Hamid Razavian1, Pedram Iranmanesh2, Hamid Mojtahedi3, Rahman Nazeri3.
Abstract
INTRODUCTION: Presence of surface defects in endodontic instruments can lead to unwanted complications such as instrument fracture and incomplete preparation of the canal. The current study was conducted to evaluate the effect of autoclave cycles on surface characteristics of S-File by scanning electron microscopy (SEM). METHODS AND MATERIALS: In this experimental study, 17 brand new S-Files (#30) were used. The surface characteristics of the files were examined in four steps (without autoclave, 1 autoclave cycle, 5 autoclave cycles and 10 autoclave cycles) by SEM under 200× and 1000× magnifications. Data were analyzed using the SPSS software and the paired sample t-test, independent sample t-test and multifactorial repeated measures ANOVA. The level of significance was set at 0.05.Entities:
Keywords: Autoclave; Endodontic Instruments; Root Canal Therapy; Scanning Electron Microscopy; Surface Characteristic
Year: 2015 PMID: 26843874 PMCID: PMC4731530 DOI: 10.7508/iej.2016.01.006
Source DB: PubMed Journal: Iran Endod J ISSN: 1735-7497
Figure 1Surface characteristics of S-File; A) before autoclave cycle (200×); B) after 10 autoclave cycles (200×); C) before autoclave cycle (1000×) and D) after 10 autoclave cycles (1000×)
Figure 2A) high and B) low thresholding SEM image
Mean (SD) of instrument surface roughness in total and at separate magnifications
|
|
|
| ||
|---|---|---|---|---|
|
|
| |||
|
| 2.36 (1.29) | 21.66 (9.43)d | 12.01 (11.87) | 0.001 |
|
| 3.84 (2.70) | 32.66 (10.64)e | 18.25 (16.60) | 0.007 |
|
| 7.82 (8.38) | 53.83 (17.27)f | 30.83 (27.05) | 0.050 |
|
| 562.53 (362.30)c | 5117.84 (2733.66)g | 2840.18 (3010.43) | 0.002 |
|
| 0.001 | 0.001 | 0.001 | |
Multifactorial repeated measures of ANOVA test (comparison of surface roughness of files at each magnification);
Independent sample t-test (paired comparison of surface roughness of files at two magnifications); Identical letters indicate insignificant difference