Literature DB >> 21872560

A novel radiofrequency ablation catheter using contact force sensing: Toccata study.

Karl-Heinz Kuck1, Vivek Y Reddy, Boris Schmidt, Andrea Natale, Petr Neuzil, Nadir Saoudi, Josef Kautzner, Claudia Herrera, Gerhard Hindricks, Pierre Jaïs, Hiroshi Nakagawa, Hendrik Lambert, Dipen C Shah.   

Abstract

OBJECTIVES: The aim of this multicenter study was to evaluate the device- and procedure-related safety of a novel force-sensing radiofrequency (RF) ablation catheter capable of measuring the real-time contact force (CF) and to present CF data and its possible implications on patient safety.
BACKGROUND: The clinical outcome of RF ablation for the treatment of cardiac arrhythmias may be affected by the CF between the catheter tip and the tissue. Insufficient CF may result in an ineffective lesion, whereas excessive CF may result in complications.
METHODS: Seventy-seven patients (43 with right-sided supraventricular tachycardia [SVT] and 34 with atrial fibrillation [AF]) received percutaneous ablation with the novel studied catheter. The CF applied and safety events related to the procedure were reported.
RESULTS: CF values at mapping ranged from 8 ± 8 to 60 ± 35 g and from 12 ± 10 to 39 ± 29 g in the SVT group and the LA group, respectively, showing a significant interinvestigator variability (P < .0001). High transient CFs (>100 g) were noted in 27 patients (79%) of the LA group. One device-related complication (tamponade, 3%) occurred in the AF group.
CONCLUSIONS: Catheter ablation using real-time CF technology is safe for the treatment of SVT and AF. High CFs may occur during catheter manipulation and not just during ablation, suggesting that measuring CF may provide additional useful information to the operator for safe catheter manipulation. In the future, CF-sensing catheters may also increase the effectiveness of RF ablations by allowing better control of the RF lesion size.
Copyright © 2012 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21872560     DOI: 10.1016/j.hrthm.2011.08.021

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  83 in total

1.  Lesion index: a novel guide in the path of successful pulmonary vein isolation.

Authors:  Antonio Dello Russo; Gaetano M Fassini; Michela Casella; Elena Romanelli; Salvatore Pala; Stefania Riva; Valentina Catto; Massimo Moltrasio; Fabrizio Tundo; Martina Zucchetti; Benedetta Majocchi; Maria Antonietta Dessanai; Francesca Pizzamiglio; Giulia Vettor; Valentina Ribatti; Alessio Gasperetti; Selene Cellucci; Gabriele Negro; Rita Sicuso; Corrado Carbucicchio; Claudio Tondo
Journal:  J Interv Card Electrophysiol       Date:  2018-12-04       Impact factor: 1.900

2.  Contact force monitoring during catheter ablation of intraatrial reentrant tachycardia in patients with congenital heart disease.

Authors:  Ulrich Krause; David Backhoff; Sophia Klehs; Heike E Schneider; Thomas Paul
Journal:  J Interv Card Electrophysiol       Date:  2016-01-07       Impact factor: 1.900

3.  THERMOCOOL® SMARTTOUCH® CATHETER - The Evidence So Far for Contact Force Technology and the Role of VISITAG™ MODULE.

Authors:  Tina Lin; Feifan Ouyang; Karl-Heinz Kuck; Roland Tilz
Journal:  Arrhythm Electrophysiol Rev       Date:  2014-05-30

4.  SmartTouch™ - The Emerging Role of Contact Force Technology in Complex Catheter Ablation.

Authors:  Stephen P Page; Mehul Dhinoja
Journal:  Arrhythm Electrophysiol Rev       Date:  2012-09

5.  MEMS-Based Flexible Force Sensor for Tri-Axial Catheter Contact Force Measurement.

Authors:  Hardik J Pandya; Jun Sheng; Jaydev P Desai
Journal:  J Microelectromech Syst       Date:  2016-12-22       Impact factor: 2.417

Review 6.  Conceptual Design and Procedure for an Autonomous Intramyocardial Injection Catheter.

Authors:  Weyland Cheng; Peter K Law
Journal:  Cell Transplant       Date:  2016-12-07       Impact factor: 4.064

7.  2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation.

Authors:  Hugh Calkins; Gerhard Hindricks; Riccardo Cappato; Young-Hoon Kim; Eduardo B Saad; Luis Aguinaga; Joseph G Akar; Vinay Badhwar; Josep Brugada; John Camm; Peng-Sheng Chen; Shih-Ann Chen; Mina K Chung; Jens Cosedis Nielsen; Anne B Curtis; D Wyn Davies; John D Day; André d'Avila; N M S Natasja de Groot; Luigi Di Biase; Mattias Duytschaever; James R Edgerton; Kenneth A Ellenbogen; Patrick T Ellinor; Sabine Ernst; Guilherme Fenelon; Edward P Gerstenfeld; David E Haines; Michel Haissaguerre; Robert H Helm; Elaine Hylek; Warren M Jackman; Jose Jalife; Jonathan M Kalman; Josef Kautzner; Hans Kottkamp; Karl Heinz Kuck; Koichiro Kumagai; Richard Lee; Thorsten Lewalter; Bruce D Lindsay; Laurent Macle; Moussa Mansour; Francis E Marchlinski; Gregory F Michaud; Hiroshi Nakagawa; Andrea Natale; Stanley Nattel; Ken Okumura; Douglas Packer; Evgeny Pokushalov; Matthew R Reynolds; Prashanthan Sanders; Mauricio Scanavacca; Richard Schilling; Claudio Tondo; Hsuan-Ming Tsao; Atul Verma; David J Wilber; Teiichi Yamane
Journal:  Heart Rhythm       Date:  2017-05-12       Impact factor: 6.343

Review 8.  Contact Force and Atrial Fibrillation Ablation.

Authors:  W Ullah; R J Schilling; T Wong
Journal:  J Atr Fibrillation       Date:  2016-02-29

9.  Which is the best catheter to perform atrial fibrillation ablation? A comparison between standard ThermoCool, SmartTouch, and Surround Flow catheters.

Authors:  Luigi Sciarra; Paolo Golia; Andrea Natalizia; Ermenegildo De Ruvo; Serena Dottori; Antonio Scarà; Alessio Borrelli; Lucia De Luca; Marco Rebecchi; Alessandro Fagagnini; Alberto Bandini; Fabrizio Guarracini; Marcello Galvani; Leonardo Calò
Journal:  J Interv Card Electrophysiol       Date:  2014-02-21       Impact factor: 1.900

10.  Impact of real-time contact force and impedance measurement in pulmonary vein isolation procedures for treatment of atrial fibrillation.

Authors:  Reza Wakili; Sebastian Clauss; Viola Schmidt; Michael Ulbrich; Anton Hahnefeld; Franziska Schüssler; Johannes Siebermair; Stefan Kääb; Heidi L Estner
Journal:  Clin Res Cardiol       Date:  2013-10-06       Impact factor: 5.460

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.