoichi Kuramitsu; Shinichi Shirai; Kenji Ando
Journal:  J Thorac Dis       Date:  2015-12       Impact factor: 2.895

9.  Comparison of the vessel healing process after everolimus-eluting stent and bare metal stent implantations in patients with ST-elevation myocardial infarction.

Authors:  Hideki Yano; Shigeo Horinaka; Manami Watahik; Tomoko Watanabe; Toshihiko Ishimitsu
Journal:  Heart Vessels       Date:  2018-11-03       Impact factor: 2.037

10.  Stent thrombosis with an aneurysm 7 years after a drug eluting stent implantation.

Authors:  Pritam Patil; Arvind Sethi; Upendra Kaul
Journal:  Indian Heart J       Date:  2014-03-27
View more