Effect of autoclaving on the surfaces of TiN -coated and conventional nickel-titanium rotary instruments.

G Spagnuolo1, G Ametrano, V D'Antò, C Rengo, M Simeone, F Riccitiello, M Amato.   

Abstract

AIM: To evaluate the effects of repeated autoclave sterilization cycles on surface topography of conventional nickel-titanium ( NiTi ) and titanium nitride ( TiN )-coated rotary instruments.
METHODOLOGY: A total of 60 NiTi rotary instruments, 30 ProTaper (Dentsply Maillefer) and 30 TiN -coated AlphaKite (Komet/Gebr. Brasseler), were analysed. Instruments were evaluated in the as-received condition and after 1, 5 and 10 sterilization cycles. After sterilization, the samples were observed using scanning electron microscope (SEM), and surface chemical analysis was performed on each instrument with energy dispersive X-ray spectroscopy (EDS). Moreover, the samples were analysed by atomic force microscopy (AFM), and roughness average (Ra) and the root mean square value (RMS) of the scanned surface profiles were recorded. Data were analysed by means of anova followed by Tukey's test. RESULT: Scanning electron microscope observations revealed the presence of pitting and deep milling marks in all instruments. EDS analysis confirmed that both types of instruments were composed mainly of nickel and titanium, whilst AlphaKite had additional nitride. After multiple autoclave sterilization cycles, SEM examinations revealed an increase in surface alterations, and EDS values indicated changes in chemical surface composition in all instruments. Ra and RMS values of ProTaper significantly increased after 5 (P = 0.006) and 10 cycles (P = 0.002) with respect to the as-received instruments, whilst AlphaKite showed significant differences compared with the controls after 10 cycles (P = 0.03).
CONCLUSIONS: Multiple autoclave sterilization cycles modified the surface topography and chemical composition of conventional and TiN -coated NiTi rotary instruments.
© 2012 International Endodontic Journal.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22757632     DOI: 10.1111/j.1365-2591.2012.02088.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  14 in total

1.  Effect of multiple autoclave cycles on the surface roughness of HyFlex CM and HyFlex EDM files: an atomic force microscopy study.

Authors:  K Yılmaz; G Uslu; T Özyürek
Journal:  Clin Oral Investig       Date:  2018-02-13       Impact factor: 3.573

2.  Effect of autoclave sterilization on cyclic fatigue and torsional fracture resistance of NiTi rotary instruments.

Authors:  Wooyoung Kim; Soram Oh; Gil-Joo Ryu; Tae-Hwan Kim; Sung-Jae Kim; Dong-Hyung Kim; Bin-Na Lee; Kee-Yeon Kum; Seok Woo Chang; Ji-Hyun Jang
Journal:  Odontology       Date:  2019-09-11       Impact factor: 2.634

3.  Microstructural, microchemical, and mechanical changes associated with the clinical reuse of two nickel-titanium endodontic instruments.

Authors:  Felipe Augusto Restrepo-Restrepo; Viviana Andrea Holguín-Vásquez; Syldana Julieth Cañas-Jiménez; Paula Andrea Villa-Machado; Sara Ochoa-Soto; Claudia Patricia Ossa-Orozco; Sergio Iván Tobón-Arroyave
Journal:  Dent Res J (Isfahan)       Date:  2021-06-22

4.  Effects of intraoral aging on surface properties of coated nickel-titanium archwires.

Authors:  Roberto Rongo; Gianluca Ametrano; Antonio Gloria; Gianrico Spagnuolo; Angela Galeotti; Sergio Paduano; Rosa Valletta; Vincenzo D'Antò
Journal:  Angle Orthod       Date:  2013-12-05       Impact factor: 2.079

5.  Effect of repeated sterilization cycles on the physical properties of scaling instruments: a scanning electron microscopy study.

Authors:  Alessandra Nogueira Porto; Álvaro Henrique Borges; Alex Semenoff-Segundo; Suzane A Raslan; Fábio Luis Miranda Pedro; Antônio Olavo Cardoso Jorge; Matheus Coelho Bandeca
Journal:  J Int Oral Health       Date:  2015-05

6.  Clinical and Spectrophotometric Evaluation of LED and Laser Activated Teeth Bleaching.

Authors:  R Lo Giudice; G Pantaleo; A Lizio; U Romeo; G Castiello; G Spagnuolo; G Lo Giudice
Journal:  Open Dent J       Date:  2016-05-31

7.  Surface and cross-sectional characterization of titanium-nitride coated nickel-titanium endodontic files.

Authors:  Youssef S Al Jabbari; Theodoros Koutsoukis; Suliman Al Hadlaq; David W Berzins; Spiros Zinelis
Journal:  J Dent Sci       Date:  2015-11-19       Impact factor: 2.080

8.  Effects of various prophylactic procedures on titanium surfaces and biofilm formation.

Authors:  Anna Di Salle; Gianrico Spagnuolo; Raffaele Conte; Alfredo Procino; Gianfranco Peluso; Carlo Rengo
Journal:  J Periodontal Implant Sci       Date:  2018-12-27       Impact factor: 2.614

9.  Effect of Autoclave Cycles on Surface Characteristics of S-File Evaluated by Scanning Electron Microscopy.

Authors:  Hamid Razavian; Pedram Iranmanesh; Hamid Mojtahedi; Rahman Nazeri
Journal:  Iran Endod J       Date:  2015-12-24

10.  The IG- file use to Gauge the Apical Diameter in Endodontics: An In Vitro Study.

Authors:  Massimo Amato; Alfredo Iandolo; Giuseppe Pantaleo; Dina Abtellatif; Michele Simeone; Angelo Lizio; Roberto Lo Giudice; Giuseppe Lo Giudice
Journal:  Open Dent J       Date:  2018-09-28