Literature DB >> 15642025

Biological and biomechanical evaluation of bone remodelling and implant stability after using an osteotome technique.

André Büchter1, Johannes Kleinheinz, Hans Peter Wiesmann, Jan Kersken, Manuel Nienkemper, Hendrik von Weyhrother, Ulrich Joos, Ulrich Meyer.   

Abstract

The influence of the osteotome technique on the osseointegration and biomechanical behaviour of cylinder implants (SLA, ITI was compared with conventional preparation of the implant site in an animal model. A total of 56 implants were placed in the cranial and caudal tibia condyle of six Gottinger minipigs. The implant site was prepared either by the conventional technique with drills (control group A) or by the osteotome technique (experimental group B). Resonance frequency measurements (RFMs) were made on each implant at the time of fixture placement and at the time of scarification. Half of the minipigs were sacrificed 7 days and 28 days after implant placement and the implants were removed with the surrounding bone. Bone tissue responses were evaluated by histological analysis and removal torque testing. For histological evaluation 30-50 microm-thick ground sections were examined. Biomechanical testing revealed a significantly higher stability of implants in the control group (A) than in the experimental group (B) (P = 0.004) at day 7. After 28 days implant stability in the control group remained significant higher (47%) than those of group B (P > 0.001). RFM demonstrated no significant difference between both groups and during the experimental course. Histological analysis demonstrated fractured trabeculae in peri-implant bone in the experimental group at day 7, while they were not posed at day 28. We conclude that the decreased implant stability by using the osteotome technique is based on microfractures in peri-implant bone.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15642025     DOI: 10.1111/j.1600-0501.2004.01081.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  23 in total

1.  Clinical indications, advantages and limits of the expansion-condensing osteotomes technique for the creation of implant bed.

Authors:  G DE Vico; M Bonino; D Spinelli; A Pozzi; A Barlattani
Journal:  Oral Implantol (Rome)       Date:  2009-12-10

2.  Orthodontic mini-implant stability at different insertion depths : Sensitivity of three stability measurement methods.

Authors:  Manuel Nienkemper; Natascha Santel; Ralf Hönscheid; Dieter Drescher
Journal:  J Orofac Orthop       Date:  2016-06-07       Impact factor: 1.938

3.  Comparison of heat production and bone architecture changes in the implant site preparation with compressive osteotomes, osseodensification technique, piezoelectric devices, and standard drills: an ex vivo study on porcine ribs.

Authors:  Nishith Bhargava; Vittoria Perrotti; Vito Carlo Alberto Caponio; Victor Haruo Matsubara; Diana Patalwala; Alessandro Quaranta
Journal:  Odontology       Date:  2022-07-19       Impact factor: 2.885

4.  Peri-implant bone reactions around immediately loaded conical implants with different prosthetic suprastructures: histological and histomorphometrical study on minipigs.

Authors:  Bartolomeo Assenza; Antonio Scarano; Vittoria Perrotti; Iole Vozza; Alessandro Quaranta; Manlio Quaranta; Adriano Piattelli; Giovanna Iezzi
Journal:  Clin Oral Investig       Date:  2009-06-04       Impact factor: 3.573

5.  In vitro and in vivo studies of osteoblast cell response to a titanium-6 aluminium-4 vanadium surface modified by neodymium:yttrium-aluminium-garnet laser and silicon carbide paper.

Authors:  M E Khosroshahi; M Mahmoodi; H Saeedinasab
Journal:  Lasers Med Sci       Date:  2008-11-19       Impact factor: 3.161

6.  Measurement of mini-implant stability using resonance frequency analysis.

Authors:  Manuel Nienkemper; Benedict Wilmes; Agamemnon Panayotidis; Alexander Pauls; Vladimir Golubovic; Frank Schwarz; Dieter Drescher
Journal:  Angle Orthod       Date:  2012-08-29       Impact factor: 2.079

7.  Comparison of Conventional and Standardized Bone Densitometry around Implants in Periapical Radiographs during a Three Months Period.

Authors:  Asieh Zamani Naser; Shahab Etemadi; Mansour Rismanchian; Mahnaz Sheikhi; Mohammadbagher Tavakoli
Journal:  Dent Res J (Isfahan)       Date:  2011

8.  Mini-implants in the anchorage armamentarium: new paradigms in the orthodontics.

Authors:  Masaru Yamaguchi; Toshihiro Inami; Ko Ito; Kazutaka Kasai; Yasuhiro Tanimoto
Journal:  Int J Biomater       Date:  2012-06-05

9.  Biological and biomechanical evaluation of interface reaction at conical screw-type implants.

Authors:  Andre Büchter; Ulrich Joos; Hans-Peter Wiesmann; László Seper; Ulrich Meyer
Journal:  Head Face Med       Date:  2006-02-21       Impact factor: 2.151

10.  Repair of microdamage in osteonal cortical bone adjacent to bone screw.

Authors:  Lei Wang; Tingjun Ye; Lianfu Deng; Jin Shao; Jin Qi; Qi Zhou; Li Wei; Shijing Qiu
Journal:  PLoS One       Date:  2014-02-20       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.