Comparison of biometric measurements obtained by the Verion Image-Guided System versus the auto-refracto-keratometer.

Cecilio Velasco-Barona1, Guadalupe Cervantes-Coste1, Erick Mendoza-Schuster1, Claudia Corredor-Ortega1, Nadia L Casillas-Chavarín1, Alejandro Silva-Moreno1, Manuel Garza-León2, Roberto Gonzalez-Salinas3.   

Abstract

PURPOSE: To compare the biometric measurements obtained from the Verion Image-Guided System to those obtained by auto-refracto-keratometer in normal eyes.
METHODS: This is a prospective, observational, comparative study conducted at the Asociación para Evitar la Ceguera en México I.A.P., Mexico. Three sets of keratometry measurements were obtained using the image-guided system to assess the coefficient of variation, the within-subject standard deviation and intraclass correlation coefficient (ICC). A paired Student t test was used to assess statistical significance between the Verion and the auto-refracto-keratometer. A Pearson's correlation coefficient (r) was obtained for all measurements, and the level of agreement was verified using Bland-Altman plots.
RESULTS: The right eyes of 73 patients were evaluated by each platform. The Verion coefficient of variation was 0.3% for the flat and steep keratometry, with the ICC being greater than 0.9 for all parameters measured. Paired t test showed statistically significant differences between groups (P = 0.0001). A good correlation was evidenced for keratometry values between platforms (r = 0.903, P = 0.0001 for K1, and r = 0.890, P = 0.0001). Bland-Altman plots showed a wide data spread for all variables.
CONCLUSION: The image-guided system provided highly repeatable corneal power and keratometry measurements. However, significant differences were evidenced between the two platforms, and although values were highly correlated, they showed a wide data spread for all analysed variables; therefore, their interchangeable use for biometry assessment is not advisable.

Entities:  

Keywords:  Cataract surgery; Corneal power; IOL calculation; Keratometry; Verion Image-Guided System

Mesh:

Year:  2017        PMID: 28444525     DOI: 10.1007/s10792-017-0541-3

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  18 in total

1.  Prevalence of refractive error in the United States, 1999-2004.

Authors:  Susan Vitale; Leon Ellwein; Mary Frances Cotch; Frederick L Ferris; Robert Sperduto
Journal:  Arch Ophthalmol       Date:  2008-08

2.  Repeatability data and agreement of keratometry with the VERION system compared to the IOLMaster.

Authors:  Gabor Nemeth; Eszter Szalai; Ziad Hassan; Agnes Lipecz; Andras Berta; Laszlo Modis
Journal:  J Refract Surg       Date:  2015-05       Impact factor: 3.573

3.  Comparison of a new image-guided system versus partial coherence interferometry, Scheimpflug imaging, and optical low-coherence reflectometry devices: Keratometry and repeatability.

Authors:  Andreas Mueller; Bettina C Thomas; Gerd U Auffarth; Mike P Holzer
Journal:  J Cataract Refract Surg       Date:  2016-05       Impact factor: 3.351

4.  Measurement agreement between a new biometer based on partial coherence interferometry and a validated biometer based on optical low-coherence reflectometry.

Authors:  Junhua Li; Hao Chen; Giacomo Savini; Weicong Lu; Xinxin Yu; Fangjun Bao; Qinmei Wang; Jinhai Huang
Journal:  J Cataract Refract Surg       Date:  2016-01       Impact factor: 3.351

5.  Assessment of corneal curvature using verion optical imaging system: a comparative study.

Authors:  Jenny L Lauschke; Michael Lawless; Gerard Sutton; Timothy V Roberts; Christopher Hodge
Journal:  Clin Exp Ophthalmol       Date:  2016-02-20       Impact factor: 4.207

6.  Comparison of corneal astigmatism and axis location in cataract patients measured by total corneal power, automated keratometry, and simulated keratometry.

Authors:  Sabong Srivannaboon; Chareenun Chirapapaisan; Pratuangsri Chonpimai
Journal:  J Cataract Refract Surg       Date:  2012-09-15       Impact factor: 3.351

7.  Influence factors of estimation errors for total corneal astigmatism using keratometric astigmatism in patients before cataract surgery.

Authors:  Tianyu Zheng; Zhanghua Chen; Yi Lu
Journal:  J Cataract Refract Surg       Date:  2016-01       Impact factor: 3.351

8.  Repeatability and reproducibility of ocular biometry using a new noncontact optical low-coherence interferometer.

Authors:  Jinhai Huang; Giacomo Savini; Fan Wu; Xinxin Yu; Jing Yang; Ayong Yu; Ye Yu; Qinmei Wang
Journal:  J Cataract Refract Surg       Date:  2015-10       Impact factor: 3.351

9.  Repeatability and comparability of corneal power and corneal astigmatism obtained from a point-source color light-emitting diode topographer, a Placido-based corneal topographer, and a low-coherence reflectometer.

Authors:  Bruna V Ventura; Zaina Al-Mohtaseb; Li Wang; Douglas D Koch; Mitchell P Weikert
Journal:  J Cataract Refract Surg       Date:  2015-10       Impact factor: 3.351

10.  Prevalence of Corneal Astigmatism in Patients before Cataract Surgery in Northern China.

Authors:  Xiaoyong Yuan; Hui Song; Gang Peng; Xia Hua; Xin Tang
Journal:  J Ophthalmol       Date:  2014-06-03       Impact factor: 1.909

View more
  1 in total

1.  Influence of Angle κ and Higher-Order Aberrations on Visual Quality Employing Two Diffractive Trifocal IOLs.

Authors:  Cecilio Velasco-Barona; Claudia Corredor-Ortega; Alejandra Mendez-Leon; Nadia L Casillas-Chavarín; Daniel Valdepeña-López Velarde; Guadalupe Cervantes-Coste; Daniel Malacara-Hernández; Roberto Gonzalez-Salinas
Journal:  J Ophthalmol       Date:  2019-11-27       Impact factor: 1.909

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.