Repeatability data and agreement of keratometry with the VERION system compared to the IOLMaster.

Gabor Nemeth, Eszter Szalai, Ziad Hassan, Agnes Lipecz, Andras Berta, Laszlo Modis.   

Abstract

PURPOSE: To analyze the repeatability of keratometric and white-to-white (WTW) distance measurements with the VERION Measurement Module (Alcon Laboratories, Inc., Fort Worth, TX) and to compare the measured data to the results of the IOLMaster (Carl Zeiss Meditec, Jena, Germany).
METHODS: Three images were captured with the VERION and the flattest and steepest keratometric data, the astigmatism axis, and the WTW distance were recorded. Subsequently, the axial length, the keratometric data with axis, and the WTW distance were measured with an IOLMaster. The repeatability data of the keratometric value of the VERION System, converted to cross cylinder J0 and J45 vector components, were analyzed. The agreement data for keratometry obtained by the VERION System and the differences regarding keratometric data and WTW distance compared to IOLMaster were calculated.
RESULTS: The measurements were conducted in 50 eyes of 50 healthy volunteers (median age: 50.32 years, range: 19.34 to 85.3 years). The flattest and the steepest keratometric data, the diopter of astigmatism, the J0 and J45 vector components, and WTW distance did not differ significantly between devices (P > .05). Intraclass correlation coefficients (range: 0.863 to 0.994) and Cronbach's alpha values (range: 0.950 to 0.998) were high for all parameters measured by the VERION System.
CONCLUSIONS: The VERION System has high repeatability and agreement with the IOLMaster, making it suitable as an alternative tool in clinical practice. Copyright 2015, SLACK Incorporated.

Entities:  

Mesh:

Year:  2015        PMID: 25974973     DOI: 10.3928/1081597X-20150424-01

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  3 in total

1.  Comparison of biometric measurements obtained by the Verion Image-Guided System versus the auto-refracto-keratometer.

Authors:  Cecilio Velasco-Barona; Guadalupe Cervantes-Coste; Erick Mendoza-Schuster; Claudia Corredor-Ortega; Nadia L Casillas-Chavarín; Alejandro Silva-Moreno; Manuel Garza-León; Roberto Gonzalez-Salinas
Journal:  Int Ophthalmol       Date:  2017-04-25       Impact factor: 2.031

2.  Comparison of keratometric measurements obtained by the Verion Image Guided System with optical biometry and auto-keratorefractometer.

Authors:  Leyla Asena; Sirel Gür Güngör; Ahmet Akman
Journal:  Int Ophthalmol       Date:  2016-06-07       Impact factor: 2.031

3.  Comparison of corneal power obtained from VERION image-guided surgery system and four other devices.

Authors:  Hung-Yuan Lin; Hsin-Yang Chen; Han Bor Fam; Ya-Jung Chuang; Ronald Yeoh; Pi-Jung Lin
Journal:  Clin Ophthalmol       Date:  2017-07-12