Stephanie Fichtner1, Tilko Reents2, Sonia Ammar2, Verena Semmler2, Susanne Kathan2, Roger Dillier2, Alexandra Buiatti2, Gabriele Hessling2, Isabel Deisenhofer2.
Abstract
INTRODUCTION: Pulmonary vein reconnection after pulmonary vein isolation (PVI) is a significant problem in the treatment of paroxysmal atrial fibrillation (AF). We report about patients who underwent contact force (CF) guided PVI using CF catheter and compared them to patients with PVI using an ablation catheter with enhanced tip irrigation.Entities:
Keywords: Ablation; Atrial fibrillation, AF; Contact force; Smarttouch; Surround flow
Year: 2015 PMID: 26937108 PMCID: PMC4750134 DOI: 10.1016/j.ipej.2015.07.004
Source DB: PubMed Journal: Indian Pacing Electrophysiol J ISSN: 0972-6292
Fig. 1Anterior (left side) and posterior view of an anatomical map of the left atrium merged with a cardiac CT image of the same patients using the Carto3 System. Pulmonary veins were circumferentially isolated using a contact force measurement system (Thermocool Smarttouch catheter, Biosense Webster). The color-coding shows good contact (green to blue) or bad contact (red) on the map. The shown number 17 on the left and 18 on the right side shows the online contact force in grams.
Baseline characteristics.
| Data | Surround Flow catheter group N = 29 | Smarttouch catheter group N = 30 | P value |
|---|---|---|---|
| Age | 61 ± 12 | 59.1 ± 10 | |
| Gender male | 48% | 53% | |
| Art. Hypertension | 52% | 55% | |
| Coronary artery disease | 10% | 17% | |
| Diabetes mellitus | 10% | 7% | |
| Previous stroke | 7% | 0 | |
| Body mass index | 22.5 ± 3 | 23.6 ± 3 | |
| Duration of AF in months | 33.1 ± 33 | 55.2 ± 53 | |
| Episodes per year | 157 ± 160 | 123.8 ± 138 | |
| Size of left atrium mmode/planimetric mm2 | 40.4 ± 4/22 ± 5 | 43.6 ± 8/22.0 ± 6 | |
| Anticoagulation with warfarin | 90% | 72% | |
| Previous beta blocker | 72% | 71% | |
| Previous class Ic or III antiarrhythmic drug | 77% | 65% |
Procedural data.
| Data | Surround Flow catheter group N = 29 | Smarttouch catheter group N = 30 | P value |
|---|---|---|---|
| Procedure time (hours) | 2.37 ± 1.13 | 2.15 ± 0.5 | |
| Fluoroscopic time (minutes) | 29 ± 18 | 25.2 ± 13 | |
| Fluoroscopic dose (cGym2) | 3038.3 ± 1997 | 2675.6 ± 1658 | |
| RF time (minutes) | 49.8 ± 19 | 46.6 ± 18 |