Electrical reconnection after pulmonary vein isolation is contingent on contact force during initial treatment: results from the EFFICAS I study.

Petr Neuzil1, Vivek Y Reddy, Josef Kautzner, Jan Petru, Dan Wichterle, Dipen Shah, Hendrik Lambert, Aude Yulzari, Erik Wissner, Karl-Heinz Kuck.   

Abstract

BACKGROUND: Pulmonary vein isolation is the most prevalent approach for catheter ablation of paroxysmal atrial fibrillation. Long-term success of the procedure is diminished by arrhythmia recurrences occurring predominantly because of reconnections in previously isolated pulmonary veins. The aim of the EFFICAS I multicenter study was to demonstrate the correlation between contact force (CF) parameters during initial procedure and the incidence of isolation gaps (gap) at 3-month follow-up. METHOD AND
RESULTS: A radiofrequency ablation catheter with integrated CF sensor (TactiCath, Endosense, Geneva, Switzerland) was used to perform pulmonary vein isolation in 46 patients with paroxysmal atrial fibrillation. During the ablation procedure, the operator was blinded to CF information. At follow-up, an interventional diagnostic procedure was performed to assess gap location as correlated to index procedure ablation parameters. At follow-up, 65% (26/40) of patients showed ≥1 gaps. Ablations with minimum Force-Time Integral (FTI) <400 gs showed increased likelihood for reconnection (P<0.001). Reconnection correlated strongly with minimum CF (P<0.0001) and minimum FTI (P=0.0007) at the site of gap. Gap occurrence showed a strong trend with lower average CF and average FTI. CF and FTI are generally higher on the right side, although the left anterior segment presents a unique challenge to achieve stable position with good CF.
CONCLUSIONS: Minimum CF and minimum FTI values are strong predictors of gap formation. Optimal CF parameter recommendations are a target CF of 20 g and a minimum FTI of 400 gs for each new lesion.

Entities:  

Mesh:

Year:  2013        PMID: 23515263     DOI: 10.1161/CIRCEP.113.000374

Source DB:  PubMed          Journal:  Circ Arrhythm Electrophysiol        ISSN: 1941-3084


  119 in total

1.  Lesion index: a novel guide in the path of successful pulmonary vein isolation.

Authors:  Antonio Dello Russo; Gaetano M Fassini; Michela Casella; Elena Romanelli; Salvatore Pala; Stefania Riva; Valentina Catto; Massimo Moltrasio; Fabrizio Tundo; Martina Zucchetti; Benedetta Majocchi; Maria Antonietta Dessanai; Francesca Pizzamiglio; Giulia Vettor; Valentina Ribatti; Alessio Gasperetti; Selene Cellucci; Gabriele Negro; Rita Sicuso; Corrado Carbucicchio; Claudio Tondo
Journal:  J Interv Card Electrophysiol       Date:  2018-12-04       Impact factor: 1.900

2.  General anesthesia improves contact force and reduces gap formation in pulmonary vein isolation: a comparison with conscious sedation.

Authors:  Akio Chikata; Takeshi Kato; Takanori Yaegashi; Satoru Sakagami; Chieko Kato; Takahiro Saeki; Keiichi Kawai; Shin-Ichiro Takashima; Hisayoshi Murai; Soichiro Usui; Hiroshi Furusho; Shuichi Kaneko; Masayuki Takamura
Journal:  Heart Vessels       Date:  2017-03-04       Impact factor: 2.037

3.  The importance of catheter stability evaluated by Visitag(TM) during pulmonary vein isolation.

Authors:  Ryudo Fujiwara; Kimitake Imamura; Yoichi Kijima; Tomoya Masano; Ryoji Nagoshi; Amane Kohzuki; Hiroyuki Shibata; Yoshiro Tsukiyama; Ryo Takeshige; Kenichi Yanaka; Shinsuke Nakano; Yusuke Fukuyama; Junya Shite
Journal:  J Interv Card Electrophysiol       Date:  2016-01-19       Impact factor: 1.900

4.  Clinical utility of automated ablation lesion tagging based on catheter stability information (VisiTag Module of the CARTO 3 System) with contact force-time integral during pulmonary vein isolation for atrial fibrillation.

Authors:  Yasuo Okumura; Ichiro Watanabe; Kazuki Iso; Koichi Nagashima; Kazumasa Sonoda; Naoko Sasaki; Rikitake Kogawa; Keiko Takahashi; Kimie Ohkubo; Toshiko Nakai; Shiro Nakahara; Yuuichi Hori; Atsushi Hirayama
Journal:  J Interv Card Electrophysiol       Date:  2016-06-09       Impact factor: 1.900

5.  Contact force and impedance decrease during ablation depends on catheter location and orientation: insights from pulmonary vein isolation using a contact force-sensing catheter.

Authors:  Sven Knecht; Tobias Reichlin; Nikola Pavlovic; Beat Schaer; Stefan Osswald; Christian Sticherling; Michael Kühne
Journal:  J Interv Card Electrophysiol       Date:  2015-04-30       Impact factor: 1.900

6.  Randomized comparison of contact force-guided versus conventional circumferential pulmonary vein isolation of atrial fibrillation: prevalence, characteristics, and predictors of electrical reconnections and clinical outcomes.

Authors:  Kohki Nakamura; Shigeto Naito; Takehito Sasaki; Masahiro Nakano; Kentaro Minami; Yosuke Nakatani; Kentaro Ikeda; Eiji Yamashita; Koji Kumagai; Nobusada Funabashi; Shigeru Oshima
Journal:  J Interv Card Electrophysiol       Date:  2015-09-19       Impact factor: 1.900

7.  Robotic ablation of atrial fibrillation with a new remote catheter system.

Authors:  Alexander Wutzler; Thomas Wolber; Abdul Shokor Parwani; Martin Huemer; Philipp Attanasio; Florian Blaschke; Laurent Haegeli; Wilhelm Haverkamp; Firat Duru; Leif-Hendrik Boldt
Journal:  J Interv Card Electrophysiol       Date:  2014-04-29       Impact factor: 1.900

8.  High-voltage zones within the pulmonary vein antra: Major determinants of acute pulmonary vein reconnections after atrial fibrillation ablation.

Authors:  Koichi Nagashima; Ichiro Watanabe; Yasuo Okumura; Kazuki Iso; Keiko Takahashi; Ryuta Watanabe; Masaru Arai; Sayaka Kurokawa; Toshiko Nakai; Kimie Ohkubo; Shunichi Yoda; Atsushi Hirayama
Journal:  J Interv Card Electrophysiol       Date:  2017-04-21       Impact factor: 1.900

9.  Impact of real-time contact force and impedance measurement in pulmonary vein isolation procedures for treatment of atrial fibrillation.

Authors:  Reza Wakili; Sebastian Clauss; Viola Schmidt; Michael Ulbrich; Anton Hahnefeld; Franziska Schüssler; Johannes Siebermair; Stefan Kääb; Heidi L Estner
Journal:  Clin Res Cardiol       Date:  2013-10-06       Impact factor: 5.460

Review 10.  Making better scar: Emerging approaches for modifying mechanical and electrical properties following infarction and ablation.

Authors:  Jeffrey W Holmes; Zachary Laksman; Lior Gepstein
Journal:  Prog Biophys Mol Biol       Date:  2015-11-23       Impact factor: 3.667

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.