Agreement between a partial coherence interferometer and 2 manual keratometers.

Mark A Bullimore1, Tobias Buehren, Wilfried Bissmann.   

Abstract

PURPOSE: To establish repeatability and validity of a partial coherence interferometry (PCI) device in patients with astigmatism.
SETTING: Industry and university.
DESIGN: Observational studies.
METHODS: A prospective study determined the agreement between the IOLMaster 500 PCI device and the Marco 1-position manual keratometer for measurement of mean corneal power, astigmatic power, and axis. Interinstrument and interoperator reproducibility was also assessed. A retrospective study determined the agreement between the PCI device and the Javal 2-position manual keratometer. Repeatability was also assessed. The 95% limits of agreement (LoA) were determined for all comparisons.
RESULTS: Comparing the PCI device and the 1-position manual keratometer, the axis was within 5 degrees in 72.1% of eyes; the cylinder power differed by 0.18 diopter (D) (95% LoA, ±0.45 D) and the mean power by 0.33 D (95% LoA, ±0.25 D). Comparing the PCI device and the 2-position manual keratometer, the axis was within 5 degrees in 77.1% of eyes. The mean difference in cylinder power between the 2 instruments was close to zero (95% LoA, ±0.37 D); however, the mean power differed by 0.34 D (95% LoA, ±0.25 D). The repeatability of the PCI device was better than that of both manual keratometers.
CONCLUSIONS: Agreement between the PCI device and the 2-position manual keratometer for measures of corneal astigmatism and axis was excellent. The 1-position manual keratometer gave slightly lower estimates of corneal astigmatism, although the agreement for axis was excellent. Thus, the PCI device is an appropriate choice for the calculation of toric intraocular lenses. FINANCIAL DISCLOSURE: Dr. Bullimore is a consultant to Carl Zeiss Meditec AG and Alcon Surgical, Inc. Drs. Buehren and Bissmann are employees of Carl Zeiss Meditec AG.
Copyright © 2013 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23876813     DOI: 10.1016/j.jcrs.2013.03.034

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  8 in total

1.  Comparison of keratometric measurements obtained by the Verion Image Guided System with optical biometry and auto-keratorefractometer.

Authors:  Leyla Asena; Sirel Gür Güngör; Ahmet Akman
Journal:  Int Ophthalmol       Date:  2016-06-07       Impact factor: 2.031

2.  Repeatability of Ophtha Top topography and comparison with IOL-Master and LenstarLS900 in cataract patients.

Authors:  Sha-Sha Yu; Hui Song; Xin Tang
Journal:  Int J Ophthalmol       Date:  2017-11-18       Impact factor: 1.779

3.  Comparison of Total Corneal Astigmatism between IOLMaster and Pentacam.

Authors:  Xiaochun Li; Xiaoguang Cao; Yongzhen Bao
Journal:  Biomed Res Int       Date:  2022-07-08       Impact factor: 3.246

4.  The effect of corneal irregularity on astigmatism measurement by automated versus ray tracing keratometry.

Authors:  Hyun Cheol Roh; Roy S Chuck; Jimmy K Lee; Choul Yong Park
Journal:  Medicine (Baltimore)       Date:  2015-04       Impact factor: 1.889

5.  Corneal Astigmatism Measurements Comparison among Ray-Tracing Aberrometry, Partial Coherence Interferometry, and Scheimpflug Imaging System.

Authors:  Yaqin Zhang; Jing Dong; Suhua Zhang; Bin Sun; Xiaoliang Wang; Maolong Tang; Xiaogang Wang
Journal:  J Ophthalmol       Date:  2020-04-01       Impact factor: 1.909

6.  Influence of Dry Eye Disease on the Measurement Repeatability of Corneal Curvature Radius and Axial Length in Patients with Cataract.

Authors:  Takahiro Hiraoka; Hiroki Asano; Tomohiro Ogami; Shinichiro Nakano; Yoshifumi Okamoto; Yoshiaki Yamada; Tetsuro Oshika
Journal:  J Clin Med       Date:  2022-01-28       Impact factor: 4.241

7.  Distribution and Repeatability of Corneal Astigmatism Measurements (Magnitude and Axis) Evaluated With Color Light Emitting Diode Reflection Topography.

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  Cornea       Date:  2015-08       Impact factor: 2.651

8.  Posterior corneal astigmatism: a review article.

Authors:  Seyed-Farzad Mohammadi; Masoud Khorrami-Nejad; Moein Hamidirad
Journal:  Clin Optom (Auckl)       Date:  2019-08-12
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.